IMPLEMENTATION OF A NEW METHOD FOR DYNAMIC MULTILEAF COLLIMATOR TRACKING OF PROSTATE MOTION IN ARC RADIOTHERAPY USING A SINGLE KV IMAGER

被引:45
作者
Poulsen, Per Rugaard [1 ,2 ]
Cho, Byungchul [1 ]
Sawant, Amit [1 ]
Keall, Paul J. [1 ]
机构
[1] Stanford Univ, Dept Radiat Oncol, Stanford, CA 94305 USA
[2] Aarhus Univ, Dept Oncol, Dept Med Phys, Aarhus, Denmark
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2010年 / 76卷 / 03期
关键词
Image-guided radiotherapy; Intrafraction motion; Prostate; Tumor tracking; REAL-TIME; MODULATED ARC; INTRAFRACTION MOTION; RESPIRATORY MOTION; TARGET TRACKING; IMRT DELIVERY; THERAPY; FEASIBILITY; POSITION; TUMOR;
D O I
10.1016/j.ijrobp.2009.06.073
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To implement a method for real-time prostate motion estimation with a single kV imager during are radiotherapy and to integrate it with dynamic multileaf collimator (DMLC) target tracking. Methods and Materials: An arc field with a circular aperture and 358 degrees gantry rotation was delivered to a motion phantom with a fiducial marker under continuous kV X-ray imaging at 5 Hz, perpendicular to the treatment beam. A pretreatment gantry rotation of 120 degrees in 20 see with continuous imaging preceded the treatment. During treatment, each kV image was first used together with all previous images to estimate the three-dimensional (3D) target probability density function and then used together with this probability density function to estimate the 3D target position. The MLC aperture was then adapted to the estimated 3D target position. Tracking was performed with five patient-measured prostate trajectories that represented characteristic prostate motion patterns. Two data sets were recorded during tracking: (1) the estimated 3D target positions, for off-line comparison with the actual phantom motion; and (2) continuous portal images, for independent off-line calculation of the 2D tracking error as the positional difference between the marker and the MLC aperture center in each portal image. All experiments were also made with 1-Hz kV imaging. Results: The mean 3D root-mean-square error of the trajectory estimation was 0.6 mm. The mean root-mean-square tracking error was 0.7 mm, both parallel and perpendicular to the MLC. The accuracy degraded slightly for 1-Hz imaging. Conclusions: Single-imager DMLC prostate tracking that allows arbitrary beam modulation during arc radiotherapy was implemented. It has submillimeter accuracy for most prostate motion types. (C) 2010 Elsevier Inc.
引用
收藏
页码:914 / 923
页数:10
相关论文
共 28 条
[1]   Prostate intrafraction motion evaluation using kV fluoroscopy during treatment delivery: A feasibility and accuracy study [J].
Adamson, Justus ;
Wu, Qiuwen .
MEDICAL PHYSICS, 2008, 35 (05) :1793-1806
[2]   FIRST DEMONSTRATION OF COMBINED KV/MV IMAGE-GUIDED REAL-TIME DYNAMIC MULTILEAF-COLLIMATOR TARGET TRACKING [J].
Cho, Byungchul ;
Poulsen, Per R. ;
Sloutsky, Alex ;
Sawant, Amit ;
Keall, Paul J. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 74 (03) :859-867
[3]   Real-time intra-fraction-motion tracking using the treatment couch: a feasibility study [J].
D'Souza, WD ;
Naqvi, SA ;
Yu, CX .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (17) :4021-4033
[4]   Intrafraction prostate motion during IMRT for prostate cancer [J].
Huang, E ;
Dong, L ;
Chandra, A ;
Kuban, DA ;
Rosen, II ;
Evans, A ;
Pollack, A .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 53 (02) :261-268
[5]   Motion adaptive x-ray therapy: a feasibility study [J].
Keall, PJ ;
Kini, VR ;
Vedam, SS ;
Mohan, R .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (01) :1-10
[6]   Multi-institutional clinical experience with the Calypso System in localization and continuous, real-time monitoring of the prostate gland during external radiotherapy [J].
Kupelian, Patrick ;
Willoughby, Twyla ;
Mahadevan, Arul ;
Djemil, Toufik ;
Weinstein, Geoffrey ;
Jani, Shirish ;
Enke, Charles ;
Solberg, Timothy ;
Flores, Nicholas ;
Liu, David ;
Beyer, David ;
Levine, Lisa .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 67 (04) :1088-1098
[7]   Observations on real-time prostate gland motion using electromagnetic tracking [J].
Langen, Katja M. ;
Willoughby, Twyla R. ;
Meeks, Sanford L. ;
Santhanam, Anand ;
Cunningham, Alexis ;
Levine, Lisa ;
Kupelian, Patrick A. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 71 (04) :1084-1090
[8]  
MALINOWSKI K, 2007, P SPIE INT SOC OPT E, V6510
[9]   A fiducial detection algorithm for real-time image guided IMRT based on simultaneous MV and kV imaging [J].
Mao, Weihua ;
Riaz, Nadeem ;
Lee, Louis ;
Wiersma, Rodney ;
Xing, Lei .
MEDICAL PHYSICS, 2008, 35 (08) :3554-3564
[10]   IMRT delivery to a moving target by dynamic MLC tracking: delivery for targets moving in two dimensions in the beam's eye view [J].
McQuaid, D. ;
Webb, S. .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (19) :4819-4839