Multicenter evaluation of the Paragon CZE™ 2000 capillary zone electrophoresis system for serum protein electrophoresis and monoclonal component typing
被引:0
|作者:
Bienvenu, J
Graziani, MS
Arpin, F
Bernon, H
Blessum, C
Marchetti, C
Righetti, G
Somenzini, M
Verga, G
Aguzzi, F
[1
]
机构:
[1] Osped Stradella, Lab Anal, I-27049 Stradella, Italy
[2] Ctr Hosp Lyon Sud, Immunol Lab, F-69495 Pierre Benite, France
[3] Osped Civile Maggiore, Chim Clin & Ematol Lab, I-37126 Verona, Italy
[4] Beckman Instruments Inc, Brea, CA 92621 USA
[5] Beckman Analyt SPA, I-20060 Cassina De Pecchi, Italy
关键词:
D O I:
暂无
中图分类号:
R446 [实验室诊断];
R-33 [实验医学、医学实验];
学科分类号:
1001 ;
摘要:
Serum protein electrophoresis and typing of monoclonal components (MCs) are routine but time-consuming and technically demanding assays. We evaluated capillary electrophoresis (Paragon CZE(TM) 2000) for automation of the two assays. CZE and cellulose acetate electrophoresis gave similar data on 794 samples. Within-run and between-run CVs were <2% for albumin and gamma-globulins and 4-7% for alpha(1)-, alpha(2)-, and beta-globulins. Bilirubin, hemoglobin, triglycerides, and fibrinogen were found not to interfere. No carryover by capillaries was detected. The detection limit for MC was <0.5 g/L. MC assessment by immunosubtraction on 403 samples identified the monoclonal type in all samples with peak concentrations >10 g/L; only 50% of MCs that could not be quantified by densitometric scan were typed.