Evaluation of the Performance of Direct Susceptibility Test by VITEK-2 from Positively Flagged Blood Culture Broth for Gram-Negative Bacilli

被引:9
作者
Kavipriya, D. [1 ]
Prakash, Suman Susan [1 ]
Dhandapani, Sarumathi [1 ]
Rajshekar, Deepashree [1 ]
Sastry, Apurba Sankar [1 ]
机构
[1] Jawaharlal Inst Postgrad Med Educ & Res, Dept Microbiol, Pondicherry, India
关键词
antimicrobial susceptibility testing; bloodstream infections; categorical agreement; essential agreement; Enterobacteriaceae; nonfermenters; Vitek-2; DISK DIFFUSION; IDENTIFICATION; DIAGNOSIS;
D O I
10.1055/s-0041-1732489
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Timely initiation of antimicrobial therapy in patients with blood stream infection is absolutely necessary to reduce mortality and morbidity. Most clinical microbiology laboratories use conventional methods for identification and antimicrobial susceptibility testing (AST) that involve biochemical methods for identification followed by AST by disk diffusion. The aim of the current study is to assess the various errors associated with direct susceptibility testing done from blood culture broth using automated AST system-Vitek-2 compact compared with the reference method of AST done from bacterial colonies. Materials and Methods The study was conducted in a tertiary care public sector 2,200-bedded hospital in South India for a period of 6 months. The study involved positively flagged blood culture bottles that yielded single morphotype of Gram- negative organism by Gram stain. A total of 120 bacterial isolates were collected that consisted of consecutively obtained first 60 isolates of Enterobacteriaceae family ( 30 Escherichia coli and 30 Klebsiella pneumoniae) and consecutively obtained first 60 nonfermenters (30 Pseudomonas aeruginosa and 30 Acinetobacter baumannii). Vitek-2 AST was done from these 120 blood culture broth, following the protocol by Biomerieux, and results were obtained. Then, Vitek-2 was done from colonies (reference method) using appropriate panel for Enterobacteriaceae and nonfermenters, and results were obtained. Both the results were compared. Results Nonfermenters showed a better categorical agreement of 97.6%, as compared to Enterobacteriaceae, which showed 97%. Among Enterobacteriaceae, both E. coli and K. pneumoniae showed categorical agreement of 97% each. Conclusion The procedure of AST directly from blood culture broth represents a simple and effective technique that can reduce the turnaround time by 24 hours, which in turn benefits the clinician in appropriate utilization of antimicrobials for better patient care.
引用
收藏
页码:374 / 379
页数:6
相关论文
共 20 条
[1]  
[Anonymous], 2077612006 ISO
[2]   A new rapid method for direct antimicrobial susceptibility testing of bacteria from positive blood cultures [J].
Barnini, Simona ;
Brucculeri, Veronica ;
Morici, Paola ;
Ghelardi, Emilia ;
Florio, Walter ;
Lupetti, Antonella .
BMC MICROBIOLOGY, 2016, 16
[3]   Direct susceptibility testing by disk diffusion on clinical samples: a rapid and accurate tool for antibiotic stewardship [J].
Coorevits, L. ;
Boelens, J. ;
Claeys, G. .
EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES, 2015, 34 (06) :1207-1212
[4]   RAPID ANTIMICROBIAL SUSCEPTIBILITY TESTING OF ISOLATES FROM BLOOD CULTURES BY DIRECT INOCULATION AND EARLY READING OF DISK DIFFUSION TESTS [J].
COYLE, MB ;
MCGONAGLE, LA ;
PLORDE, JJ ;
CLAUSEN, CR ;
SCHOENKNECHT, FD .
JOURNAL OF CLINICAL MICROBIOLOGY, 1984, 20 (03) :473-477
[5]   Rapid and accurate direct antibiotic susceptibility testing of blood culture broths using MALDI Sepsityper combined with the BD Phoenix automated system [J].
Hazelton, Briony ;
Thomas, Lee C. ;
Olma, Thomas ;
Kok, Jen ;
O'Sullivan, Matthew ;
Chen, Sharon C. -A. ;
Iredell, Jonathan R. .
JOURNAL OF MEDICAL MICROBIOLOGY, 2014, 63 :1590-1594
[6]   Automated direct screening for resistance of Gram-negative blood cultures using the BD Kiestra WorkCell [J].
Heather, C. S. ;
Maley, M. .
EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES, 2018, 37 (01) :117-125
[7]   Rapid antibiotic susceptibility testing in blood culture diagnostics performed by direct inoculation using the VITEK®-2 and BD Phoenix platforms [J].
Hoering, Steffen ;
Massarani, Alain Sami ;
Loeffler, Bettina ;
Roedel, Juergen .
EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES, 2019, 38 (03) :471-478
[8]   Rapid antimicrobial susceptibility testing by VITEK®2 directly from blood cultures in patients with Gram-negative rod bacteremia [J].
Hogan, C. A. ;
Watz, N. ;
Budvytiene, I. ;
Banaei, N. .
DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2019, 94 (02) :116-121
[9]   Current Concepts in Laboratory Testing to Guide Antimicrobial Therapy [J].
Jenkins, Stephen G. ;
Schuetz, Audrey N. .
MAYO CLINIC PROCEEDINGS, 2012, 87 (03) :290-308
[10]   SELECTION CRITERIA FOR AN ANTIMICROBIAL SUSCEPTIBILITY TESTING SYSTEM [J].
JORGENSEN, JH .
JOURNAL OF CLINICAL MICROBIOLOGY, 1993, 31 (11) :2841-2844