Agreement and repeatability of objective systems for assessment of the tear film

被引:26
作者
Fernandez, Joaquin [1 ,2 ]
Rodriguez-Vallejo, Manuel [1 ]
Martinez, Javier [1 ]
Tauste, Ana [1 ]
Garcia-Montesinos, Javier [1 ]
Pinero, David P. [3 ,4 ]
机构
[1] Vithas Virgen Mar Hosp, Dept Ophthalmol Qvis, Almeria 04120, Spain
[2] Torrecardenas Hosp Complex, Dept Ophthalmol, Almeria 04009, Spain
[3] Univ Alicante, Dept Opt Pharmacol & Anat, Alicante, Spain
[4] Vithas Medimar Int Hosp, Dept Ophthalmol OFTALMAR, Alicante, Spain
关键词
Tear film stability; Objective systems; Ocular scatter index; Non-invasive break-up time; Laser refractive surgery; DRY EYE; REFRACTIVE SURGERY; DISEASE; RELIABILITY; CANDIDATES; STABILITY; MENISCI;
D O I
10.1007/s00417-018-3986-9
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
To assess the agreement and repeatability of two objective systems for measuring the tear film stability. Retrospective analysis of the tear film stability of 99 healthy right eyes measured with a videokeratoscope (VK) and the Optical Quality Analysis System (OQAS, Visiometrics). Two consecutive measures were taken with both systems, with an interval of 10 min between them. Variables included in the study were first and mean non-invasive break-up times (NIBUT and MNIBUT) measured with VK, and mean and standard deviation of the optical scattering index (OSIm and OSIsd) measured with OQAS. The agreement and repeatability of grading scales provided by both devices were also evaluated using the Cohen's k with quadratic weights. The Ocular Surface Disease index (OSDI) questionnaire was also passed out to all subjects. Correlations and associations between subjective and objective metrics were analyzed. Significant differences were found between consecutive measurements of NIBUT (p = 0.04) and MNIBUT (p = 0.01), but not for OSIm (p = 0.11) and OSIsd (p = 0.50). Grading scales resulted in fair (k = 0.20) or poor agreement (k = 0.04) between systems depending if the first or second trial was considered. The repeatability of the grading scale was good for OQAS (k = 0.59) and fair for VK (k = 0.37). No significant correlations or associations were found between OSDI and any of the metrics obtained with both devices (p ae<yen> 0.36). The two devices evaluated cannot be used interchangeably for the assessment of tear film stability. Good intrasession repeatability was obtained for tear film grading of the OQAS whereas it was fair for VK.
引用
收藏
页码:1535 / 1541
页数:7
相关论文
共 29 条
[1]   Noninvasive Imaging of Tear Film Dynamics in Eyes With Ocular Surface Disease [J].
Abdelfattah, Nizar Saleh ;
Dastiridou, Anna ;
Sadda, SriniVas R. ;
Lee, Olivia L. .
CORNEA, 2015, 34 :S48-S52
[2]   An Objective Scatter Index Based on Double-Pass Retinal Images of a Point Source to Classify Cataracts [J].
Artal, Pablo ;
Benito, Antonio ;
Perez, Guillermo M. ;
Alcon, Encarna ;
De Casas, Alvaro ;
Pujol, Jaume ;
Marin, Jose M. .
PLOS ONE, 2011, 6 (02)
[3]   Clinical evaluation of the Oculus Keratograph [J].
Best, N. ;
Drury, L. ;
Wolffsohn, J. S. .
CONTACT LENS & ANTERIOR EYE, 2012, 35 (04) :171-174
[4]   Non-invasive assessment of tear film stability with a novel corneal topographer in Indian subjects [J].
Bhandari, Vipul ;
Reddy, Jagdeesh K. ;
Relekar, Kirti ;
Ingawale, Ameya ;
Shah, Neeraj .
INTERNATIONAL OPHTHALMOLOGY, 2016, 36 (06) :781-790
[5]  
Bland JM, 1999, STAT METHODS MED RES, V8, P135, DOI 10.1177/096228029900800204
[6]   Reduced ocular surface disease index (OSDI) scores in patients with isotretinoin treatment [J].
Caglar, Cagatay ;
Senel, Engin ;
Sabancilar, Emine ;
Durmus, Mustafa .
INTERNATIONAL OPHTHALMOLOGY, 2017, 37 (01) :197-202
[7]   Agreement between Automated and Traditional Measures of Tear Film Breakup [J].
Cox, Stephanie M. ;
Nichols, Kelly K. ;
Nichols, Jason J. .
OPTOMETRY AND VISION SCIENCE, 2015, 92 (09) :E257-E263
[8]   Tear Function Evaluation in Candidates of Corneal Laser Refractive Surgery for Myopia [J].
Farahi, Azadeh ;
Hashemi, Hassan ;
Mehravaran, Shiva ;
Tavakolizadeh, Sepideh ;
Khabazkhoob, Mehdi .
EYE & CONTACT LENS-SCIENCE AND CLINICAL PRACTICE, 2014, 40 (02) :91-94
[9]  
Hong J, 2014, OPTOMETRY VISION SCI, V91, P1
[10]   Assessment of Tear Film Stability in Dry Eye With a Newly Developed Keratograph [J].
Hong, Jiaxu ;
Sun, Xinghuai ;
Wei, Anji ;
Cui, Xinhan ;
Li, Yimin ;
Qian, Tingting ;
Wang, Wentao ;
Xu, Jianjiang .
CORNEA, 2013, 32 (05) :716-721