Tolerance Measurements on Internal- and External-Hexagon Implants

被引:23
作者
Braian, Michael [1 ]
De Bruyn, Hugo [2 ]
Fransson, Hakan [3 ]
Christersson, Cecilia [4 ]
Wennerberg, Ann [1 ]
机构
[1] Malmo Univ, Fac Odontol, Dept Prosthodont, S-20506 Malmo, Sweden
[2] Univ Ghent, Fac Med & Hlth Sci, B-9000 Ghent, Belgium
[3] Malmo Univ, Dept Technol, S-20506 Malmo, Sweden
[4] Malmo Univ, S-20506 Malmo, Sweden
关键词
dental implants; external-connection implant; internal-connection implant; machining tolerances; misfit; FIXED PARTIAL DENTURES; IN-VITRO; OSSEOINTEGRATED IMPLANTS; MARGINAL FIT; IMPRESSION TECHNIQUES; TITANIUM FRAMEWORKS; PASSIVE FIT; SCREW; COMPLICATIONS; COMPONENTS;
D O I
10.11607/jomi.3242
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: To measure the horizontal machining tolerances of the interface between internal-and external-hexagon implants and analogs with corresponding components after delivery from the manufacturer. These values may be a valuable tool for evaluating increasing misfit caused by fabrication, processing, and wear. Materials and Methods: Seven implants and seven analogs with external-and internal-hexagon connections (Biomet 3i) with corresponding prefabricated gold cylinders and gold screws, prefabricated cylindric plastic cylinders, and laboratory screws were studied. One set of components from the external and internal groups was measured manually and digitally. Measurements from the test subjects were compared with identical measurements from the virtual model to obtain threshold values. The virtual model was then used to obtain optimally oriented cuts. Results: The horizontal machining tolerances for castable plastic abutments on external implants were 12 +/- 89 mu m, and for internal implants they were 86 +/- 47 mu m. Tolerance measurements on prefabricated gold abutments for external implants were 44 +/- 9 mu m, and for internal implants they were 58 +/- 28 mu m. Conclusion: The groups with metallic components showed the smallest tolerance at < 50 mu m for the external group and < 90 mu m for the internal group. The prefabricated plastic cylinder groups ranged from < 100 mu m for external and < 130 mu m for internal connection.
引用
收藏
页码:846 / 852
页数:7
相关论文
共 50 条
[1]   MARGINAL TISSUE-REACTIONS AT OSSEOINTEGRATED TITANIUM FIXTURES .1. A 3-YEAR LONGITUDINAL PROSPECTIVE-STUDY [J].
ADELL, R ;
LEKHOLM, U ;
ROCKLER, B ;
BRANEMARK, PI ;
LINDHE, J ;
ERIKSSON, B ;
SBORDONE, L .
INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1986, 15 (01) :39-52
[2]  
Al Quran FA, 2012, QUINTESSENCE INT, V43, P119
[3]  
Al-Fadda SA, 2007, INT J PROSTHODONT, V20, P125
[4]   Biological and technical complications and failures with fixed partial dentures (FPD) on implants and teeth after four to five years of function [J].
Brägger, U ;
Aeschlimann, S ;
Bürgin, W ;
Hämmerle, CHF ;
Lang, NP .
CLINICAL ORAL IMPLANTS RESEARCH, 2001, 12 (01) :26-34
[5]  
Branemark P I, 1977, Scand J Plast Reconstr Surg Suppl, V16, P1
[6]  
Carr A B, 1996, Int J Oral Maxillofac Implants, V11, P589
[7]   An in vivo quantitative analysis of the fit of Nobel Biocare implant superstructures [J].
Cheshire, PD ;
Hobkirk, JA .
JOURNAL OF ORAL REHABILITATION, 1996, 23 (11) :782-789
[8]  
Cox J F, 1987, Int J Oral Maxillofac Implants, V2, P91
[9]   Correlation between vertical misfits and stresses transmitted to implants from metal frameworks [J].
de Torres, Erica Miranda ;
Seabra Barbosa, Gustavo Augusto ;
Bernardes, Sergio Rocha ;
Chiarello de Mattos, Maria da Gloria ;
Ribeiro, Ricardo Faria .
JOURNAL OF BIOMECHANICS, 2011, 44 (09) :1735-1739
[10]   The effect of commercially pure titannium and alternative dental alloys on the marginal fit of one-piece cast implant frameworks [J].
de Torres, Erica Miranda ;
Silveira Rodrigues, Renata Cristina ;
Chiarello de Mattos, Maria da Gloria ;
Ribeiro, Ricardo Faria .
JOURNAL OF DENTISTRY, 2007, 35 (10) :800-805