Comparison of monocular and binocular infrared pupillometers under mesopic lighting conditions

被引:12
作者
Scheffel, Magdalena [1 ]
Kuehne, Christoph [1 ]
Kohnen, Thomas [1 ]
机构
[1] Goethe Univ Frankfurt, Dept Ophthalmol, D-60590 Frankfurt, Germany
关键词
PHAKIC INTRAOCULAR-LENS; PUPIL SIZE; VISION; DIAMETER; IMPLANTATION; QUALITY; LASIK;
D O I
10.1016/j.jcrs.2009.11.006
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To compare 3 infrared pupillometers under mesopic lighting conditions in refractive surgery candidates with moderate to high myopia SETTING: Johann Wolfgang Goethe-University, Department of Ophthalmology, Frankfurt/Main. Germany. METHODS: In this prospective 1-day study, pupils of refractive surgery candidates with moderate to high myopia were measured with 3 infrared pupillometers. the monocular Colvard (monocular pupillometer A), the monocular Pupil Scan II (monocular pupillometer B), and the binocular Procyon P2000SA Mesopic lighting conditions were consistent with the International Organization for Standardization for phakic intraocular lenses and manufacturer instructions as follows: 3 cd/m(2) (both monocular pupillometers). 0 4 lux (binocular pupillometer mesopic low setting), and user defined (binocular pupillometer user setting). Pupil diameters were measured twice per pupillometer by 2 independent examiners. RESULTS: The study included 40 subjects (80 eyes) The mean pupil diameters were similar except with the mesopic low setting of the binocular pupillometer, which measured pupil diameter approximately 1 2 mm smaller. There was a high correlation between the 2 monocular pupillometers (0.83, P < 0001) The correlation between the monocular methods was stronger than that between the binocular user setting and either monocular method (monocular pupillometer A, r(2) = 0 665, P<.0001; monocular pupillometer B, r(2) = 0 594, P<0001) No statistically significant difference was found between monocular pupillometer A. monocular pupillometer B, and the binocular pupillometer user setting CONCLUSION: Both monocular pupillometers and the binocular pupillometer user setting measured pupil diameter under mesopic conditions similarly in eyes with moderate to high myopia.
引用
收藏
页码:625 / 630
页数:6
相关论文
共 19 条
[1]  
Baikoff G, 1998, J REFRACT SURG, V14, P282
[2]   Dark adaptation is critical for accurate pupil measurement [J].
Brown, Sandra M. ;
Bradley, Jay C. ;
Khanani, Arshad M. .
ARCHIVES OF OPHTHALMOLOGY, 2008, 126 (04) :584-584
[3]  
Campion M, 1998, J CATARACT REFR SURG, V24, P293, DOI 10.1016/S0886-3350(98)80309-X
[4]  
Charles KC, 1998, J CATARACT REFR SURG, V24, P294, DOI 10.1016/S0886-3350(98)80312-X
[5]   Change in pupil size after implantation of an iris-fixated toric phakic intraocular lens [J].
Dick, HB ;
Aliyeva, S ;
Tehrani, M .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (02) :302-307
[6]  
International Organization for Standardization, 2006, 1197910 ISO
[7]   Pupil diameter changes and reaction after posterior chamber phakic intraocular lens implantation [J].
Keuch, RJ ;
Bleckmann, H .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (12) :2170-2172
[8]   Night vision after LASIK: The pupil proclaims innocence [J].
Klyce, SD .
OPHTHALMOLOGY, 2004, 111 (01) :1-2
[9]   Comparison of a digital and a handheld infrared pupillometer for determining scotopic pupil diameter [J].
Kohnen, T ;
Terzi, E ;
Buhren, J ;
Kohnen, EM .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2003, 29 (01) :112-117
[10]   Monocular versus binocular pupillometry [J].
Kurz, S ;
Krummenauer, F ;
Pfeiffer, N ;
Dick, HB .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2004, 30 (12) :2551-2556