Paraplegic standing supported by FES-controlled ankle stiffness

被引:33
作者
Jaime, RP
Matjacic, Z
Hunt, KJ
机构
[1] Univ Glasgow, Ctr Syst & Control, Glasgow G12 8QQ, Lanark, Scotland
[2] Aalborg Univ, Ctr Sensory Motor Interact, Aalborg, Denmark
关键词
ankle joint stiffness; FES; functional electrical stimulation; integrated voluntary control; multipurpose rehabilitation frame; paraplegic standing;
D O I
10.1109/TNSRE.2002.806830
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The objective of this paper was to investigate whether a paraplegic subject-is able to maintain balance during standing by means of voluntary and reflex activity of the upper body while being supported by closed loop controlled ankle stiffness using FES. The knees and hips of the subject were held in extended positions by a mechanical apparatus, which restricted movement to the sagittal plane. The subject underwent several training sessions where the appropriate level of stiffness around the ankles was maintained by the mechanical apparatus. This enabled the subject to learn how to use the upper body for. balancing. After the subject gained adequate skills closed-loop FES was employed to regulate ankle stiffness, replacing the stiffness provided by the apparatus. A method to control antagonist muscle moment was implemented. In subsequent standing sessions, the subject had no difficulties in maintaining balance. When the FES, support was withheld, the ability to balance was lost.
引用
收藏
页码:239 / 248
页数:10
相关论文
共 33 条
[1]  
Astrom K., 1997, COMPUTER CONTROLLED
[2]   ANKLE STIFFNESS OF STANDING HUMANS IN RESPONSE TO IMPERCEPTIBLE PERTURBATION - REFLEX AND TASK-DEPENDENT COMPONENTS [J].
FITZPATRICK, RC ;
TAYLOR, JL ;
MCCLOSKEY, DI .
JOURNAL OF PHYSIOLOGY-LONDON, 1992, 454 :533-547
[3]  
GOLLEE H, 2001, P 6 ANN C INT FUNCT
[4]   Design of feedback controllers for paraplegic standing [J].
Hunt, KJ ;
Gollee, H ;
Jaime, RP ;
Donaldson, ND .
IEE PROCEEDINGS-CONTROL THEORY AND APPLICATIONS, 2001, 148 (02) :97-108
[5]   Robust control of electrically-stimulated muscle using polynomial H∞ design [J].
Hunt, KJ ;
Jaime, RP ;
Gollee, H .
CONTROL ENGINEERING PRACTICE, 2001, 9 (03) :313-328
[6]   Control of paraplegic ankle joint stiffness using FES while standing [J].
Hunt, KJ ;
Gollee, H ;
Jaime, RP .
MEDICAL ENGINEERING & PHYSICS, 2001, 23 (08) :541-555
[7]   Optimal control of ankle joint moment: Toward unsupported standing in paraplegia [J].
Hunt, KJ ;
Munih, M ;
Donaldson, N ;
Barr, FMD .
IEEE TRANSACTIONS ON AUTOMATIC CONTROL, 1998, 43 (06) :819-832
[8]  
HUNT KJ, 2000, P 5 ANN C INT FUNCT
[9]   PrP genetics in sheep and the implications for scrapie and BSE [J].
Hunter, N .
TRENDS IN MICROBIOLOGY, 1997, 5 (08) :331-334
[10]   DESIGN AND SIMULATION OF CLOSED-LOOP ELECTRICAL-STIMULATION ORTHOSES FOR RESTORATION OF QUIET STANDING IN PARAPLEGIA [J].
JAEGER, RJ .
JOURNAL OF BIOMECHANICS, 1986, 19 (10) :825-835