Multichannel SQUID magnetometry using double relaxation oscillation SQUID's

被引:5
|
作者
vanDuuren, MJ
Lee, YH
Adelerhof, DJ
Kawai, J
Kado, H
Flokstra, J
Rogalla, H
机构
[1] KOREAN RES INST STAND & SCI, TAEJON 305606, SOUTH KOREA
[2] SUPERCONDUCTING SENSOR LAB, CHIBA 27013, JAPAN
关键词
D O I
10.1109/77.488279
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A highly sensitive first-order gradiometer based on double relaxation oscillation SQUID's (DROS's) for multichannel use is presented, The white flux noise level of the bare DROS's is 4.5 mu Phi(0)/root Hz (epsilon = 275 h), With wire-wound first-order gradiometers, having a baseline of 40 mm, the white magnetic field noise equals 4 fT/root Hz. As a result of the high flux-to-voltage transfer of the DROS's of about 1 mV/Phi(0), this low noise level could be obtained with simple room-temperature flux locked loop electronics based on direct voltage readout. The relaxation frequency of the present DROS's is approximately 1 GHz, No degradation of the DROS characteristics caused by interference between relaxation oscillations in two adjacent channels has been observed, although the gradiometers are spaced by less than 1 mm, This shows that DROS's fan be used in multichannel SQUID magnetometers.
引用
收藏
页码:38 / 44
页数:7
相关论文
共 50 条
  • [1] Double relaxation oscillation SQUID systems for biomagnetic multichannel measurements
    Lee, YH
    Kwon, H
    Kim, JM
    Kim, K
    Kim, IS
    Park, YK
    IEICE TRANSACTIONS ON ELECTRONICS, 2005, E88C (02): : 168 - 174
  • [2] A multichannel double relaxation oscillation SQUID magnetometer system for magnetoencephalogram.
    Kang, CS
    Lee, YH
    Kwon, H
    Kim, JM
    Kim, K
    Lee, SG
    Park, YK
    PHYSICA STATUS SOLIDI A-APPLICATIONS AND MATERIALS SCIENCE, 2004, 201 (08): : 1956 - 1960
  • [3] A multichannel SQUID magnetometer system based on double relaxation oscillation SQUIDs
    Lee, YH
    Kwon, H
    Kim, JM
    Kang, CS
    Kim, K
    Kim, IS
    Park, YK
    Lee, SG
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2003, 13 (02) : 755 - 758
  • [4] Double relaxation oscillation squid with reference junction for biomagnetic multichannel applications
    Lee, YH
    Kwon, HC
    Kim, JM
    Park, YK
    Park, JC
    APPLIED SUPERCONDUCTIVITY, 1997, 5 (7-12) : 413 - 418
  • [5] Noise Properties of Digital SQUID Using Double Relaxation Oscillation SQUID Comparator With Relaxation Oscillation Resonant Circuit
    Myoren, Hiroaki
    Kobayashi, Ryota
    Kumagai, Kenji
    Naruse, Masato
    Taino, Tohru
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2017, 27 (04)
  • [6] Proposal of a digital double relaxation oscillation SQUID
    Myoren, H
    Nakamura, M
    Iizuka, T
    Takada, S
    IEICE TRANSACTIONS ON ELECTRONICS, 2001, E84C (01) : 49 - 54
  • [7] Design of a fast digital double relaxation oscillation SQUID
    Podt, M
    Mieog, AJ
    Flokstra, J
    Rogalla, H
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2001, 11 (01) : 1235 - 1238
  • [8] An integrated planar gradiometer based on a double relaxation oscillation SQUID
    Lee, YH
    Kwon, HC
    Kim, JM
    Park, YK
    Park, JC
    SUPERCONDUCTOR SCIENCE & TECHNOLOGY, 1996, 9 (4A): : A34 - A37
  • [9] A digital Double Relaxation Oscillation SQUID for particle detector readout
    Podt, M
    Keizer, D
    Flokstra, J
    Rogalla, H
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2000, 444 (1-2): : 120 - 123
  • [10] Design of a fast digital Double Relaxation Oscillation SQUID (Corrected)
    Podt, M
    Mieog, AJ
    Flokstra, J
    Rogalla, H
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2001, 11 (02) : 4054 - 4057