Surface heating by transvaginal transducers

被引:13
作者
Calvert, J.
Duck, F. [1 ]
Clift, S.
Azaime, H.
机构
[1] Royal United Hosp, Dept Med Phys, Bath BA1 3NG, Avon, England
[2] Univ Bath, Dept Mech Engn, Bath BA2 7AY, Avon, England
关键词
thermal safety; transducer self-heating; transvaginal ultrasound scanning;
D O I
10.1002/uog.3973
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Objective This safety study was designed to investigate tissue heating close to the surface of transvaginal ultrasound transducers, with the objective of assessing the validity of manufacturing safety standards set by the International Electrotechnical Commission (IEC). Methods The transducers investigated in this study were held in contact with a layered soft-tissue mimicking material (TMM), and the temperature increase was measured at various depths using a miniature thermocouple. The temperature rise at 200s was recorded, and the measured profiles of temperature rise with depth were compared with profiles predicted from both analytical and numeric models. Two transvaginal transducers of different manufacturers were investigated, operating in B-mode imaging, color-flow imaging and pulsed Doppler modes, using scanner settings giving acoustic output power towards the upper end of that available. Results The greatest heating always occurred at the interface between the transducer and the TMM, and it reduced to about 0.1 times the surface temperature rise at a depth of 1 cm. A local maximum was observed in pulsed Doppler mode. A three-dimensional finite-element model which accounted for transducer dimensions gave a better prediction of temperature increase than a simple analytical model. The temperature profiles were compared with the depth of fetal tissue measured from a small survey of clinical scans. Conclusions It is provisionally concluded that the transducer surface temperature rise of 6 C allowed to manufacturers by the IEC may give rise to an associated worst-case contribution to temperature rise due to the transducer, in fetal tissue, of between 0.5 and 1 degrees C at 1-cm depth. The contribution to tissue heating at 2 cm and deeper is negligible. Copyright (c) 2007 ISUOG. Published by John Wiley & Sons, Ltd.
引用
收藏
页码:427 / 432
页数:6
相关论文
共 22 条
[1]   Self-heating of Diagnostic Ultrasound Transducers in Air and in Contact with Tissue Mimics [J].
Calvert, Justine ;
Duck, Francis .
ULTRASOUND, 2006, 14 (02) :100-108
[2]   SOFT-TISSUE TEMPERATURE RISE CAUSED BY SCANNED, DIAGNOSTIC ULTRASOUND [J].
CURLEY, MG .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1993, 40 (01) :59-66
[3]   ARRHENIUS RELATIONSHIPS FROM THE MOLECULE AND CELL TO THE CLINIC [J].
DEWEY, WC .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1994, 10 (04) :457-483
[4]   Comparison of finite element and heated disc models of tissue heating by ultrasound [J].
Doody, C ;
Duck, FA ;
Humphrey, VF .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2000, 26 (08) :1347-1355
[5]   SURFACE HEATING OF DIAGNOSTIC ULTRASOUND TRANSDUCERS [J].
DUCK, FA ;
STARRITT, HC ;
TERHAAR, GR ;
LUNT, MJ .
BRITISH JOURNAL OF RADIOLOGY, 1989, 62 (743) :1005-1013
[6]  
FRASER JD, 1996, Patent No. 555887
[7]   THE ROLE OF APOPTOSIS IN THE RESPONSE OF CELLS AND TUMORS TO MILD HYPERTHERMIA [J].
HARMON, BV ;
TAKANO, YS ;
WINTERFORD, CM ;
GOBE, GC .
INTERNATIONAL JOURNAL OF RADIATION BIOLOGY, 1991, 59 (02) :489-501
[8]   On the development of a method to measure the surface temperature of ultrasonic diagnostic transducers. [J].
Hekkenberg, RT ;
Bezemer, RA .
AMUM 2004: Advanced Metrology for Ultrasound in Medicine 2004, 2004, 1 :84-89
[9]  
HEKKENBERG RT, 2003, TNO PREVENTION HLTH
[10]  
*ICRU, 1998, 61 ICRU