Patient and operating room staff radiation dose during fenestrated/branched endovascular aneurysm repair using premanufactured devices

被引:14
作者
Kirkwood, Melissa L. [1 ]
Chamseddin, Khalil [1 ]
Arbique, Gary M. [2 ]
Guild, Jeffrey B. [2 ]
Timaran, David [1 ]
Anderson, Jon A. [2 ]
Timaran, Carlos [1 ]
机构
[1] Univ Texas Southwestern Med Ctr Dallas, Dept Surg, Div Vasc & Endovasc Surg, Dallas, TX 75390 USA
[2] Univ Texas Southwestern Med Ctr Dallas, Dept Radiol, Div Med Phys, Dallas, TX 75390 USA
关键词
Fluoroscopically guided interventions; Fenestrated endovascular aneurysm repair; FEVAR; Radiation dose; Occupational radiation dose; ABDOMINAL AORTIC-ANEURYSMS; STENT GRAFT; EXPOSURE;
D O I
10.1016/j.jvs.2018.02.031
中图分类号
R61 [外科手术学];
学科分类号
摘要
Introduction: Fenestrated endovascular aneurysm repair (FEVAR) is the highest radiation dose procedure performed by vascular surgeons. We sought to characterize the radiation dose to patients and staff during FEVAR procedures with different premanufactured devices. Methods: A single-center prospective study of FEVARs was performed over 24 months. Three FEVAR devices were included: off-the-shelf (OTS; t-Branch, p-Branch), Zenith Fenestrated (ZFen), and investigational custom-made devices (CMDs). Radiation doses to the surgeon, trainee, anesthesiologist, and scrub/circulating nurses were measured using a personal dosimetry system (DoseAware, Philips Healthcare, Amsterdam, The Netherlands). Procedure type, patient body mass index (BMI), reference air kerma (RAK), and kerma area product (KAP) were recorded. RAK and KAP were corrected for BMI based on an exponential fit of fluoroscopy dose rate and the dose per radiographic frame. Operator dose was corrected for BMI by the ratio of corrected to actual KAP. A one-sided Wilcox rank-sum test was used to compare personnel radiation doses, RAKs, and KAPs between procedure types. Statistical significance was set at P <=.05. Results: There were 80 FEVARs performed by a single surgeon on a Philips Allura XperFD20 fluoroscopy system equipped with Clarity technology. Average BMI was 27 kg/m2. Sixty CMDs (36 four-, 21 three-, and 3 two-vessel fenestrations), 11 ZFens (8 three-and 3 two-vessel fenestrations), and 9 OTS devices (4 p-Branch, 5 t-Branch) were included. ZFens had significantly lower patient (1800 mGy vs 2950 mGy; P=.004), operator (120 mu Sv vs 370 mu Sv; P=.004), assistant (60 mu Sv vs 210 mu Sv; P=.003), circulator (10 mu Sv vs 30 mu Sv; P=.049), and scrub nurse dose (10 mu Sv vs 40 mu Sv; P=.02) compared with CMDs. OTS devices had significantly lower operator (220 mu Sv vs 370 mu Sv; P=.04), assistant (110 mu Sv vs 210 mu Sv; P=.02), and circulator doses (4 mu Sv vs 30 mu Sv; P=.001) compared with CMDs. Four-vessel fenestrated devices had significantly higher patient dose (3020 mGy) compared with three-vessel FEVARs (2670 mGy; P=.03) and two-vessel FEVARs (1600 mGy; P=.0007), and significantly higher operator dose (440 mu Sv) compared with three-vessel FEVARs (170 mu Sv; P=.0005). Patient dose was lowest with ZFens. Operating room personnel dose was lower with ZFens and OTS devices compared with CMDs. Fourvessel fenestrations required significantly more radiation compared with those involving three-vessel fenestrations; however, the dose increase was only 12% and should not preclude operators fromextending coverage, if anatomically required. Conclusions: Overall, patient and personnel radiation doses during FEVAR with all devices were within acceptable limits and lower in our series than previously reported.
