In order to effectively remove the magnetocardiography (MCG) signal noise and improve the accuracy of heart disease diagnosis, a cardiac Fourier-wavelet denoising method based on sparse representation is used to process the cardiac magnetic signal. The noise reduction method combines the advantages of Fourier and wavelet. The initial denoising process is performed in Fourier domain, and then the residual noise is removed in the wavelet domain. Based on the theory of sparse representation, the wavelet denoising is transformed into an optimization problem, and a denoising model based on sparse representation is established. By eliminating the sparsity of wavelet coefficients, the purpose of noise removal is achieved. The experimental results verify that the method of the paper can effectively improve the denoising effect of the cardiac magnetic signal.
引用
收藏
页码:266 / 268
页数:3
相关论文
共 8 条
[1]  
[Anonymous], 2004, MODERN SIGNAL PROCES
[2]   Hybrid Fourier-wavelet image denoising [J].
Jiang, S. ;
Hao, X. .
ELECTRONICS LETTERS, 2007, 43 (20) :1081-1082
[3]   A THEORY FOR MULTIRESOLUTION SIGNAL DECOMPOSITION - THE WAVELET REPRESENTATION [J].
MALLAT, SG .
IEEE TRANSACTIONS ON PATTERN ANALYSIS AND MACHINE INTELLIGENCE, 1989, 11 (07) :674-693
[4]  
[唐锐 Tang Rui], 2013, [激光与红外, Laser and Infrared], V43, P592
[5]  
Wan Hong, 2008, MICROCOMPUTER INFORM, V24, P156
[6]  
Wu Juzhuo, 2018, INSTRUMENTATION
[7]  
Yali Qiao, THESIS BEIJING JIAOT
[8]   Applications of improved complementary pair adaptive noise cancellation to MCG analysis [J].
Zhao, L ;
Chen, GH ;
Zhang, LH ;
Huang, XG ;
Zhai, GJ ;
Li, JW ;
Tang, YL ;
Feng, J .
ACTA PHYSICA SINICA, 2004, 53 (12) :4420-4427