Molecularly imprinted solid-phase extraction combined with molecularly imprinted polymer-sensor: a diagnostic tool applicable to creatine deficiency syndrome

被引:12
作者
Sharma, Piyush Sindhu [1 ]
Lakshmi, Dhana [1 ]
Prasad, Bhim Bali [1 ]
机构
[1] Banaras Hindu Univ, Dept Chem, Analyt Div, Varanasi 221005, Uttar Pradesh, India
关键词
molecularly imprinted polymer; creatine; molecularly imprinted polymer-modified HMDE sensor; molecularly imprinted solid-phase extraction; blood serum;
D O I
10.1002/bmc.842
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Primary creatine deficiency syndromes (CDS) are a new group of disorders caused by guanidinoacetate methyltransferase (GAMT) deficiency, which affects endogenous creatine biosynthesis with depletion of body creatine. A deficiency in creatine can be corrected by treatment with oral creatine supplementation and this necessitates a simple and sensitive screening method for early detection of creatine in dilute physiologic fluids. In this work an artificial receptor, molecularly imprinted polymer (MIP), for creatine was used both as a material for solid-phase extraction (SPE) and as a sensing element in a voltammetric sensor. Using the combination of molecularly imprinted solid-phase extraction (MISPE) with a complementary MIP sensor, the minimum detectable amount was found to be 0.0015 ng mL(-1) (RSD = 1.3%, S/N = 3). The MISPE-MIP sensor combination provided up to 60-fold preconcentration, which was more than sufficient for achieving the required quantification limit 50 ng mL(-1) (or 0.0025 ng mL(-1) after 2 x 10(4)-fold dilution) for creatine in human blood serum. Copyright 0 2007 John Wiley & Sons, Ltd.
引用
收藏
页码:976 / 986
页数:11
相关论文
共 46 条
[11]  
HERBERT K, 1974, CLIN CHIM ACTA, V53, P209
[12]   GC-MS determination of guanidinoacetate in urine and plasma [J].
Hunneman, DH ;
Hanefeld, F .
JOURNAL OF INHERITED METABOLIC DISEASE, 1997, 20 (03) :450-452
[13]   Liquid chromatography of guanidino compounds using a porous graphite carbon column and application to their analysis in serum [J].
Inamoto, Y ;
Inamoto, S ;
Hanai, T ;
Tokuda, M ;
Hatase, O ;
Yoshii, K ;
Sugiyama, N ;
Kinoshita, T .
JOURNAL OF CHROMATOGRAPHY B, 1998, 707 (1-2) :111-120
[14]   An enzymatic assay for erythrocyte creatine as an index of the erythrocyte life time [J].
Jiao, YF ;
Okumiya, T ;
Saibara, T ;
Tsubosaki, E ;
Matsumura, H ;
Park, K ;
Sugimoto, K ;
Kageoka, T ;
Sasaki, M .
CLINICAL BIOCHEMISTRY, 1998, 31 (02) :59-65
[15]   HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY OF GUANIDINO COMPOUNDS USING BENZOIN AS A PRECOLUMN FLUORESCENT DERIVATIZATION REAGENT [J].
KAI, M ;
MIYAZAKI, T ;
YAMAGUCHI, M ;
OHKURA, Y .
JOURNAL OF CHROMATOGRAPHY, 1983, 268 (03) :417-424
[16]   HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC MEASUREMENT OF GUANIDINO COMPOUNDS OF CLINICAL IMPORTANCE IN HUMAN-URINE AND SERUM BY PRE-COLUMN FLUORESCENCE DERIVATIZATION USING BENZOIN [J].
KAI, M ;
MIYAZAKI, T ;
OHKURA, Y .
JOURNAL OF CHROMATOGRAPHY, 1984, 311 (02) :257-266
[17]  
KORALL H, 2000, J INHERIT METAB DIS, V23, P213
[18]  
LAKSHMI D, 2006, BIOSENS BIOELECTRON, DOI DOI 10.1016/JBIOS2006.12.011
[19]   Analysis of creatine, creatinine, creatine-d3 and creatinine-d3 in urine, plasma, and red blood cells by HPLC and GC-MS to follow the fate of ingested creatine-d3 [J].
MacNeil, L ;
Hill, L ;
MacDonald, D ;
Keefe, L ;
Cormier, JF ;
Burke, DG ;
Smith-Palmer, T .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2005, 827 (02) :210-215
[20]  
Marymont J H Jr, 1968, Tech Bull Regist Med Technol, V38, P15