Evaluation of implantable actuators by means of a middle ear simulation model

被引:47
作者
Bornitz, Matthias [1 ]
Hardtke, Hans-Juergen [2 ]
Zahnert, Thomas [1 ]
机构
[1] Tech Univ Dresden, Dept Med, Clin Otolaryngol, D-01307 Dresden, Germany
[2] Tech Univ Dresden, Dept Mech Engn, Inst Solid Mech, D-01307 Dresden, Germany
关键词
Middle ear; Implantable hearing aids; Finite element model; Actuator; HEARING DEVICE; TRANSDUCER; CRYSTAL;
D O I
10.1016/j.heares.2010.02.007
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
The extension of indication of implantable hearing aids to cases of conductive hearing loss pushed the development of these devices. There is now a great variety of devices available with different actuator concepts and different attachment points to the middle ear or inner ear fluid. But there is little comparative data available about the devices to provide an insight into advantages and disadvantages of different types of actuators and attachment points at the ossicular chain. This paper investigates two principle (idealized) types of actuators in respect of attachments points at the ossicular chain and direction of excitation. Other parts of implantable hearing aids like microphone, amplifier and signal processing electronics were not incorporated into this study. Investigations were performed by means of a mathematical simulation model of the middle ear (finite element model). Actuator performance and theoretical gain were calculated by harmonic analysis in the frequency range of 100-6000 Hz and were compared for the different situations. The stapes head proofed to be an ideal attachment point for actuators of both types as this position is very insensitive to changes in the direction of excitation. The implantable actuators showed higher ratio of equivalent sound pressure to radiated sound pressure compared to an open hearing aid transducer and should therefore allow for more functional gain. (C) 2010 Elsevier B.V. All rights reserved.
引用
收藏
页码:145 / 151
页数:7
相关论文
共 23 条
  • [11] Koike T, 2004, PROCEEDINGS OF THE 3RD SYMPOSIUM ON MIDDLE EAR MECHANICS IN RESEARCH AND OTOLOGY, P68
  • [12] Fully implantable hearing systems
    Maurer, J.
    [J]. HNO, 2009, 57 (03) : 199 - +
  • [13] In vitro study of a multi-layer piezoelectric crystal attic hearing implant
    Mills, RP
    Wang, ZG
    Abel, EW
    Eng, C
    [J]. JOURNAL OF LARYNGOLOGY AND OTOLOGY, 2001, 115 (05) : 359 - 362
  • [14] Earlens tympanic contact transducer: A new method of sound transduction to the human ear
    Perkins, R
    [J]. OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 1996, 114 (06) : 720 - 728
  • [15] Testing a method for quantifying the output of implantable middle ear hearing devices
    Rosowski, J. J.
    Chien, W.
    Ravicz, M. E.
    Merchant, S. N.
    [J]. AUDIOLOGY AND NEURO-OTOLOGY, 2007, 12 (04) : 265 - 276
  • [16] Spindel Jonathan H, 2002, Am J Audiol, V11, P104, DOI 10.1044/1059-0889(2002/019)
  • [17] Computer assisted optimization of an electromagnetic transducer design for implantable hearing aids
    Stieger, C
    Wäckerlin, D
    Bernhard, H
    Stahel, A
    Kompis, M
    Häusler, R
    Burger, E
    [J]. COMPUTERS IN BIOLOGY AND MEDICINE, 2004, 34 (02) : 141 - 152
  • [18] Human temporal bones versus mechanical model to evaluate three middle ear transducers
    Stieger, Christof
    Bernhard, Hans
    Waeckerlin, Daniela
    Kornpis, Martin
    Burger, Juergen
    Haeusler, Rudolf
    [J]. JOURNAL OF REHABILITATION RESEARCH AND DEVELOPMENT, 2007, 44 (03) : 407 - 415
  • [19] Hearing rehabilitation by means of implantable hearing devices
    Tisch, M.
    [J]. HNO, 2009, 57 (03) : 196 - +
  • [20] Acoustic responses of the human middle ear
    Voss, SE
    Rosowski, JJ
    Merchant, SN
    Peake, WT
    [J]. HEARING RESEARCH, 2000, 150 (1-2) : 43 - 69