A review of factors affecting patient doses for barium enemas and meals

被引:21
作者
Martin, CJ [1 ]
机构
[1] Gartnavel Royal Hosp, Dept Clin Phys & Bioengn, Glasgow G12 0XH, Lanark, Scotland
关键词
D O I
10.1259/bjr/99143061
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A study of patient doses for barium enema and meat examinations has been carried out for hospitals in the West of Scotland to assess the impact of dose reduction facilities on new X-ray equipment. Dose-area product (DAP) information has been collected on examinations for groups of patients at 20 hospitals and results reviewed together with data on equipment performance measurements. Median DAPs for barium enemas and meals were 15.7 Gy cm(2) and 4.8 Gy cm(2), respectively, and effective doses estimated from the results are 3.5+/-0.7 mSv and 1.5+/-0.5 mSv, respectively. These doses are lower than those reported in earlier studies and in previous surveys in the West of Scotland. The reduction in dose is linked primarily to the low dose facilities available on newer X-ray equipment, such as low dose pulsed fluoroscopy, digital imaging facilities and use of copper filtration. Use of the image intensifier for decubitus images on C-arm units employed for barium enemas also gives a significantly lower dose. Equipment with copper filtration had the lowest doses. The reduction in effective dose will be significantly less than the reduction in DAP for units in which a copper filter is included and the adoption of lower diagnostic reference levels is proposed for units with this facility. It is important that the operators are aware of the low dose imaging options on their equipment in order that techniques can be fully optimized.
引用
收藏
页码:864 / 868
页数:5
相关论文
共 20 条
[1]   THE IMPACT OF DIGITAL IMAGING ON PATIENT DOSES DURING BARIUM STUDIES [J].
BROADHEAD, DA ;
CHAPPLE, CL ;
FAULKNER, K .
BRITISH JOURNAL OF RADIOLOGY, 1995, 68 (813) :992-996
[2]   Radiation doses for barium enema and barium meal examinations in Ireland: potential diagnostic reference levels [J].
Carroll, EM ;
Brennan, PC .
BRITISH JOURNAL OF RADIOLOGY, 2003, 76 (906) :393-397
[3]  
CORBETT RH, 2000, P 10 INT C IRPA
[4]   Reducing dose at barium enema: radiographers do it digitally [J].
Crawley, MT ;
Booth, A .
BRITISH JOURNAL OF RADIOLOGY, 2002, 75 (896) :652-656
[5]   Patient dose due to colon examination: Dose assessment and results from a survey in the Netherlands [J].
Geleijns, J ;
Broerse, JJ ;
Shaw, MPC ;
Schultz, FW ;
Teeuwisse, W ;
vanUnnik, JG ;
Zoetelief, J .
RADIOLOGY, 1997, 204 (02) :553-559
[6]   ESTIMATION OF EFFECTIVE DOSE FROM DOSE-AREA PRODUCT MEASUREMENTS FOR BARIUM MEALS AND BARIUM ENEMAS [J].
HART, D ;
WALL, BF .
BRITISH JOURNAL OF RADIOLOGY, 1994, 67 (797) :485-489
[7]  
Hart D., 1994, R262 NRPB
[8]  
Hart D, 2002, W14 NRPB
[9]  
HART D, 1996, R289 NRPB
[10]  
HART D, 1996, R279 NRPB