Adaptable switching schemes for time-encoded multichannel optical coherence tomography

被引:7
|
作者
Wartak, Andreas [1 ]
Beer, Florian [1 ,2 ]
Baumann, Bernhard [1 ]
Pircher, Michael [1 ]
Hitzenberger, Christoph K. [1 ]
机构
[1] Med Univ Vienna, Ctr Med Phys & Biomed Engn, Vienna, Austria
[2] Vienna Univ Technol, Inst Appl Phys, Vienna, Austria
基金
奥地利科学基金会;
关键词
optical coherence tomography; optical switching; Doppler; ophthalmology; retinal imaging; OCULAR BLOOD-FLOW; SWEPT-SOURCE; DOPPLER OCT; RANGE; INTERFEROMETRY; GLAUCOMA; MIRROR; LENGTH; MHZ;
D O I
10.1117/1.JBO.23.5.056010
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
We introduce the approach of variable time encoding for multichannel optical coherence tomography (OCT). High-speed fiber optical switches are applied for sequential sample arm switching to enable quasisimultaneous image acquisition from three different orientation angles. In comparison with previous multichannel OCT (using simultaneous sample illumination), time-encoded multichannel OCT has no need for division of illumination power among the respective channels to satisfy laser safety requirements. Especially for ophthalmic applications-in particular retinal imaging, which the presented prototype was developed for-this advantage strongly influences image quality through an enhanced sensitivity. Nevertheless, time encoding comes at the cost of a decrease in imaging speed due to sequential channel illumination. For the typical multichannel OCT modality Doppler OCT, this results in a reduction of the maximum unambiguously determinable Doppler velocity. However, we demonstrate that this drawback can be overcome by adaptation of the illumination channel switching scheme. Thus, a re-extension of the maximum unambiguously determinable Doppler frequency to the full A-scan rate of the tunable light source is presented. The performance of the technique is demonstrated by flow phantom experiments and measurements of retinal blood flow in the eyes of healthy human volunteers. (C) The Authors. Published by SPIE under a Creative Commons Attribution 3.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication.
引用
收藏
页数:12
相关论文
共 50 条
  • [41] Interlaced spectrally encoded confocal scanning laser ophthalmoscopy and spectral domain optical coherence tomography
    Tao, Yuankai K.
    Farsiu, Sina
    Izatt, Joseph A.
    BIOMEDICAL OPTICS EXPRESS, 2010, 1 (02): : 431 - 440
  • [42] Real-time phase resolved optical coherence tomography and optical Doppler tomography using optical Hilbert Transformation
    Zhao, YH
    Ding, ZH
    Ren, HW
    Miao, JJ
    Jung, WG
    Nelson, JS
    Chen, ZP
    COHERENCE DOMAIN OPTICAL METHODS IN BIOMEDICAL SCIENCE AND CLINICAL APPLICATIONS VI, 2002, 4619 : 115 - 119
  • [43] Dynamic properties of wavelength switching in a widely tunable semiconductor laser for optical coherence tomography
    Pajkovic, Rastko
    Garbi, Daniel
    Williams, Kevin
    Bentel, Erwin
    SEMICONDUCTOR LASERS AND LASER DYNAMICS IX, 2020, 11356
  • [44] Integrated optical imaging system composed of optical time-of-flight and optical coherence tomography
    Sanchez, Juan Manuel Franco
    Shimamoto, Yuki
    Cervantes, Joel
    Kagawa, Keiichiro
    Nagahara, Hajime
    Hayasaki, Yoshio
    BIOMEDICAL IMAGING AND SENSING CONFERENCE 2021, 2021, 11925
  • [45] Polarization sensitive en face optical coherence tomography using multichannel acousto-optic deflectors
    Zurauskas, Mantas
    Rogers, John
    Podoleanu, Adrian Gh.
    OPTICAL COHERENCE TOMOGRAPHY AND COHERENCE DOMAIN OPTICAL METHODS IN BIOMEDICINE XVII, 2013, 8571
  • [46] A linearly optical scanning mirror for real-time imaging of optical coherence tomography
    Lin, CH
    Hou, MTK
    Hsu, IJ
    Wen, DC
    IEEE/LEOS OPTICAL MEMS 2005: INTERNATIONAL CONFERENCE ON OPTICAL MEMS AND THEIR APPLICATIONS, 2005, : 59 - 60
  • [47] All reflective optical delay line for real-time optical coherence tomography
    Kao, Chia-Yun
    Chang, Chao-Ming
    Hou, Max Ti-Kuang
    Hsu, I-Jen
    PHOTONICS NORTH 2006, PTS 1 AND 2, 2006, 6343
  • [48] Explicit and Semi-implicit Complex-Diffusion Schemes for Optical Coherence Tomography Despeckling
    Rodrigues, Pedro
    Guimaraes, Pedro
    Araujo, Aderito
    Barbeiro, Silvia
    Bernardes, Rui
    Serranho, Pedro
    IMAGE ANALYSIS AND RECOGNITION, 2013, 7950 : 282 - 289
  • [49] Multiscan time-domain optical coherence tomography for retina imaging
    Rosa, Carla Carmelo
    Rogers, John
    Pedro, Justin
    Rosen, Richard
    Podoleanu, Adrian
    APPLIED OPTICS, 2007, 46 (10) : 1795 - 1808
  • [50] Continuous focus tracking for real-time optical coherence tomography
    Cobb, MJ
    Liu, XM
    Li, XD
    OPTICS LETTERS, 2005, 30 (13) : 1680 - 1682