Using a human visual system model to optimize soft-copy mammography display: Influence of display phosphor

被引:36
|
作者
Krupinski, E
Johnson, J
Roehrig, H
Lubin, J
机构
[1] Univ Arizona, Dept Radiol, Tucson, AZ 85724 USA
[2] Sarnoff Corp, Princeton, NJ USA
关键词
breast radiography; technology; cathode ray tubes; diagnostic radiology; observer performance; images; display;
D O I
10.1016/S1076-6332(03)80040-5
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives. The authors developed an efficient method for optimizing cathode ray tube performance for soft-copy digital mammography displays, based on correlation between the performance of human observers and the performance of a mathematical computer model of the human visual system. The authors measured radiologist performance on soft-copy display monitors with different phosphors and used these results to validate the human visual performance model. Materials and Methods. Six radiologists viewed a series of 250 mammographic images with microcalcifications of different contrast levels. They viewed images on two soft-copy display monitors with phosphor luminescence-one with P45 and the other with P104. The same images were analyzed with the JNDmetrix Visual Discrimination Model, which is based on psychophysical just-noticeable difference measurement principles and on frequency-channel vision-modeling principles. Receiver operating characteristic curves were generated for the human and model observers' performances, and results were compared statistically. Results. Both human and model performance (area under the receiver operating characteristic curve) was better overall with the P45 than with the P104 monitor, especially for microcalcifications in the midlevel contrast range. There was high correlation between the human and model observers. Conclusion. The results indicate that the type of phosphor in a display monitor can influence observer performance significantly and that a model based on characteristics of the human visual system can be used to predict human observer performance accurately.
引用
收藏
页码:161 / 166
页数:6
相关论文
共 15 条
  • [1] Using a human visual system model to optimize soft-copy mammography display: Influence of veiling glare
    Krupinski, EA
    Lubin, J
    Roehrig, H
    Johnson, J
    Nafziger, J
    ACADEMIC RADIOLOGY, 2006, 13 (03) : 289 - 295
  • [2] Using a human visual system model to optimize soft-copy mammography display: influence of MTF compensation
    Krupinski, EA
    Johnson, J
    Roehrig, H
    Engstrom, M
    Fan, JH
    Nafziger, J
    Lubin, J
    Dallas, WJ
    ACADEMIC RADIOLOGY, 2003, 10 (09) : 1030 - 1035
  • [3] Optimizing softcopy mammography displays using a human visual system model: influence of display phosphor
    Krupinski, E
    Johnson, J
    Roehrig, H
    Lubin, J
    MEDICAL IMAGING 2002: IMAGE PERCEPTION, OBSERVER PERFORMANCE, AND TECHNOLOGY ASSESSMENT, 2002, 4686 : 331 - 337
  • [4] Effect of soft-copy display supported by CAD on mammography screening performance
    Roelofs, AAJ
    van Woudenberg, S
    Otten, JDM
    Hendriks, JHCL
    Bödicker, A
    Evertsz, CJG
    Karssemeijer, N
    EUROPEAN RADIOLOGY, 2006, 16 (01) : 45 - 52
  • [5] Effect of soft-copy display supported by CAD on mammography screening performance
    Antonius A. J. Roelofs
    Sander van Woudenberg
    Johannes D. M. Otten
    Jan H. C. L. Hendriks
    Anke Bödicker
    Carl J. G. Evertsz
    Nico Karssemeijer
    European Radiology, 2006, 16 : 45 - 52
  • [6] Comparison of calcification specificity in digital mammography using soft-copy display versus screen-film mammography screen
    Kim, Hak Hee
    Pisano, Etta D.
    Cole, Elodia B.
    Jiroutek, Michael R.
    Muller, Keith E.
    Zheng, Yuanshui
    Kuzmiak, Cherie M.
    Koomen, Marcia A.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2006, 187 (01) : 47 - 50
  • [7] Interpretation of digital mammograms: Comparison of speed and accuracy of soft-copy versus printed-film display
    Pisano, ED
    Cole, EB
    Kistner, EO
    Muller, KE
    Hemminger, BM
    Brown, ML
    Johnston, RE
    Kuzmiak, CM
    Braeuning, MP
    Freimanis, RI
    Soo, MS
    Baker, JA
    Walsh, R
    RADIOLOGY, 2002, 223 (02) : 483 - 488
  • [8] Usefulness of reversed display of soft-copy abdominal radiographs for urinary calculi detection
    Kang, SS
    Kim, JK
    Ryu, JA
    Choi, N
    Bae, SJ
    Kim, B
    ACTA RADIOLOGICA, 2004, 45 (03) : 351 - 356
  • [9] Comparison of observer performance on soft-copy reading of digital chest radiographs: High resolution liquid-crystal display monitors versus cathode-ray tube monitors
    Park, Chang Min
    Lee, Hyun Ju
    Goo, Jin Mo
    Han, Dae Hee
    Kim, Ji Hoon
    Lim, Keun Young
    Kim, Seung Ho
    Kang, Jimmy Jayoung
    Kim, Kwang Gi
    Lee, Chang Hyun
    Chun, Eun Ju
    Im, Jung-Gi
    EUROPEAN JOURNAL OF RADIOLOGY, 2008, 66 (01) : 13 - 18
  • [10] Use of a human visual system model to predict the effects of display veiling glare on observer performance
    Krupinski, EA
    Johnson, J
    Roehrig, H
    Nafziger, J
    Fan, JH
    Lubin, J
    MEDICAL IMAGING 2004: IMAGE PERCEPTION, OBSERVER PERFORMANCE, AND TECHNOLOGY ASSESSMENT, 2004, 5372 : 423 - 429