Quantification of key folate forms in serum using stable-isotope dilution ultra performance liquid chromatography-tandem mass spectrometry

被引:53
|
作者
Kirsch, Susanne H. [1 ]
Knapp, Jean-Pierre [1 ]
Herrmann, Wolfgang [1 ]
Obeid, Rima [1 ]
机构
[1] Saarland Univ Hosp, Dept Clin Chem & Lab Med, D-66421 Homburg, Germany
关键词
Folate; 5-Methyltetrahydrofolate; Folic acid; Homocysteine; Tandem mass spectrometry; NEURAL-TUBE DEFECTS; METHYLENETETRAHYDROFOLATE REDUCTASE; FOLIC-ACID; THERMAL-DESTRUCTION; SAMPLE PREPARATION; VASCULAR-DISEASE; COMMON MUTATION; RISK-FACTOR; RADIOASSAY; FOLACIN;
D O I
10.1016/j.jchromb.2009.11.021
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Folates act as essential coenzymes in many biological pathways. Alteration in folate form distribution might have biological significance, especially in relation to certain genetic polymorphisms. We developed a stable-isotope dilution ultra performance liquid chromatography-mass spectrometry (UPLC-MS/MS) method for quantification of the folate forms 5-methyltetrahydrofolate (5-methylTHF), 5-formylTHF, 5,10-methenylTHF, THF, and folic acid in serum. After extraction using an ion exchange and mixed mode solid-phase, samples were separated and detected using an UPLC-MS/MS system. The quantification limits were between 0.17 nmol/L (5-formylTHF) and 1.79 nmol/L (THF), and the assay was linear up to 100 nmol/L (5-methylTHF) and 10 nmol/L (5-formylTHF, 5,10-methenylTHF, THF, and folic acid). The intraassay CVs for 5-methylTHF and 5-formylTHF were 2.0% and 7.2%, respectively. Mean recoveries were between 82.3% for THF and 110.8% for 5,10-methenylTHF. Concentrations of total folate measured by the new method showed a strong correlation with those measured by an immunologic assay (r = 0.939; p, < 0.001). The mean total folate from 32 apparently healthy subjects was 18.09 nmol/L, of which 87.23% was 5-methylTHF. Concentrations of homocysteine showed a better correlation to the total folate measured by the new method compared to that obtained by an immunologic assay. We also confirmed that MTHFR polymorphism has a significant effect on folate distribution in this small population of non-supplemented subjects. (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:68 / 75
页数:8
相关论文
共 50 条
  • [41] Quantification of levetiracetam in plasma of neonates by ultra performance liquid chromatography-tandem mass spectrometry
    Blonk, Maren I.
    van der Nagel, Bart C.
    Smit, Liesbeth S.
    Mathot, Ron A. A.
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2010, 878 (7-8): : 675 - 681
  • [42] Quantification of dsRNA using stable isotope labeling dilution liquid chromatography/mass spectrometry
    Kung, An-Wen
    Kilby, Peter M.
    Portwood, David E.
    Dickman, Mark J.
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2018, 32 (07) : 590 - 596
  • [43] A reference measurement procedure for amino acids in blood using isotope dilution ultra-performance liquid chromatography-tandem mass spectrometry
    Kim, Juok
    Tran, Thi Thanh Huong
    Hong, Seon-Pyo
    Jeong, Ji-Seon
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2017, 1055 : 72 - 80
  • [44] Stable Isotope Dilution Analysis of Gibberellin Residues in Tomato Paste by Liquid Chromatography-Tandem Mass Spectrometry
    SUN Li1
    2. School of Food and Biological Engineering
    3. Xinjiang Extry-exit Inspection and Quarantine Bureau
    Chemical Research in Chinese Universities, 2012, 28 (05) : 797 - 801
  • [45] Quantification of folate metabolites in serum using ultraperformance liquid chromatography tandem mass spectrometry
    Wang, Xiuwei
    Zhang, Ting
    Zhao, Xin
    Guan, Zhen
    Wang, Zhen
    Zhu, Zhiqiang
    Xie, Qiu
    Wang, Jianhua
    Niu, Bo
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2014, 962 : 9 - 13
  • [46] Optimization of Enzyme Digestion Conditions for Quantification of Glycated Hemoglobin Using Isotope Dilution Liquid Chromatography-Tandem Mass Spectrometry
    Jeong, Ji-Seon
    MASS SPECTROMETRY LETTERS, 2014, 5 (02) : 52 - 56
  • [47] Stable Isotope Dilution Analysis of Gibberellin Residues in Tomato Paste by Liquid Chromatography-Tandem Mass Spectrometry
    Sun Li
    Zhao Yan-sheng
    Nie Xue-mei
    Ling Yun
    Chu Xiao-gang
    Shang De-jun
    Dong Ying
    CHEMICAL RESEARCH IN CHINESE UNIVERSITIES, 2012, 28 (05) : 797 - 801
  • [48] Liquid Chromatography-Tandem Mass Spectrometry Analysis of Folate and Folate Catabolites in Human Serum
    Hannisdal, Rita
    Ueland, Per Magne
    Svardal, Asbjorn
    CLINICAL CHEMISTRY, 2009, 55 (06) : 1147 - 1154
  • [49] Quantification of recombinant human erythropoietin by amino acid analysis using isotope dilution liquid chromatography-tandem mass spectrometry
    Yim, Jung-Hyuk
    Yoon, Ina
    Yang, Hyo-Jin
    Kim, Sook-Kyung
    Park, Sang-Ryoul
    Lee, Yong-Moon
    Jeong, Ji-Seon
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2014, 406 (18) : 4401 - 4409
  • [50] Simultaneous measurement of amiodarone and desethylamiodarone in human plasma and serum by stable isotope dilution liquid chromatography-tandem mass spectrometry assay
    Kuhn, Joachim
    Goetting, Christian
    Kleesiek, Knut
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2010, 51 (01) : 210 - 216