New concept of traction force applied to biliary self-expandable metallic stents

被引:9
|
作者
Hori, Yasuki [1 ]
Hayashi, Kazuki [1 ]
Yoshida, Michihiro [1 ]
Naitoh, Itaru [1 ]
Nakazawa, Takahiro [1 ]
Miyabe, Katsuyuki [1 ]
Shimizu, Shuya [1 ]
Kondo, Hiromu [1 ]
Nishi, Yuji [1 ]
Umemura, Shuichiro [1 ]
Kato, Akihisa [1 ]
Ohara, Hirotaka [2 ]
Joh, Takashi [1 ]
机构
[1] Nagoya City Univ, Grad Sch Med Sci, Dept Gastroenterol & Metab, Nagoya, Aichi, Japan
[2] Nagoya City Univ, Dept Community Based Med Educ, Grad Sch Med Sci, Nagoya, Aichi, Japan
关键词
RANDOMIZED-TRIAL; MECHANICAL-PROPERTIES; EUROPEAN-SOCIETY; OBSTRUCTION; FEASIBILITY; CHOICE;
D O I
10.1055/s-0041-111566
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
Background and study aims: Various mechanical properties of self-expandable metallic stents (SEMSs) have been reported. They can be classified into the device behavior during and after deployment. While there have been several reports on the latter, information on the former is insufficient. During deployment, the position is maintained by retracting the delivery catheter. We propose that this pulling force be called the traction force and that the magnitude of traction force is termed the traction momentum. The aim of this study was to measure these parameters in order to clarify the properties of SEMSs in terms of their deployment. Material and methods: The traction force, traction momentum, and shortening rate of 10 different SEMSs were measured. Traction force was measured using in-house equipment, and the shortening rate was determined by measuring the stents. Results: The shortening rate was closely related to the stent structure. The traction force varied between 1.5 N and 9.4 N, and the traction momentum was significantly elevated in covered and braided stents. A high traction force did not imply a high traction momentum. Conclusions: A low or constant traction force and a minimal shortening rate significantly facilitated SEMS deployment to optimal positions. Traction force could be an important element for new ideal SEMS design.
引用
收藏
页码:472 / 476
页数:5
相关论文
共 50 条
  • [21] Efficacy and Safety of Fully Covered Self-Expandable Metallic Stents in Biliary Complications After Liver Transplantation: A Preliminary Study
    Traina, Mario
    Tarantino, Ilaria
    Barresi, Luca
    Volpes, Riccardo
    Gruttadauria, Salvatore
    Petridis, Ioannis
    Gridelli, Bruno
    LIVER TRANSPLANTATION, 2009, 15 (11) : 1493 - 1498
  • [22] Self-expandable Metallic Stents in Patients with Stage IV Obstructing Colorectal Cancer
    Lamazza, Antonietta
    Fiori, Enrico
    Schillaci, Alberto
    DeMasi, Ercole
    Pontone, Stefano
    Sterpetti, Antonio V.
    WORLD JOURNAL OF SURGERY, 2012, 36 (12) : 2931 - 2936
  • [23] Self-expandable metal stents versus plastic stents for malignant biliary obstruction: a meta-analysis
    Sawas, Tarek
    Al Halabi, Shadi
    Parsi, Mansour A.
    Vargo, John J.
    GASTROINTESTINAL ENDOSCOPY, 2015, 82 (02) : 256 - +
  • [24] No Benefit of Endoscopic Sphincterotomy Before Biliary Placement of Self-Expandable Metal Stents for Unresectable Pancreatic Cancer
    Hayashi, Tsuyoshi
    Kawakami, Hiroshi
    Osanai, Manabu
    Ishiwatari, Hirotoshi
    Naruse, Hirohito
    Hisai, Hiroyuki
    Yanagawa, Nobuyuki
    Kaneto, Hiroyuki
    Koizumi, Kazuya
    Sakurai, Tamaki
    Sonoda, Tomoko
    CLINICAL GASTROENTEROLOGY AND HEPATOLOGY, 2015, 13 (06) : 1151 - +
  • [25] Evaluation of anti-migration properties of biliary covered self-expandable metal stents
    Minaga, Kosuke
    Kitano, Masayuki
    Imai, Hajime
    Harwani, Yogesh
    Yamao, Kentaro
    Kamata, Ken
    Miyata, Takeshi
    Omoto, Shunsuke
    Kadosaka, Kumpei
    Sakurai, Toshiharu
    Nishida, Naoshi
    Kudo, Masatoshi
    WORLD JOURNAL OF GASTROENTEROLOGY, 2016, 22 (30) : 6917 - 6924
  • [26] Evaluation of the mechanical properties of current biliary self- expandable metallic stents: axial and radial force, and axial force zero border
    Yamagata, Wataru
    Fujisawa, Toshio
    Sasaki, Takashi
    Ishibashi, Rei
    Saito, Tomotaka
    Yoshida, Shuntaro
    No, Shizuka
    Inoue, Kouta
    Nakai, Yousuke
    Sasahira, Naoki
    Isayama, Hiroyuki
    CLINICAL ENDOSCOPY, 2023, 56 (05) : 633 - 649
  • [27] Sequential or Simultaneous Placement of Self-expandable Metallic Stents for Palliation of Malignant Biliary and Duodenal Obstruction Due to Unresectable Pancreatic Head Carcinoma
    Katsinelos, Panagiotis
    Kountouras, Jannis
    Germanidis, George
    Paroutoglou, George
    Paikos, Dimitris
    Lazaraki, Georgia
    Pilpilidis, Ioannis
    Chatzimavroudis, Grigoris
    Fasoulas, Kostas
    Zavos, Christos
    SURGICAL LAPAROSCOPY ENDOSCOPY & PERCUTANEOUS TECHNIQUES, 2010, 20 (06) : 410 - 415
  • [28] A tissue attached to self-expandable metal stents for biliary stricture could be useful to find malignancy
    Kwon, Hyungjun
    Jung, Min Kyu
    Park, Sung Jin
    Heo, Jun
    Lee, Dong Wook
    Kang, Yoo Na
    Han, Jaryung
    Han, Young Seok
    Lee, Won Kee
    SCIENTIFIC REPORTS, 2022, 12 (01)
  • [29] Comparison of plastic stents with self-expandable metal stents in palliative treatment of malignant biliary obstruction: a meta-analysis
    Yuan, T. -W.
    Liu, H. -Q.
    Wang, S. -B.
    Cao, J.
    EUROPEAN REVIEW FOR MEDICAL AND PHARMACOLOGICAL SCIENCES, 2017, 21 (12) : 2847 - 2857
  • [30] Outcome of self-expandable metallic stents in low-grade versus advanced hilar obstruction
    Rerknimitr, Rungsun
    Kongkam, Pradermchai
    Kullavanijaya, Pinit
    JOURNAL OF GASTROENTEROLOGY AND HEPATOLOGY, 2008, 23 (11) : 1695 - 1701