Effect of surface topography and residual stress on the taper connection stability in total hip arthroplasty

被引:4
作者
Mai, Phuong Thao [1 ]
Bormann, Therese [2 ]
Mueller, Ulrike [2 ]
Kretzer, Jan Philippe [2 ]
Gibmeier, Jens [1 ]
机构
[1] Karlsruhe Inst Technol, Inst Appl Mat, D-76131 Karlsruhe, Germany
[2] Heidelberg Univ Hosp, Lab Biomech & Implant Res, Clin Orthoped & Trauma Surg, D-69118 Heidelberg, Germany
关键词
Topography; Head-stem junction; Fretting fatigue; Joining process; Taper contact situation; Total hip arthroplasty; SIMULATED INFLAMMATORY CONDITIONS; HEAD; CORROSION; ALLOY; REPLACEMENT; TITANIUM; JUNCTION; WEAR; STRENGTH; STATES;
D O I
10.1016/j.jmbbm.2022.105119
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In the present work, the influence of the trunnion surface topography and the near-surface residual stresses on the joining process of a taper connection is examined using a replicate of the realistic taper connection as it occurs in conventional hip joint implants. The focus of the work is on the surface of the taper trunnion made of Ti6Al4V ELI and its effect on the connection stability with a CoCrMo counterpart. In this regard, the interrelation between surface topography, residual stresses, the joining behavior and the corrosion behavior under dynamic loading have been systematically investigated. For this purpose, taper trunnions produced by means of three different machining processes were considered, i.e. fine machining, rough machining and a novel furrowing process. These mechanical surface treatments result in different surface topographies and near-surface work hardening and residual stress states. The results show that the primary taper stability is hardly altered by the different types of trunnion surfaces. For all three surface states, the joining/dismantling procedure did not change the residual stress state at the surface. After corrosion testing under dynamic loading, the fine machined taper surface exhibits the highest stability. Moreover, fine machined tapers consolidated during the dynamic corrosion experiment as the ratio between joining and dismantling force increased from 0.49 +/- 0.04 to 0.83 +/- 0.08. For the furrowed and rough machined taper surfaces, the connection stability showed a tendency towards increase and decrease, respectively, in the course of dynamic corrosion testing. The results indicate that for choosing an optimal taper trunnion surface, the effects of corrosion must be taken into account.
引用
收藏
页数:11
相关论文
共 52 条
[1]  
[Anonymous], 2007, ERMUDUNGSFESTIGKEIT
[2]   Do Stem Taper Microgrooves Influence Taper Corrosion in Total Hip Arthroplasty? A Matched Cohort Retrieval Study [J].
Arnholt, Christina M. ;
MacDonald, Daniel W. ;
Underwood, Richard J. ;
Guyer, Eric P. ;
Rimnac, Clare M. ;
Kurtz, Steven M. ;
Mont, Michael A. ;
Klein, Gregg R. ;
Lee, Gwo-Chin ;
Chen, Antonia F. ;
Hamlin, Brian R. ;
Cates, Harold E. ;
Malkani, Arthur L. ;
Kraay, Matthew J. .
JOURNAL OF ARTHROPLASTY, 2017, 32 (04) :1363-1373
[3]   Atomic Scale Origin of Metal Ion Release from Hip Implant Taper Junctions [J].
Balachandran, Shanoob ;
Zachariah, Zita ;
Fischer, Alfons ;
Mayweg, David ;
Wimmer, Markus A. ;
Raabe, Dierk ;
Herbig, Michael .
ADVANCED SCIENCE, 2020, 7 (05)
[4]  
Bargel H-J., 2012, Werkstoffkunde, V11th
[5]  
Baxmanna M., 2017, Biotribology, V12, P1, DOI DOI 10.1016/J.BIOTRI.2017.07.002
[6]   Corrosion Behavior of Surface-Treated Metallic Implant Materials [J].
Bormann, Therese ;
Phuong Thao Mai ;
Gibmeier, Jens ;
Sonntag, Robert ;
Mueller, Ulrike ;
Kretzer, J. Philippe .
MATERIALS, 2020, 13 (09)
[7]   Shorter, rough trunnion surfaces are associated with higher taper wear rates than longer, smooth trunnion surfaces in a contemporary large head metal-on-metal total hip arthroplasty system [J].
Brock, Timothy M. ;
Sidaginamale, Raghavendra ;
Rushton, Steven ;
Nargol, Antoni V. F. ;
Bowsher, John G. ;
Savisaar, Christina ;
Joyce, Tom J. ;
Deehan, David J. ;
Lord, James K. ;
Langton, David J. .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2015, 33 (12) :1868-1874
[8]  
Committee,, PRACT FRETT CORR TES
[9]   Surface integrity in high speed end milling of titanium alloy Ti-6Al-4V [J].
Daymi, A. ;
Boujelbene, M. ;
Ben Amara, A. ;
Bayraktar, E. ;
Katundi, D. .
MATERIALS SCIENCE AND TECHNOLOGY, 2011, 27 (01) :387-394
[10]   The Influence of Head Size on Corrosion and Fretting Behaviour at the Head-Neck Interface of Artificial Hip Joints [J].
Dyrkacz, Richard M. R. ;
Brandt, Jan-M. ;
Ojo, Olanrewaju A. ;
Turgeon, Thomas R. ;
Wyss, Urs P. .
JOURNAL OF ARTHROPLASTY, 2013, 28 (06) :1036-1040