Development of a multiplex immunochromatographic strip test and ultrasensitive electrochemical immunosensor for hepatitis B virus screening

被引:36
作者
Akkapinyo, Chutimon [1 ]
Khownarumit, Porntip [2 ]
Waraho-Zhmayev, Dujduan [1 ]
Poo-arporn, Rungtiva P. [1 ]
机构
[1] King Mongkuts Univ Technol Thonburi, Biol Engn Program, Bangkok 10140, Thailand
[2] King Mongkuts Univ Technol Thonburi, PDTI, Sensor Technol, Bangkok 10150, Thailand
关键词
Hepatitis B; Immunosensor; HBsAg; Impedance spectroscopy; Test strip; SURFACE-ANTIGEN; IMPEDANCE IMMUNOSENSOR; ANTIBODY; IMMOBILIZATION; GOLD; NANOPARTICLES; FUNCTIONALIZATION; IMMUNOASSAY; POLYPYRROLE; ELECTRODES;
D O I
10.1016/j.aca.2019.10.016
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This research proposes two methods for hepatitis B diagnosis including rapid testing and electrochemical assay. For the first method, a multiplex hepatitis B test strip was fabricated to serve as a rapid test for hepatitis B screening. It was developed to simultaneously test three essential serological markers of hepatitis B virus infection including hepatitis B surface antigen (HBsAg), hepatitis B surface antibody (Anti-HBs) and hepatitis B core antibody (Anti-HBc). Gold nanoparticles (GNPs) were used as the signal generator on the test strip. Furthermore, a part of a paper network was incorporated on the strip for the gold-silver enhancement process. This paper network helped in decreasing the analysis time of enhancement and makes the enhancement process easier for rapid testing. The developed test strip was specific for each serological marker. The detection limits of HBsAg, Anti-HBs and Anti-HBc were obtained at 0.5, 0.3 and 0.1 mu g mL(-1), respectively. For the second method, electrochemical impedance spectroscopy (EIS) was applied for HBsAg detection. This method was proposed for quantitative hepatitis B detection. Anti-HBs antibodies were immobilized on a carbon screen printed electrode (SPCE) via the N-ethyl-N'-(3-(dimethylamino)propyl)carbo-diimide/N-hydroxy succinimide (EDC/NHS) couple reaction which reacted with the carboxyl group of the BSA cross-linked film on the electrode. The electrode modification process was characterized by EIS. A linear relationship between delta charge transfer resistance (Delta Rct) and HBsAg concentration was obtained in the range of 5-3000 ng mL(-1) with a detection limit of 2.1 ng mL(-1). This work is appropriate for quantitative analysis because it is a simple and low-cost method to implement as the SPCE is disposable. Therefore, we hope that this research will be useful to improve hepatitis B detection in the future. (C) 2019 Elsevier B.V. All rights reserved.
引用
收藏
页码:162 / 171
页数:10
相关论文
共 41 条
  • [1] On the physical interpretation of constant phase elements
    Abouzari, M. R. Shoar
    Berkemeier, F.
    Schmitz, G.
    Wilmer, D.
    [J]. SOLID STATE IONICS, 2009, 180 (14-16) : 922 - 927
  • [2] [Anonymous], [No title captured]
  • [3] [Anonymous], [No title captured]
  • [4] Ochratoxin A Detection on Antibody-Immobilized on BSA-Functionalized Gold Electrodes
    Badea, Mihaela
    Floroian, Laura
    Restani, Patrizia
    Cobzac, Simona Codruta Aurora
    Moga, Marius
    [J]. PLOS ONE, 2016, 11 (07):
  • [5] Electrochemical impedance immunosensor for rapid detection of stressed pathogenic Staphylococcus aureus bacteria
    Bekir, Karima
    Barhoumi, Houcine
    Braiek, Mohamed
    Chrouda, Amani
    Zine, Nadia
    Abid, Nabil
    Maaref, Abdelrazek
    Bakhrouf, Amina
    Ben Ouada, Hafedh
    Jaffrezic-Renault, Nicole
    Ben Mansour, Hedi
    [J]. ENVIRONMENTAL SCIENCE AND POLLUTION RESEARCH, 2015, 22 (20) : 15796 - 15803
  • [6] Non-specific binding of antibodies in immunohistochemistry: fallacies and facts
    Buchwalow, Igor
    Samoilova, Vera
    Boecker, Werner
    Tiemann, Markus
    [J]. SCIENTIFIC REPORTS, 2011, 1
  • [7] Label-free electrochemical impedance immunosensor based on modified screen-printed gold electrodes for the diagnosis of canine visceral leishmaniasis
    Cordeiro, Tais A. R.
    Goncalves, Marcus V. C.
    Franco, Diego L.
    Reis, Alexandre B.
    Martins, Helen R.
    Ferreira, Lucas F.
    [J]. TALANTA, 2019, 195 : 327 - 332
  • [8] Immunochromatographic methods in food analysis
    Dzantiev, Boris B.
    Byzova, Nadezhda A.
    Urusov, Alexandr E.
    Zherdev, Anatoly V.
    [J]. TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2014, 55 : 81 - 93
  • [9] Evaluation of commercialized rapid diagnostic testing for some Hepatitis B biomarkers in an area of intermediate endemicity
    EI-Ghitany, Engy Mohamed
    Farghaly, Azza Galal
    [J]. JOURNAL OF VIROLOGICAL METHODS, 2013, 194 (1-2) : 190 - 193
  • [10] Two-Dimensional Paper Network Format That Enables Simple Multistep Assays for Use in Low-Resource Settings in the Context of Malaria Antigen Detection
    Fu, Elain
    Liang, Tinny
    Spicar-Mihalic, Paolo
    Houghtaling, Jared
    Ramachandran, Sujatha
    Yager, Paul
    [J]. ANALYTICAL CHEMISTRY, 2012, 84 (10) : 4574 - 4579