A retrospective analysis of discordances between international normalized ratio (INR) self-testing and INR laboratory testing in a pediatric patient population

被引:0
作者
Bitan, Joan [1 ]
Bajolle, Fanny [2 ]
Harroche, Annie [3 ,4 ]
Cannet, Pauline [2 ]
Braems, Alice [2 ]
Taleb, Sofiane [1 ]
Bonnet, Damien [2 ,3 ]
Borgel, Delphine [1 ,5 ]
Lasne, Dominique [1 ,5 ]
机构
[1] Hop Univ Necker Enfants Malad, AP HP, Hematol Lab, Paris, France
[2] Hop Univ Necker Enfants Malad, AP HP, M3C Necker Cardiol Congenitale & Pediat, Paris, France
[3] Univ Paris, Paris, France
[4] Hop Univ Necker Enfants Malad, AP HP, Ctr Traitement Hemophilie, Paris, France
[5] Univ Paris Saclay, INSERM, UMR S 1176, HITh, Le Kremlin Bicetre, France
关键词
Anticoagulants; international normalized ratio; point-of-care systems; quality control; thromboplastin; ORAL ANTICOAGULANT TREATMENT; VITAMIN-K ANTAGONISTS; QUALITY ASSESSMENT; PROTHROMBIN TIME; COAGUCHEK XS; MANAGEMENT; POINT; CHILDREN; THERAPY; PROGRAM;
D O I
10.1111/ijlh.13652
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction The lack of quality control procedures for home point-of-care (POC) international normalized ratio (INR) devices is a concern. Concomitant laboratory and POC INR testing may be proposed to overcome the lack of quality control. However, a difference between the POC INR and the laboratory INR is not necessarily due to failure of the POC device. This study aimed to identify variables associated with a significant deviation between the POC INR and the laboratory INR. Methods Children included in this retrospective cohort study performed at least one concomitant laboratory and POC INRs. Clinical and laboratory variables were assessed for an association with significant deviation within pairs of INR. Results A significant deviation was noted for 30 (15.3%) of the 196 pairs of INR measurements from 124 children. Relative to patients without deviations, patients with deviations were younger (odds ratio =0.91; P = .020), less experienced in the use of POC INR devices (odds ratio =0.89; P = .098), and more likely to have received an INR result from a laboratory using animal thromboplastin (odds ratio =2.81 vs. 0.37 for laboratories using human thromboplastin; P = .016). In a multivariate analysis, younger age and the laboratory's use of animal thromboplastin were associated with significant deviations. Conclusions Although most children had coherent pairs of INR values, the occurrence of deviations raises the question of the origin of the thromboplastin used in the laboratory and emphasizes the need to provide specific quality control procedures for POC INR devices.
引用
收藏
页码:1575 / 1584
页数:10
相关论文
共 37 条
  • [1] Quality of life in children participating in a non-selective INR self-monitoring VKA-education programme
    Amedro, Pascal
    Bajolle, Fanny
    Bertet, Helena
    Cheurfi, Radhia
    Lasne, Dominique
    Nogue, Erika
    Auquier, Pascal
    Picot, Marie-Christine
    Bonnet, Damien
    [J]. ARCHIVES OF CARDIOVASCULAR DISEASES, 2018, 111 (03) : 180 - 188
  • [2] [Anonymous], ARRETE 28 JUILLET 20
  • [3] [Anonymous], ARRETE 18 JUIN 2008
  • [4] [Anonymous], 175932007 ISO
  • [5] Pharmacology and management of the vitamin K antagonists
    Ansell, Jack
    Hirsh, Jack
    Hylek, Elaine
    Jacobson, Alan
    Crowther, Mark
    Palareti, Gualtiero
    [J]. CHEST, 2008, 133 (06) : 160S - 198S
  • [6] Home point-of-care international normalised ratio monitoring sustained by a non-selective educational program in children
    Bajolle, Fanny
    Lasne, Dominique
    Elie, Caroline
    Cheurfi, Radhia
    Grazioli, Aurelie
    Traore, Maladon
    Souillard, Patrick
    Boudjemline, Younes
    Jourdain, Patrick
    Bonnet, Damien
    [J]. THROMBOSIS AND HAEMOSTASIS, 2012, 108 (04) : 710 - 718
  • [7] EMPoWARed: Edmonton pediatric warfarin self-management study
    Bauman, M. E.
    Massicotte, M. P.
    Kuhle, S.
    Siddons, S.
    Bruce, A. A. K.
    [J]. THROMBOSIS RESEARCH, 2015, 136 (05) : 887 - 893
  • [8] Monitoring of treatment with vitamin K antagonists: recombinant thromboplastins are more sensitive to factor VII than tissue-extract thromboplastins
    Biedermann, J. S.
    van den Besselaar, A. M. H. P.
    de Maat, M. P. M.
    Leebeek, F. W. G.
    Kruip, M. J. H. A.
    [J]. JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2017, 15 (03) : 500 - 506
  • [9] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [10] Long-term follow-up of the St Jude medical prosthesis in pediatric patients
    Cabalka, AK
    Emery, RW
    Petersen, RJ
    Helseth, HK
    Jakkula, M
    Arom, KV
    Nicoloff, DM
    [J]. ANNALS OF THORACIC SURGERY, 1995, 60 (06) : S618 - S623