Assessment of arterially hyper-enhancing liver lesions using virtual monoenergetic images from spectral detector CT: phantom and patient experience

被引:60
作者
Hokamp, N. Grosse [1 ,2 ,3 ]
Hoeink, A. J. [1 ]
Doerner, J. [1 ]
Jordan, D. W. [2 ,3 ]
Pahn, G. [4 ]
Persigehl, T. [1 ]
Maintz, D. [1 ]
Haneder, S. [1 ]
机构
[1] Univ Hosp Cologne, Dept Diagnost & Intervent Radiol, Kerpener Str 62, D-50937 Cologne, Germany
[2] Univ Hosp Cleveland, Dept Radiol, 2074 Abington Rd, Cleveland, OH 44106 USA
[3] Case Western Reserve Univ, Sch Med, Dept Radiol, Cleveland, OH 44106 USA
[4] Philips Clin Sci CT, Hamburg, Germany
关键词
Spectral detector computed tomography; Image quality; Virtual monoenergetic images; Hyper-attenuating liver lesions; DUAL-ENERGY CT; KILOELECTRON VOLT SETTINGS; TO-NOISE RATIO; ITERATIVE RECONSTRUCTION; DOSE REDUCTION; PULMONARY ANGIOGRAPHY; COMPUTED-TOMOGRAPHY; ABDOMINAL CT; CONTRAST; QUALITY;
D O I
10.1007/s00261-017-1411-1
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To investigate a benefit from virtual monoenergetic reconstructions (VMIs) for assessment of arterially hyper-enhancing liver lesions in phantom and patients and to compare hybrid-iterative and spectral image reconstructions of conventional images (CI-IR and CI-SR). All imaging was performed on a SDCT (Philips Healthcare, Best, The Netherlands). Images of a non-anthropomorphic phantom with a lesion-mimicking insert (containing iodine in water solution) and arterial-phase images from contrast-enhanced patient examinations were evaluated. VMIs (40-200 keV, 10 keV increment), CI-IR, and CI-SR were reconstructed using different strengths of image denoising. ROIs were placed in lesions, liver/matrix, muscle; signal-to-noise, contrast-to-noise, and lesion-to-liver ratios (SNR, CNR, and LLR) were calculated. Qualitatively, 40, 70, and 110 keV and CI images were assessed by two radiologists on five-point Likert scales regarding overall image quality, lesion assessment, and noise. In phantoms, SNR was increased threefold by VMI40keV compared with CI-IR/SR (5.8 +/- 1.1 vs. 18.8 +/- 2.2, p ae<currency> 0.001), while no difference was found between CI-IR and CI-SR (p = 1). Denoising was capable of noise reduction by 40%. In total, 20 patients exhibiting 51 liver lesions were assessed. Attenuation was the highest in VMI40keV, while image noise was comparable to CI-IR resulting in a threefold increase of CNR/LLR (CI-IR 1.3 +/- 0.8/4.4 +/- 2.0, VMI40keV: 3.8 +/- 2.7/14.2 +/- 7.5, p ae<currency> 0.001). Subjective lesion delineation was the best in VMI40keV image (p ae<currency> 0.01), which also provided the lowest perceptible noise and the best overall image quality. VMIs improve assessment of arterially hyper-enhancing liver lesions since they increase lesion contrast while maintaining low image noise throughout the entire keV spectrum. These data suggest that to consider VMI screening after arterially hyper-enhancing liver lesions.
引用
收藏
页码:2066 / 2074
页数:9
相关论文
共 34 条
  • [1] Advanced image-based virtual monoenergetic dual-energy CT angiography of the abdomen: optimization of kiloelectron volt settings to improve image contrast
    Albrecht, Moritz H.
    Scholtz, Jan-Erik
    Huesers, Kristina
    Beeres, Martin
    Bucher, Andreas M.
    Kaup, Moritz
    Martin, Simon S.
    Fischer, Sebastian
    Bodelle, Boris
    Bauer, Ralf W.
    Lehnert, Thomas
    Vogl, Thomas J.
    Wichmann, Julian L.
    [J]. EUROPEAN RADIOLOGY, 2016, 26 (06) : 1863 - 1870
  • [2] Dual-energy-CT of hypervascular liver lesions in patients with HCC: investigation of image quality and sensitivity
    Altenbernd, Jens
    Heusner, Till A.
    Ringelstein, Adrian
    Ladd, Susanne C.
    Forsting, Michael
    Antoch, Gerald
    [J]. EUROPEAN RADIOLOGY, 2011, 21 (04) : 738 - 743
  • [3] Value of monoenergetic low-kV dual energy CT datasets for improved image quality of CT pulmonary angiography
    Apfaltrer, Paul
    Sudarski, Sonja
    Schneider, David
    Nance, John W., Jr.
    Haubenreisser, Holger
    Fink, Christian
    Schoenberg, Stefan O.
    Henzler, Thomas
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2014, 83 (02) : 322 - 328
  • [4] Improvements to image quality using hybrid and model-based iterative reconstructions: a phantom study
    Aurumskjold, Marie-Louise
    Ydstrom, Kristina
    Tingberg, Anders
    Soderberg, Marcus
    [J]. ACTA RADIOLOGICA, 2017, 58 (01) : 53 - 61
  • [5] HYPERVASCULAR HEPATIC METASTASES - CT EVALUATION
    BRESSLER, EL
    ALPERN, MB
    GLAZER, GM
    FRANCIS, IR
    ENSMINGER, WD
    [J]. RADIOLOGY, 1987, 162 (01) : 49 - 51
  • [6] CONTRAST ENHANCEMENT OF FOCAL HEPATIC-LESIONS IN CT - EFFECT OF SIZE AND HISTOLOGY
    BURGENER, FA
    HAMLIN, DJ
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1983, 140 (02) : 297 - 301
  • [7] AJR teaching file:: Hypervascular metastasis or hepatic hemangioma?
    Burkholz, Kimberly J.
    Silva, Alvin C.
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2008, 190 (06) : S53 - S56
  • [8] The AAPM/RSNA physics tutorial for residents - X-ray interactions
    Bushberg, JT
    [J]. RADIOGRAPHICS, 1998, 18 (02) : 457 - 468
  • [9] Carrascosa PM, 2015, CARDIOVASC IMAGING, DOI [10.1007/978-3-319-21227-2, DOI 10.1007/978-3-319-21227-2]
  • [10] Assessment of a Model-Based, Iterative Reconstruction Algorithm (MBIR) Regarding Image Quality and Dose Reduction in Liver Computed Tomography
    Chang, Won
    Lee, Jeong Min
    Lee, Kyunghee
    Yoon, Jeong Hee
    Yu, Mi Hye
    Han, Joon Koo
    Choi, Byung Ihn
    [J]. INVESTIGATIVE RADIOLOGY, 2013, 48 (08) : 598 - 606