Cyclic Fatigue and Fracture Characteristics of Ground and Twisted Nickel-Titanium Rotary Files

被引:132
作者
Kim, Hyeon-Cheol [2 ]
Yum, Jiwan [2 ]
Hur, Bock [2 ]
Cheung, Gary Shun-Pan [1 ]
机构
[1] Univ Hong Kong, Fac Dent, Area Endodont, Hong Kong, Hong Kong, Peoples R China
[2] Pusan Natl Univ, Sch Dent, Dept Conservat Dent, Pusan, South Korea
关键词
Cyclic fatigue; electropolishing; fracture; machining grooves; nickel-titanium; rotary instrument; surface defect; Twisted File; SCANNING-ELECTRON-MICROSCOPY; STAINLESS-STEEL HAND; ENDODONTIC INSTRUMENTS; BENDING PROPERTIES; CLINICAL-USE; PHASE-TRANSFORMATION; SHAPING ABILITY; PROFILE; DEFECTS; BEHAVIOR;
D O I
10.1016/j.joen.2009.09.037
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Introduction: The purposes of this study were to compare the fatigue resistance of traditional, ground nickel-titanium rotary instruments with the Twisted File and to examine the fracture characteristics of the fatigued fragment. Methods: Size #25, 0.06 tapered, TF (SybronEndo), RaCe (FKG Dentaire), Helix (Dia Dent), and Pro Taper F1 (Dentsply Maillefer) were examined with scanning electron microscope for surface characteristics before subjected to a cyclic (rotational bending) fatigue test. The time until fracture was recorded to calculate the number of revolutions for each instrument. The data were compared for differences by using analysis of variance and post hoc Scheffe test. The fragments were examined with scanning electron microscope both in lateral view and fractographically. Results: TF showed a significantly higher resistance to cyclic fatigue than other nickel-titanium files that were manufactured with a grinding process (P < .05). The path of crack propagation appeared to be different for electropolished (IF and RaCe) versus non-electropolished (Helix and Pro Taper) instruments. Conclusions: Although all specimens showed similar fractographic appearance, which indicated a similar fracture mechanism, instruments with abundant machining grooves seemed to have a higher risk of fatigue. (J Endod 2010;36:147-152)
引用
收藏
页码:147 / 152
页数:6
相关论文
共 45 条
[1]   SEM observations of nickel-titanium rotary endodontic instruments that fractured during clinical use [J].
Alapati, SB ;
Brantley, WA ;
Svec, TA ;
Powers, JM ;
Nusstein, JM ;
Daehn, GS .
JOURNAL OF ENDODONTICS, 2005, 31 (01) :40-43
[2]   Scanning electron microscope observations of new and used nickel-titanium rotary files [J].
Alapati, SB ;
Brantley, WA ;
Svec, TA ;
Powers, JM ;
Mitchell, JC .
JOURNAL OF ENDODONTICS, 2003, 29 (10) :667-669
[3]   K3 endo, ProTaper, and ProFile systems: Breakage and distortion, in severely curved roots of molars [J].
Ankrum, MT ;
Hartwell, GR ;
Truitt, JE .
JOURNAL OF ENDODONTICS, 2004, 30 (04) :234-237
[4]   Evaluation of single-use rotary nickel-titanium instruments [J].
Arens, FC ;
Hoen, MM ;
Steiman, HR ;
Dietz, GC .
JOURNAL OF ENDODONTICS, 2003, 29 (10) :664-666
[5]   Influence of simulated clinical use on the torsional behavior of nickel-titanium rotary endodontic instruments [J].
Bahia, MGA ;
Melo, MCC ;
Buono, VTL .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 2006, 101 (05) :675-680
[6]   A comparison of stainless steel hand and rotary nickel-titanium instrumentation using a silicone impression technique [J].
Chen, JL ;
Messer, HH .
AUSTRALIAN DENTAL JOURNAL, 2002, 47 (01) :12-20
[7]  
Cheung G., 2009, Endododontic Topics, V16, P1, DOI 10.1111/j.1601-1546.2009.00239.x
[8]   Fatigue testing of a NiTi rotary instrument. Part 2: fractographic analysis [J].
Cheung, G. S. P. ;
Darvell, B. W. .
INTERNATIONAL ENDODONTIC JOURNAL, 2007, 40 (08) :619-625
[9]   Does electropolishing improve the low-cycle fatigue behavior of a nickel-titanium rotary instrument in hypochlorite? [J].
Cheung, Gary S. P. ;
Shen, Ya ;
Darvell, Brian W. .
JOURNAL OF ENDODONTICS, 2007, 33 (10) :1217-1221
[10]   Defects in ProTaper S1 instruments after clinical use: fractographic examination [J].
Cheung, GSP ;
Peng, B ;
Bian, Z ;
Shen, Y ;
Darvell, BW .
INTERNATIONAL ENDODONTIC JOURNAL, 2005, 38 (11) :802-809