Relative importance of rate and place: Experiments using pitch scaling techniques with cochlear implant recipients

被引:13
作者
Fearn, R [1 ]
Wolfe, J [1 ]
机构
[1] Univ New S Wales, Dept Biophys, Sydney, NSW, Australia
关键词
D O I
10.1177/0003489400109S1221
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Pitch scaling was used to determine the dependence of perceived pitch on rate and place of stimulation in postlingually deafened adult subjects using cochlear implants. For stimulation rates below about 500 pulses per second (pps), perceived pitch is a strong function of both rate and place. In this range, perceived pitch increases logarithmically with stimulation rate, but decreases with distance from the round window. A 2-mm displacement into the cochlea has an effect similar to that of halving the stimulation rate. Place resolution in this context is comparable with the interelectrode spacing (0.75 mm). At rates approaching 1,000 pps, rate has little effect on perceived pitch. An average of bipolar quality judgments showed that periodic pulsatile stimulation is least pleasant when low frequencies are applied to the legion closest to the window.
引用
收藏
页码:51 / 53
页数:3
相关论文
共 13 条
[1]   PITCH PERCEPTION FOR DIFFERENT MODES OF STIMULATION USING THE COCHLEAR MULTIPLE-ELECTRODE PROSTHESIS [J].
BUSBY, PA ;
WHITFORD, LA ;
BLAMEY, PJ ;
RICHARDSON, LM ;
CLARK, GM .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1994, 95 (05) :2658-2669
[2]   Comparison of electrode discrimination, pitch ranking, and pitch scaling data in postlingually deafened adult cochlear implant subjects [J].
Collins, LM ;
Zwolan, TA ;
Wakefield, GH .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1997, 101 (01) :440-455
[3]   THE PITCH OF ELECTRICALLY PRESENTED SINUSOIDS [J].
DORMAN, MF ;
SMITH, M ;
SMITH, L ;
PARKIN, JL .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1994, 95 (03) :1677-1679
[4]  
EDDINGTON DK, 1978, ANN OTOL RHINOL S53, V87
[5]   Musical pitch perception with electrical stimulation of the cochlea [J].
McDermott, HJ ;
McKay, CM .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1997, 101 (03) :1622-1631
[6]   PITCH RANKING WITH NONSIMULTANEOUS DUAL-ELECTRODE ELECTRICAL-STIMULATION OF THE COCHLEA [J].
MCDERMOTT, HJ ;
MCKAY, CM .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1994, 96 (01) :155-162
[7]   PITCH PERCEPTS ASSOCIATED WITH AMPLITUDE-MODULATED CURRENT PULSE TRAINS IN COCHLEAR IMPLANTEES [J].
MCKAY, CM ;
MCDERMOTT, HJ ;
CLARK, GM .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1994, 96 (05) :2664-2673
[8]   Pulse rate matching by cochlear implant patients: Effects of loudness randomization and electrode position [J].
Pijl, S .
EAR AND HEARING, 1997, 18 (04) :316-325
[9]   MELODY RECOGNITION AND MUSICAL INTERVAL PERCEPTION BY DEAF SUBJECTS STIMULATED WITH ELECTRICAL PULSE TRAINS THROUGH SINGLE COCHLEAR IMPLANT ELECTRODES [J].
PIJL, S ;
SCHWARZ, DWF .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1995, 98 (02) :886-895
[10]  
Schulz E., 1994, ADV COCHLEAR IMPLANT, P326