Transthoracic defibrillation potential gradients in a closed chest porcine model of prolonged spontaneous and electrically induced ventricular fibrillation
被引:6
|作者:
Niemann, James T.
[1
,2
]
Rosborough, John P.
[1
]
Youngquist, Scott T.
[3
]
Shah, Atman P.
[4
]
机构:
[1] Harbor UCLA Med Ctr, Dept Emergency Med, Torrance, CA 90509 USA
[2] Univ Calif Los Angeles, David Geffen Sch Med, Los Angeles, CA 90095 USA
[3] Univ Utah, Dept Surg, Div Emergency Med, Salt Lake City, UT USA
[4] Univ Chicago, Dept Med, Div Cardiol, Chicago, IL 60637 USA
基金:
美国国家卫生研究院;
关键词:
Ischaemia;
Cardiac arrest;
Ventricular fibrillation;
Defibrillation;
CARDIAC-ARREST;
EXTERNAL DEFIBRILLATION;
ENERGY-REQUIREMENTS;
ACUTE-ISCHEMIA;
RESUSCITATION;
D O I:
10.1016/j.resuscitation.2009.12.027
中图分类号:
R4 [临床医学];
学科分类号:
1002 ;
100602 ;
摘要:
Objective: The purpose of this study was to measure the local electrical field or potential gradient, measured with a catheter-based system, required to terminate long duration electrically or ischaemically induced ventricular fibrillation (VF). We hypothesized that prolonged ischaemic VF would be more difficult to terminate when compared to electrically induced VF of similar duration. Methods: Thirty anesthetized and instrumented swine were randomized to electrically induced VF or spontaneous, ischaemically induced VF, produced by balloon occlusion of the left anterior descending coronary artery. After 7 min of VF, chest compressions were initiated and rescue shocks were attempted 1 min later. The potential gradient for each shock was measured and the mean values required for defibrillation compared for the VF groups. Results: The number of shocks and the shock strength required for termination of VF were not significantly different for the groups. The potential gradient of the first successful defibrillating shock was significantly greater in the spontaneous, occlusion-induced VF group (12.80 +/- 2.82 V/cm vs 9.60 +/- 2.48 V/cm, p = 0.002). The number of refibrillations was greater in the ischaemic group than in the non-ischaemic electrical group (6 +/- 4 vs 1 1, p < 0.001). The number of animals requiring a shock at 360 J was 2.5 times greater for the ischaemic group. Conclusions: Defibrillation of prolonged VF produced by acute myocardial ischaemia requires a significantly greater potential gradient to terminate than prolonged VF induced by electrical stimulation of the right ventricular endocardium. The VF duration used in this study approximates that occurring in victims of out-of-hospital cardiac arrest. Our findings may be of clinical importance in the management of such patients. (C) 2010 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:477 / 480
页数:4