Evaluation of GAFCHROMIC® EBT film for CyberKnife® dosimetry

被引:114
作者
Wilcox, Ellen E. [1 ]
Daskalov, George M. [1 ]
机构
[1] St Francis Hosp & Med Ctr, Dept Radiat Oncol, Hartford, CT 06105 USA
关键词
EBT film dosimetry; stereotactic radiosurgery;
D O I
10.1118/1.2734384
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
External beam therapy (EBT) GAFCHROMIC (R) film is evaluated for dosimetry and characterization of the CyberKnife (R) radiation beams. Percentage depth doses, lateral beam profiles, and output factors are measured in solid water using EBT GAFCHROMIC (R) film (International Specialty Products, Wayne, NJ) for the 6 MV radiation beams of diameter 5 to 60 rum produced by the CyberKnife (R) (Accuray, Sunnyvale, CA). The data are compared to those measured with the PTW 60008 diode and the Wellhofer CC01 ion chamber in water. For the small radiation field sizes used in stereotactic radiosurgery, lateral electronic disequilibrium and steep dose gradients exist in a large portion of these fields, requiring the use of high-resolution measurement techniques. For small beams, the detector size approaches the dimensions of the beam and adversely affects measurement accuracy in regions where the gradient varies across the detector. When film is the detector, the scanning system is usually the resolution-limiting component. Radiographic films based upon silver halide (AgH) emulsions are widely used for relative dosimetry of external radiation treatment beams in the megavoltage energy range, because of their good spatial resolution and capability to provide integrated dosimetry over two dimensions. Film dosimetry, however, has drawbacks due to its steep energy dependence at low photon energies as well as film processor and densitometer artifacts. EBT radiochromic film, introduced in 2004 specifically for IMRT dosimetry, may be a detector of choice for the characterization of small radiosurgical beams, because of its near-tissue equivalence, radiation beam energy independence, high spatial resolution, and self developing properties. For radiation beam sizes greater than 10 rum, the film measurements were identical to those of the diode and ion chamber. For the smaller beam diameters of 7.5 and 5 rum, however, there were differences in the data measured with the different detectors, which are attributed to their different spatial resolution and non-water-equivalence. (c) 2007 American Association of Physicists in Medicine.
引用
收藏
页码:1967 / 1974
页数:8
相关论文
共 54 条
[1]   Monte Carlo study of a Cyberknife stereotactic radiosurgery system [J].
Araki, Fujio .
MEDICAL PHYSICS, 2006, 33 (08) :2955-2963
[2]   ABSORBED DOSE PERTURBATION CAUSED BY DIODES FOR SMALL-FIELD PHOTON DOSIMETRY [J].
BEDDAR, AS ;
MASON, DJ ;
OBRIEN, PF .
MEDICAL PHYSICS, 1994, 21 (07) :1075-1079
[3]  
BJARNGARD BE, 1990, MED PHYS, V17, P794, DOI 10.1118/1.596475
[4]   Interinstitutional variations of sensitometric curves of radiographic dosimetric films [J].
Bos, LJ ;
Danciu, C ;
Cheng, CW ;
Brugmans, MJP ;
van der Horst, A ;
Minken, A ;
Mijnheer, BJ .
MEDICAL PHYSICS, 2002, 29 (08) :1772-1780
[5]   Energy dependence of response of new high sensitivity radiochromic films for megavoltage and kilovoltage radiation energies [J].
Chiu-Tsao, ST ;
Ho, Y ;
Shankar, R ;
Wang, L ;
Harrison, LB .
MEDICAL PHYSICS, 2005, 32 (11) :3350-3354
[6]  
DARCIU C, 2001, MED PHYS, V28, P966
[7]   ANALYSIS OF PHYSICAL PARAMETERS ASSOCIATED WITH THE MEASUREMENT OF HIGH-ENERGY X-RAY PENUMBRA [J].
DAWSON, DJ ;
HARPER, JM ;
AKINRADEWO, AC .
MEDICAL PHYSICS, 1984, 11 (04) :491-497
[8]   Quantitative optical densitometry with scanning-laser film digitizers [J].
Dempsey, JF ;
Low, DA ;
Kirov, AS ;
Williamson, JF .
MEDICAL PHYSICS, 1999, 26 (08) :1721-1731
[9]   Validation of a precision radiochromic film dosimetry system for quantitative two-dimensional imaging of acute exposure dose distributions [J].
Dempsey, JF ;
Low, DA ;
Mutic, S ;
Markman, J ;
Kirov, AS ;
Nussbaum, GH ;
Williamson, JF .
MEDICAL PHYSICS, 2000, 27 (10) :2462-2475
[10]   Precise radiochromic film dosimetry using a flat-bed document scanner [J].
Devic, S ;
Seuntjens, J ;
Sham, E ;
Podgorsak, EB ;
Schmidtlein, CR ;
Kirov, AS ;
Soares, CG .
MEDICAL PHYSICS, 2005, 32 (07) :2245-2253