Computed tomography arthrography of the shoulder with tin filter-based spectral shaping at 100 kV and 140 kV

被引:5
作者
Choi, Yun Seok [1 ]
Choo, Hye Jung [1 ]
Lee, Sun Joo [1 ]
Kim, Dong Wook [1 ]
Han, Ji-Yeon [1 ]
Kim, Da Som [1 ]
机构
[1] Inje Univ, Dept Radiol, Busan Paik Hosp, Busan 47392, South Korea
关键词
Computed tomography; arthrography; shoulder; radiation exposure; tin filter; MULTIDETECTOR CT ARTHROGRAPHY; RADIATION-DOSE REDUCTION; DUAL-ENERGY CT; IMAGE QUALITY; MR-ARTHROGRAPHY; ACCURACY; BEAM;
D O I
10.1177/0284185120965551
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background Tin filter-based spectral shaping has been used for low-dose and ultra-low-dose computed tomography (CT) in several body parts. However, studies of shoulder CT arthrography with spectral shaping are limited. Purpose To investigate image quality and radiation dose of shoulder CT arthrography with tin filter-based spectral shaping at 100 kV (Sn 100 kV) and 140 kV (Sn 140 kV) in comparison with the conventional protocol. Material and Methods Ninety-nine shoulder CT arthrographies with protocols of Sn 100 kV (n = 32), Sn 140 kV (n = 25), and conventional 120 kV (n = 42) were retrospectively evaluated. Qualitative image quality, CT attenuations of intra-articular contrast mixture and tissues, background noise, contrast-to-noise ratios (CNRs), and figures of merit were assessed. Radiation doses were compared. Results CT arthrographies with Sn 100 kV and Sn 140 kV yielded approximately 70% and 60% radiation dose reduction, respectively, compared with the conventional 120 kV (P < 0.001). Qualitative image noise and quantitative background noise of Sn 100 kV and Sn 140 kV were significantly less than those of the conventional protocol. Qualitative image contrast, CT attenuations of intra-articular contrast mixture and tissues, and CNRs for Sn 100 were similar to those of the conventional 120 kV. However, Sn 140 kV showed significantly lower qualitative contrast and CNRs than 120 kV. Sn 100 kV was the most dose efficient among the three protocols. Conclusion Shoulder CT arthrography with Sn 100 kV substantially reduced radiation dose and image noise and maintained image contrast, compared with the conventional protocol.
引用
收藏
页码:1349 / 1357
页数:9
相关论文
共 30 条
[1]   Preoperative Imaging of Anterior Shoulder Instability: Diagnostic Effectiveness of MDCT Arthrography and Comparison With MR Arthrography and Arthroscopy [J].
Acid, Souad ;
Le Corroller, Thomas ;
Aswad, Richard ;
Pauly, Vanessa ;
Champsaur, Pierre .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2012, 198 (03) :661-667
[2]   Optimizing radiation dose parameters in MDCT arthrography of the shoulder: illustration of basic concepts in a cadaveric study [J].
Aguet, Julien ;
Becce, Fabio ;
Dunet, Vincent ;
Vlassenbroek, Alain ;
Coche, Emmanuel E. ;
Omoumi, Patrick .
SKELETAL RADIOLOGY, 2019, 48 (08) :1261-1268
[3]   Comparison of Image Quality of Shoulder CT Arthrography Conducted Using 120 kVp and 140 kVp Protocols [J].
Ahn, Se Jin ;
Hong, Sung Hwan ;
Chai, Jee Won ;
Choi, Ja-Young ;
Yoo, Hye Jin ;
Kim, Sae Hoon ;
Kang, Heung Sik .
KOREAN JOURNAL OF RADIOLOGY, 2014, 15 (06) :739-745
[4]   Intravenous Contrast Medium Administration and Scan Timing at CT: Considerations and Approaches [J].
Bae, Kyongtae T. .
RADIOLOGY, 2010, 256 (01) :32-61
[5]   Radiation dose reduction initiative: Effect on image quality in shoulder CT imaging [J].
Boothe, Ethan K. ;
Tenorio, Lulu L. ;
Zabak, Elaina M. ;
Omar, Hythem A. ;
Guild, Jeffrey ;
Xi, Yin ;
Abbara, Suhny ;
Chhabra, Avneesh .
EUROPEAN JOURNAL OF RADIOLOGY, 2017, 95 :118-123
[6]   Chest CT using spectral filtration: radiation dose, image quality, and spectrum of clinical utility [J].
Braun, Franziska M. ;
Johnson, Thorsten R. C. ;
Sommer, Wieland H. ;
Thierfelder, Kolja M. ;
Meinel, Felix G. .
EUROPEAN RADIOLOGY, 2015, 25 (06) :1598-1606
[7]   Multidetector computed tomography arthrography of the shoulder: Diagnostic accuracy and indications [J].
De Filippo, M. ;
Bertellini, A. ;
Sverzellati, N. ;
Pogliacomi, F. ;
Costantino, C. ;
Vitale, M. ;
Zappia, M. ;
Corradi, D. ;
Garlaschi, G. ;
Zompatori, M. .
ACTA RADIOLOGICA, 2008, 49 (05) :540-549
[8]   Assessment of the long head of the biceps tendon of the shoulder with 3T magnetic resonance arthrography and CT arthrography [J].
De Maeseneer, Michel ;
Boulet, Cedric ;
Pouliart, Nicole ;
Kichouh, Mimoun ;
Buls, Nico ;
Verhelle, Filip ;
De Mey, Johan ;
Shahabpour, Maryam .
EUROPEAN JOURNAL OF RADIOLOGY, 2012, 81 (05) :934-939
[9]   Multisection CT Protocols: Sex- and Age-specific Conversion Factors Used to Determine Effective Dose from Dose-Length Product [J].
Deak, Paul D. ;
Smal, Yulia ;
Kalender, Willi A. .
RADIOLOGY, 2010, 257 (01) :158-166
[10]   Low-dose abdominal computed tomography for detection of urinary stone disease - Impact of additional spectral shaping of the X-ray beam on image quality and dose parameters [J].
Dewes, Patricia ;
Frellesen, Claudia ;
Scholtz, Jan-Erik ;
Fischer, Sebastian ;
Vogl, Thomas J. ;
Bauer, Ralf W. ;
Schulz, Boris .
EUROPEAN JOURNAL OF RADIOLOGY, 2016, 85 (06) :1058-1062