引用
收藏
页码:1281 / 1286
页数:6
相关论文
共 14 条
[1]   Results from multiple prospective single-center clinical trials of the off-the-shelf p-Branch fenestrated stent graft [J].
Farber, Mark A. ;
Eagleton, Matthew J. ;
Mastracci, Tara M. ;
McKinsey, James F. ;
Vallabhaneni, Raghuveer ;
Sonesson, Bjorn ;
Dias, Nuno ;
Resch, Timothy .
JOURNAL OF VASCULAR SURGERY, 2017, 66 (04) :982-990
[2]   Assessing the anatomic applicability of the multibranched endovascular repair of thoracoabdominal aortic aneurysm technique [J].
Gasper, Warren J. ;
Reilly, Linda M. ;
Rapp, Joseph H. ;
Grenon, S. Marlene ;
Hiramoto, Jade S. ;
Sobel, Julia D. ;
Chuter, Timothy A. M. .
JOURNAL OF VASCULAR SURGERY, 2013, 57 (06) :1553-1558
[3]   Risk of Radiation Exposure during Endovascular Aortic Repair [J].
Howells, P. ;
Eaton, R. ;
Patel, A. S. ;
Taylor, P. ;
Modarai, B. .
EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2012, 43 (04) :393-397
[4]   Comparison of outcomes for double fenestrated endovascular aneurysm repair versus triple or quadruple fenestrated endovascular aneurysm repair in the treatment of complex abdominal aortic aneurysms [J].
Katsargyris, Athanasios ;
Oikonomou, Kyriakos ;
Kouvelos, George ;
Mufty, Hozan ;
Ritter, Wolfgang ;
Verhoeven, Eric L. G. .
JOURNAL OF VASCULAR SURGERY, 2017, 66 (01) :29-36
[5]   Factors affecting radiation injury [J].
Killewich, Lois A. ;
Falls, Garietta ;
Mastracci, Tara M. ;
Brown, Kellie R. .
JOURNAL OF VASCULAR SURGERY, 2011, 53 :9S-14S
[6]   New image-processing and noise-reduction software reduces radiation dose during complex endovascular procedures [J].
Kirkwood, Melissa L. ;
Guild, Jeffrey B. ;
Arbique, Gary M. ;
Tsai, Shirling ;
Modrall, J. Gregory ;
Anderson, Jon A. ;
Rectenwald, John ;
Timaran, Carlos .
JOURNAL OF VASCULAR SURGERY, 2016, 64 (05) :1357-1364
[7]   Surgeon radiation dose during complex endovascular procedures [J].
Kirkwood, Melissa L. ;
Guild, Jeffrey B. ;
Arbique, Gary M. ;
Anderson, Jon A. ;
Valentine, R. James ;
Timaran, Carlos .
JOURNAL OF VASCULAR SURGERY, 2015, 62 (02) :457-463
[8]   Surgeon education decreases radiation dose in complex endovascular procedures and improves patient safety [J].
Kirkwood, Melissa L. ;
Arbique, Gary M. ;
Guild, Jeffrey B. ;
Timaran, Carlos ;
Chung, Jayer ;
Anderson, Jon A. ;
Valentine, R. James .
JOURNAL OF VASCULAR SURGERY, 2013, 58 (03) :715-721
[9]   Off-the-shelf fenestrated and branched stent graft designs for abdominal aortic aneurysm repair [J].
Mendes, Bernardo C. ;
Rathore, Animesh ;
Ribeiro, Mauricio S. ;
Oderich, Gustavo S. .
SEMINARS IN VASCULAR SURGERY, 2016, 29 (1-2) :74-83
[10]   Radiation exposure to operating room personnel and patients during endovascular procedures [J].
Mohapatra, Abhisekh ;
Greenberg, Roy K. ;
Mastracci, Tara M. ;
Eagleton, Matthew J. ;
Thornsberry, Brett .
JOURNAL OF VASCULAR SURGERY, 2013, 58 (03) :702-709