CT dose reduction using Automatic Exposure Control and iterative reconstruction: A chest paediatric phantoms study

被引:30
作者
Greffier, Jol [1 ]
Pereira, Fabricio [1 ]
Macri, Francesco [1 ]
Beregi, Jean-Paul [1 ]
Larbi, Ahmed [1 ]
机构
[1] Univ Nimes Hosp, Dept Radiol, Med Imaging Grp Nimes, EA 2415, Bd Prof Robert Debre, F-30029 Nimes, France
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2016年 / 32卷 / 04期
关键词
MDCT; Automatic Exposure Control; Radiation dose; Children; Iterative reconstruction; TUBE CURRENT MODULATION; COMPUTED-TOMOGRAPHY; MULTIDETECTOR CT; IMAGE QUALITY; OPTIMIZATION; CHILDREN; RISK;
D O I
10.1016/j.ejmp.2016.03.007
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To evaluate the impact of Automatic Exposure Control (AEC) on radiation dose and image quality in paediatric chest scans (MDCT), with or without iterative reconstruction (IR). Methods: Three anthropomorphic phantoms representing children aged one, five and 10-year-old were explored using AEC system (CARE Dose 4D) with five modulation strength options. For each phantom, six acquisitions were carried out: one with fixed mAs (without AEC) and five each with different modulation strength. Raw data were reconstructed with Filtered Back Projection (FBP) and with two distinct levels of IR using soft and strong kernels. Dose reduction and image quality indices (Noise, SNR, CNR) were measured in lung and soft tissues. Noise Power Spectrum (NPS) was evaluated with a Catphan 600 phantom. Results: The use of AEC produced a significant dose reduction (p < 0.01) for all anthropomorphic sizes employed. According to the modulation strength applied, dose delivered was reduced from 43% to 91%. This pattern led to significantly increased noise (p < 0.01) and reduced SNR and CNR (p < 0.01). However, IR was able to improve these indices. The use of AEC/IR preserved image quality indices with a lower dose delivered. Doses were reduced from 39% to 58% for the one-year-old phantom, from 46% to 63% for the five-year-old phantom, and from 58% to 74% for the 10-year-old phantom. In addition, AEC/IR changed the patterns of NPS curves in amplitude and in spatial frequency. Conclusions: In chest paediatric MDCT, the use of AEC with IR allows one to obtain a significant dose reduction while maintaining constant image quality indices. (C) 2016 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:582 / 589
页数:8
相关论文
共 33 条
  • [1] [Anonymous], 2001, Health Physics, V80, P291
  • [2] Iterative reconstruction methods in X-ray CT
    Beister, Marcel
    Kolditz, Daniel
    Kalender, Willi A.
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2012, 28 (02): : 94 - 108
  • [3] Boone J.M., 2011, SIZE SPECIFIC DOSE E
  • [4] Current concepts - Computed tomography - An increasing source of radiation exposure
    Brenner, David J.
    Hall, Eric J.
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (22) : 2277 - 2284
  • [5] Estimated risks of radiation-induced fatal cancer from pediatric CT
    Brenner, DJ
    Elliston, CD
    Hall, EJ
    Berdon, WE
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 176 (02) : 289 - 296
  • [6] CT exposure from pediatric MDCT: results from the 2007-2008 SFIPP/ISRN survey
    Brisse, H. J.
    Aubert, B.
    [J]. JOURNAL DE RADIOLOGIE, 2009, 90 (02): : 207 - 215
  • [7] Radiation risk to children from computed tomography
    Brody, Alan S.
    Frush, Donald P.
    Huda, Walter
    Brent, Robert L.
    [J]. PEDIATRICS, 2007, 120 (03) : 677 - 682
  • [8] Buls N., 2010, CT paediatric doses in Belgium: a multi-centre study: Results from a dosimetry audit in 2007-2009. s.l
  • [9] Reducing radiation dose associated with pediatric CT by decreasing unnecessary examinations
    Donnelly, LF
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2005, 184 (02) : 655 - 657
  • [10] [FRANCAISE JODLR MINISTERE DU TRAVAIL DLEEDLS], DECR ARR CIRC, P1