Three-Dimensional Synthetic Aperture Focusing Using a Rocking Convex Array Transducer

被引:21
作者
Andresen, Henrik [1 ]
Nikolov, Svetoslav Ivanov [2 ]
Pedersen, Mads Moller [3 ]
Buckton, Daniel [4 ]
Jensen, Jorgen Arendt [1 ]
机构
[1] Tech Univ Denmark, DTU Elektro, Ctr Fast Ultrasound Imaging, DK-2800 Lyngby, Denmark
[2] BK Med, R&D Dept, Herlev, Denmark
[3] Rigshosp, Dept Radiol, DK-2100 Copenhagen, Denmark
[4] Kretztechnik, GE Med Syst, Zipf, Austria
关键词
MODULATED EXCITATION SIGNALS; VOLUMETRIC IMAGING-SYSTEM; MEDICAL ULTRASOUND; DESIGN;
D O I
10.1109/TUFFC.2010.1517
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Volumetric imaging can be performed using 1-D arrays in combination with mechanical motion. Outside the elevation focus of the array, the resolution and contrast quickly degrade compared with the lateral plane, because of the fixed transducer focus. This paper shows the feasibility of using synthetic aperture focusing for enhancing the elevation focus for a convex rocking array. The method uses a virtual source (VS) for defocused multi-element transmit, and another VS in the elevation focus point. This allows a direct time-of-flight to be calculated for a given 3-D point. To avoid artifacts and increase SNR at the elevation VS, a plane-wave VS approach has been implemented. Simulations and measurements using an experimental scanner with a convex rocking array show an average improvement in resolution of 26% and 33%, respectively. This improvement is also seen in in vivo measurements. An evaluation of how a change in transducer design will affect the resolution improvement shows a potential for using a modified transducer for 3-D imaging with improved elevation focusing and contrast.
引用
收藏
页码:1051 / 1063
页数:13
相关论文
共 28 条
[1]   Precise time-of-flight calculation for 3D synthetic aperture focusing [J].
Andresen, Henrik ;
Nikolov, Svetoslav Ivanov ;
Jensen, Jorgen Arendt .
2007 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1-6, 2007, :224-227
[2]   Precise Time-of-Flight Calculation For 3-D Synthetic Aperture Focusing [J].
Andresen, Henrik ;
Nikolov, Svetoslav Ivanov ;
Jensen, Jorgen Arendt .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2009, 56 (09) :1880-1887
[3]  
BENNETT S, 1982, ACOUST IMAGING, V10, P669
[4]  
CHEN JNC, 1994, Patent No. 5285789
[5]  
FDA, 1997, INF MAN SEEK MARK CL
[6]  
FINSTERWALD MP, 1996, Patent No. 9611753
[7]  
FLAHERTY JJ, 1970, Patent No. 3548642
[8]   Synthetic aperture techniques with a virtual source element [J].
Frazier, CH ;
O'Brien, WD .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1998, 45 (01) :196-207
[9]   Multielement synthetic transmit aperture imaging using temporal encoding [J].
Gammelmark, KL ;
Jensen, JA .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2003, 22 (04) :552-563
[10]  
Jensen J. A., 1996, Medical & Biological Engineering & Computing, V34, P351