Evaluation of the efficacy of a portable LIBS system for detection of CWA on surfaces

被引:14
|
作者
L'Hermite, D. [1 ]
Vors, E. [1 ]
Vercouter, T. [1 ]
Moutiers, G. [1 ]
机构
[1] CEA, DEN, DPC, SEARS, F-91191 Gif Sur Yvette, France
关键词
LIBS; Portable; Laser; CWA; Chemicals; Detection; Contamination; Surface; INDUCED BREAKDOWN SPECTROSCOPY; PHASE;
D O I
10.1007/s11356-016-6305-1
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Laser-induced breakdown spectroscopy (LIBS) is a laser-based optical technique particularly suited for in situ surface analysis. A portable LIBS instrument was tested to detect surface chemical contamination by chemical warfare agents (CWAs). Test of detection of surface contamination was carried out in a toxlab facility with four CWAs, sarin (GB), lewisite (L1), mustard gas (HD), and VX, which were deposited on different substrates, wood, concrete, military green paint, gloves, and ceramic. The CWAs were detected by means of the detection of atomic markers (As, P, F, Cl, and S). The LIBS instrument can give a direct response in terms of detection thanks to an integrated interface for non-expert users or so called end-users. We have evaluated the capability of automatic detection of the selected CWAs. The sensitivity of our portable LIBS instrument was confirmed for the detection of a CWA at surface concentrations above 15 mu g/cm(2). The simultaneous detection of two markers may lead to a decrease of the number of false positive.
引用
收藏
页码:8219 / 8226
页数:8
相关论文
共 50 条
  • [21] Research on the Portable Virtual Detection System for the Vehicle Performance
    Wu, Ping
    2015 2ND INTERNATIONAL SYMPOSIUM ON ENGINEERING TECHNOLOGY, EDUCATION AND MANAGEMENT (ISETEM 2015), 2015, : 179 - 185
  • [22] Conversion of an industrial LIBS system to double-pulse operation to improve sulphur detection in rubber-based parts
    Klie, B.
    Schwabe, S.
    Lenth, C.
    Giese, U.
    Wackerbarth, H.
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2024, 215
  • [23] Evaluation of a Portable Microchip Electrophoresis Fluorescence Detection System for the Analysis of Amino Acid Neurotransmitters in Brain Dialysis Samples
    Oborny, Nathan J.
    Melo Costa, Elton E.
    Suntornsuk, Leena
    Abreu, Fabiane C.
    Lunte, Susan M.
    ANALYTICAL SCIENCES, 2016, 32 (01) : 35 - 40
  • [24] Portable standoff Raman system for fast detection of homemade explosives through glass, plastic and water
    Misra, Anupam K.
    Sharma, Shiv K.
    Acosta, Tayro E.
    Porter, John N.
    Lucey, Paul G.
    Bates, David E.
    CHEMICAL, BIOLOGICAL, RADIOLOGICAL, NUCLEAR, AND EXPLOSIVES (CBRNE) SENSING XIII, 2012, 8358
  • [25] Portable cell-based biosensor system for toxin detection
    Pancrazio, JJ
    Bey, PP
    Cuttino, DS
    Kusel, JK
    Borkholder, DA
    Shaffer, KM
    Kovacs, GTA
    Stenger, DA
    SENSORS AND ACTUATORS B-CHEMICAL, 1998, 53 (03) : 179 - 185
  • [26] A Portable Impedance Immunosensing System for Rapid Detection of Salmonella Typhimurium
    Wen, Tao
    Wang, Ronghui
    Sotero, America
    Li, Yanbin
    SENSORS, 2017, 17 (09):
  • [27] Development of a detection tablet for the portable NO2 monitoring system
    Nakano, N
    Ishiguro, G
    Kawabe, T
    Kono, M
    Muramatsu, S
    INDOOR AIR 2005: PROCEEDINGS OF THE 10TH INTERNATIONAL CONFERENCE ON INDOOR AIR QUALITY AND CLIMATE, VOLS 1-5, 2005, : 1952 - 1956
  • [28] Development of a portable total nitrogen detection system based on microelectrodes
    Bian, C. (cbian@mail.ie.ac.cn), 1600, SAGE Publications Ltd (228): : 46 - 51
  • [29] A Novel Portable Seizure Detection Alarm System: Preliminary Results
    Kramer, Uri
    Kipervasser, Svetlana
    Shlitner, Arie
    Kuzniecky, Ruben
    JOURNAL OF CLINICAL NEUROPHYSIOLOGY, 2011, 28 (01) : 36 - 38
  • [30] Radon fast detection and environmental monitoring with a portable wireless system
    Nodari, Benedetta
    Caldara, Michele
    Re, Valerio
    Fabris, Lorenzo
    2015 6TH IEEE INTERNATIONAL WORKSHOP ON ADVANCES IN SENSORS AND INTERFACES (IWASI), 2015, : 254 - 259