Accuracy and clinical applicability of a passive marker based frameless neuronavigation system

被引:34
|
作者
Muacevic, A [1 ]
Uhl, E [1 ]
Steiger, HJ [1 ]
Reulen, HJ [1 ]
机构
[1] Univ Munich, Klinikum Grosshadern, Dept Neurosurg, D-81377 Munich, Germany
关键词
computer-aided surgery; neuronavigation; frameless stereotaxy; passive marker systems;
D O I
10.1054/jocn.1999.0226
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
A passive marker system permits the inclusion of an unlimited number of instruments and other devices during frameless stereotaxy. The aim of this study was to evaluate the accuracy and clinical applicability of a passive marker based frameless image guided system (VectorVision; BrainLab, Heimstetten, Germany) developed for surgical planning and intraoperative image guidance. The system was first applied to a plastic phantom to determine the accuracy of the system by measuring the difference between the actual probe position and its analogous position on the monitor screen. The navigational device was subsequently applied to 40 procedures for brain tumours and cavernomas. The mean error value between the image on the monitor screen and the real location in the phantom and the clinical study was 1.45 mm (+/- 0.99) and 4.05 mm (+/- 3.62), respectively. Many different instruments could be employed as pointing devices. It was helpful in minimising the size of the craniotomy. An average lengthening of the surgical procedure of 20 minutes was experienced. The neuronavigation system proved to be a useful surgical tool to approach and detect lesions larger than 5 mm in diameter. The passive marker technology is intuitive and enables the surgeon to use his or her own instruments at any time as a pointing device, thus avoiding further costs for specially designed surgical equipment. (C) 2000 Harcourt Publishers Ltd.
引用
收藏
页码:414 / 418
页数:5
相关论文
共 50 条
  • [31] Image fusion of MR images and real-time ultrasonography: evaluation of fusion accuracy combining two commercial instruments, a neuronavigation system and a ultrasound system
    Schlaier, JR
    Warnat, J
    Dorenbeck, U
    Proescholdt, M
    Schebesch, KM
    Brawanski, A
    ACTA NEUROCHIRURGICA, 2004, 146 (03) : 271 - 277
  • [32] Characterization of the Targeting Accuracy of a Neuronavigation-Guided Transcranial FUS System In Vitro, In Vivo, and In Silico
    Xu, Lu
    Pacia, Christopher P.
    Gong, Yan
    Hu, Zhongtao
    Chien, Chih-Yen
    Yang, Leqi
    Gach, H. Michael
    Hao, Yao
    Comron, Hassanzadeh
    Huang, Jiayi
    Leuthardt, Eric C.
    Chen, Hong
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2023, 70 (05) : 1528 - 1538
  • [33] Comparison of accuracy and precision between frame-based and frameless stereotactic navigation for deep brain stimulation electrode implantation
    Bjartmarz, Hjalmar
    Rehncrona, Stig
    STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2007, 85 (05) : 235 - 242
  • [34] Application of neuronavigation system to brain tumor surgery with clinical experience of 420 cases
    Jung, T. -Y.
    Jung, S.
    Kim, I. -Y.
    Park, S. -J.
    Kang, S. -S.
    Kim, S. -H.
    Lim, S. -C.
    MINIMALLY INVASIVE NEUROSURGERY, 2006, 49 (04) : 210 - 215
  • [35] Surface-based facial scan registration in neuronavigation procedures: a clinical study Clinical article
    Shamir, Reuben R.
    Freiman, Moti
    Joskowicz, Leo
    Spektor, Sergey
    Shoshan, Yigal
    JOURNAL OF NEUROSURGERY, 2009, 111 (06) : 1201 - 1206
  • [36] Clinical experience of image-guided neurosurgery with a frameless navigation system (StealthStation™)
    Morioka, T
    Nishio, S
    Ikezaki, K
    Natori, Y
    Inamura, T
    Muratani, H
    Muraishi, M
    Hisada, K
    Mihara, F
    Matsushima, T
    Fukui, M
    NEUROLOGICAL SURGERY, 1999, 27 (01): : 33 - 40
  • [37] Intraoperative computed tomography registration and electromagnetic neuronavigation for transsphenoidal pituitary surgery: accuracy and time effectiveness Clinical article
    Eboli, Paula
    Shafa, Bob
    Mayberg, Marc
    JOURNAL OF NEUROSURGERY, 2011, 114 (02) : 329 - 335
  • [38] Comparison of laser surface scanning and fiducial marker-based registration in frameless stereotaxy - Technical note
    Schicho, Kurt
    Figl, Michael
    Seemann, Rudolf
    Donat, Markus
    Pretterklieber, Michael L.
    Birkfellner, Wolfgang
    Reichwein, Astrid
    Wanschitz, Felix
    Kainberger, Franz
    Bergmann, Helmar
    Wagner, Arne
    Ewers, Rolf
    JOURNAL OF NEUROSURGERY, 2007, 106 (04) : 704 - 709
  • [39] APPLICATION AND MACHINE ACCURACY OF A NEW FRAMELESS COMPUTED TOMOGRAPHY-GUIDED STEREOTACTIC BRAIN BIOPSY SYSTEM IN DOGS
    Taylor, Amanda R.
    Cohen, Noah D.
    Fletcher, Stephen
    Griffin, John F.
    Levine, Jonathan M.
    VETERINARY RADIOLOGY & ULTRASOUND, 2013, 54 (04) : 332 - 342
  • [40] New prototype neuronavigation system based on preoperative imaging and intraoperative freehand ultrasound: system description and validation
    Mercier, Laurence
    Del Maestro, Rolando F.
    Petrecca, Kevin
    Kochanowska, Anna
    Drouin, Simon
    Yan, Charles X. B.
    Janke, Andrew L.
    Chen, Sean Jy-Shyang
    Collins, D. Louis
    INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2011, 6 (04) : 507 - 522