Gliclazide voltammetric sensor based on electropolymerized molecularly imprinted polypyrrole film onto glassy carbon electrode

被引:33
作者
Hrichi, Hajer [1 ]
Louhaichi, Mohamed Radhouan [2 ]
Monser, Lotfi [1 ]
Adhoum, Nafaa [1 ,3 ]
机构
[1] Carthage Univ, Natl Inst Appl Sci & Technol, Ctr Urbain Nord, Tunis 1080, Tunisia
[2] Natl Lab Pharmaceut Control, Tunis, Tunisia
[3] Kairouan Univ, Preparatory Sch Engn Studies, Kairouan, Tunisia
关键词
Molecularly imprinted polymer; Polypyrrole; Glassy carbon electrode; Electropolymerization; Gliclazide; PERFORMANCE LIQUID-CHROMATOGRAPHY; ELECTROCHEMICAL SENSORS; DOSAGE FORMS; GLIPIZIDE; OXIDATION; METFORMIN; POLYMERS; STRATEGY; COLUMN; ACID;
D O I
10.1016/j.snb.2014.07.056
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Gliclazide sensing was investigated based on differential pulse voltammetry measurements of an electropolymerized molecularly imprinted polymer (E-MIP) film. The E-MIP polymer was prepared via anodic electropolymerization of pyrrole in the presence of GLZ onto glassy carbon electrodes using cyclic voltammetry (CV). GLZ molecules are successfully trapped into the polypyrrole matrix creating, after their subsequent removal, shape-complementary artificial recognition sites. The effect of several significant operational parameters (monomer and template concentrations, number of CV scans, pH and incubation time) on film analytical performances were investigated and optimized. Under optimized conditions, the sensor response exhibited high sensitivity toward the target template and was linearly proportional to GLZ concentration (R-2 = 0.998) over the range 5 x 10(-11)4 x 10(-10) M, with a detection limit (3 sigma/m) of 1.2 x 10(-11) M. The precision of the method (R.S.D., n = 6) for within and between-days is better than 1.4% and 2.48%, respectively at 10(-10) M. Moreover, the selectivity of E-MIP sensor, against potentially competing molecules (Imp. B, Imp. E, glipizide, glibenclamide, glimepiride), was demonstrated. The developed E-MIP sensor was successfully applied to the determination of GLZ in three pharmaceutical products and gave results in close agreement with the reference HPLC method with mean recoveries between 95.4 and 98.8%, showing promising potential in practical applications. (C) 2014 Elsevier B.V. All rights reserved.
引用
收藏
页码:42 / 49
页数:8
相关论文
共 41 条
[1]  
Abd-Elgawad R., 2010, ELECTROANAL, V22, P2991
[2]   The development and validation of liquid chromatography method for the simultaneous determination of metformin and glipizide, gliclazide, glibenclamide or glimperide in plasma [J].
AbuRuz, S ;
Millership, J ;
McElnay, J .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2005, 817 (02) :277-286
[3]   Controlling of systemic absorption of gliclazide through incorporation into alginate beads [J].
Al-Kassas, Raida S. ;
Al-Gohary, Omaimah M. N. ;
Al-Faadhel, Monirah M. .
INTERNATIONAL JOURNAL OF PHARMACEUTICS, 2007, 341 (1-2) :230-237
[4]   A capacitive biosensor for ultra-trace level urea determination based on nano-sized urea-imprinted polymer receptors coated on graphite electrode surface [J].
Alizadeh, Taher ;
Akbari, Aezam .
BIOSENSORS & BIOELECTRONICS, 2013, 43 :321-327
[5]   Electropolymerized Molecularly Imprinted Polymer Film: EIS Sensing of Bisphenol A [J].
Apodaca, Dahlia C. ;
Pernites, Roderick B. ;
Ponnapati, Ramakrishna ;
Del Mundo, Florian R. ;
Advincula, Rigoberto C. .
MACROMOLECULES, 2011, 44 (17) :6669-6682
[6]   Forced degradation study on gliclazide and application of validated stability-indicating HPLC-UV method in stability testing of gliclazide tablets [J].
Bansal, Gulshan ;
Singh, Manjeet ;
Jindal, Kaur Chand .
CHROMATOGRAPHIA, 2007, 66 (9-10) :751-755
[7]   Cogrinding as an approach to enhance dissolution rate of a poorly water-soluble drug (gliclazide) [J].
Barzegar-Jalali, Mohammad ;
Valizadeh, Hadi ;
Shadbad, Mohammad-Reza Siahi ;
Adibkia, Khosro ;
Mohammadi, Ghobad ;
Farahani, Amin ;
Arash, Zeynab ;
Nokhodchi, Ali .
POWDER TECHNOLOGY, 2010, 197 (03) :150-158
[8]   Electrochemical sensors based on molecularly imprinted polymers [J].
Blanco-López, MC ;
Lobo-Castañón, MJ ;
Miranda-Ordieres, AJ ;
Tuñón-Blanco, P .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2004, 23 (01) :36-48
[9]   Voltammetric response of diclofenac-molecularly imprinted film modified carbon electrodes [J].
Blanco-López, MC ;
Lobo-Castañón, MJ ;
Miranda-Ordieres, AJ ;
Tuñón-Blanco, P .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2003, 377 (02) :257-261
[10]   Effects of gliclazide on endothelial function in patients with newly diagnosed type 2 diabetes [J].
Chen, Lu-Lu ;
Yu, Fan ;
Zeng, Tian-shu ;
Liao, Yun-fei ;
Li, Yu-ming ;
Ding, Hong-cheng .
EUROPEAN JOURNAL OF PHARMACOLOGY, 2011, 659 (2-3) :296-301