Delivery of Inhaled Nitric Oxide During MRI to Ventilated Neonates and Infants

被引:1
作者
Carter, Bradley G. [1 ]
Swain, Rachel [1 ]
Hislop, Jaime [1 ]
Escudie, Mathilde [1 ]
Williams, Rachel H. [1 ]
机构
[1] Royal Childrens Hosp, Clin Technol Serv, Neonatal & Paediat Intens Care Units, Parkville, Vic, Australia
关键词
nitric oxide; magnetic resonance imaging; mechanical ventilators; critical care; infant; newborn child; BLOOD-FLOW;
D O I
10.4187/respcare.08408
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
BACKGROUND: Many pediatric and neonatal ICU patients receive nitric oxide (NO), with some also requiring magnetic resonance imaging (MRI) scans. MRI-compatible NO delivery devices are not always available. We describe and bench test a method of delivering NO during MRI using standard equipment in which a NO delivery device was positioned in the MRI control room with the NO blender component connected to oxygen and set to 80 ppm and delivering flow via 12 m of tubing to a MRI-compatible ventilator, set up inside the MRI scanner magnet room. METHODS: For our bench test, the ventilator was set up normally and connected to an infant test lung to simulate several patients of differing weight (ie, 4 kg, 10 kg, 20 kg). The NO blender delivered flows of 2-10 L/min to the ventilator to achieve a range of NO and oxygen concentrations monitored via extended tubing. The measured values were compared to calculated values. RESULTS: A range of NO concentrations (12-41 ppm) and F-IO2 values (0.67-0.97) were achieved during the bench testing. The additional flow increased delivered peak inspiratory pressure and PEEP by 1-5 cm H2O. Calculated values were within acceptable ranges and were used to create a lookup table. CONCLUSIONS: In clinical use, this system can safely generate a range of NO flows of 15-42 ppm with an accompanying F-IO2 range of 0.34-0.98. (C) 2021 Daedalus Enterprises.
引用
收藏
页码:1254 / 1262
页数:9
相关论文
共 50 条
  • [1] Inhaled Nitric Oxide Delivery Systems for Mechanically Ventilated and Nonintubated Patients: A Review
    Gianni, Stefano
    Carroll, Ryan W.
    Kacmarek, Robert M.
    Berra, Lorenzo
    RESPIRATORY CARE, 2021, 66 (06) : 1021 - 1028
  • [2] Delivery and monitoring of inhaled nitric oxide
    Young, JD
    Dyar, OJ
    INTENSIVE CARE MEDICINE, 1996, 22 (01) : 77 - 86
  • [3] Inhaled nitric oxide delivery and monitoring
    Montgomery, FJ
    Berssenbrugge, AD
    JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1999, 15 (05) : 325 - 335
  • [4] Response to Inhaled Nitric Oxide in Premature and Term Neonates
    Thomas Hoehn
    Martin F. Krause
    Drugs, 2001, 61 : 27 - 39
  • [5] Inhaled nitric oxide in infants and children with ARDS
    G Zobel
    S Rödl
    M Trop
    HM Grubbauer
    Critical Care, 4 (Suppl 1):
  • [6] Evaluation of nitrogen dioxide scavengers during delivery of inhaled nitric oxide
    Lindberg, L
    Rydgren, G
    BRITISH JOURNAL OF ANAESTHESIA, 1998, 81 (03) : 404 - 408
  • [7] Inhaled nitric oxide therapy for severe hypoxemia in hyperinflated mechanically ventilated bronchiolitis patient
    DonaireGarcia, Alvaro
    Dachepally, Rashmitha
    Hanna, William
    Latifi, Samir Q.
    Agarwal, Hemant S.
    RESPIRATORY MEDICINE CASE REPORTS, 2022, 37
  • [8] Inhaled nitric oxide for respiratory failure in preterm infants
    Barrington, Keith J.
    Finer, Neil
    COCHRANE DATABASE OF SYSTEMATIC REVIEWS, 2010, (12):
  • [9] Inhaled Nitric Oxide in Preterm Infants: A Systematic Review
    Donohue, Pamela K.
    Gilmore, Maureen M.
    Cristofalo, Elizabeth
    Wilson, Renee F.
    Weiner, Jonathan Z.
    Lau, Brandyn D.
    Robinson, Karen A.
    Allen, Marilee C.
    PEDIATRICS, 2011, 127 (02) : E414 - E422
  • [10] Inhaled nitric oxide for respiratory failure in preterm infants
    Barrington, Keith J.
    Finer, Neil
    Pennaforte, Thomas
    COCHRANE DATABASE OF SYSTEMATIC REVIEWS, 2017, (01):