On the potential of ROI imaging in x-ray CT - A comparison of novel dynamic beam attenuators with current technology

被引:2
作者
Huck, Sascha Manuel [1 ,2 ]
Fung, George S. K. [3 ]
Parodi, Katia [2 ]
Stierstorfer, Karl [1 ]
机构
[1] Siemens Healthcare GmbH, Forchheim, Germany
[2] Ludwig Maximilians Univ Munchen, Fac Phys, Dept Med Phys, Garching, Germany
[3] Siemens Med Solut USA Inc, Malvern, PA 19355 USA
关键词
computed tomography; x-ray fluence filtration; fluence field modulation; dynamic beam attenuator; region-of-interest imaging; dose reduction; image variance; noise power spectrum; TUBE CURRENT MODULATION; NOISE POWER SPECTRUM; DOSE REDUCTION; COMPUTED-TOMOGRAPHY; DETECTOR PERFORMANCE; PULSE PILEUP; REALIZATION; MANAGEMENT; CHILDHOOD; PHANTOMS;
D O I
10.1002/mp.14879
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: In this work, we explore the potential of region-of-interest (ROI) imaging in x-ray computed tomography (CT). Using two dynamic beam attenuator (DBA) concepts for fluence field modulation (FFM) previously developed, we investigate and evaluate the potential dose savings in comparison with current FFM technology. Methods: ROI imaging is a special application of FFM where the bulk of x-ray radiation is propagated toward a certain anatomical target (ROI), specified by the imaging task, while the surrounding tissue is spared from radiation. We introduce a criterion suitable to quantitatively describe the balance between image quality inside an ROI and total radiation dose with respect to a given ROI imaging task. It accounts for the mean image variance at the ROI and the effective patient dose calculated from Monte Carlo simulations. The criterion is further used to compile task-specific DBA trajectories determining the primary x-ray fluence, and eventually used for comparing different FFM techniques, namely the sheet-based dynamic beam attenuator (sbDBA), the z-aligned sbDBA (z-sbDBA), and an adjustable static operation mode of the z-sbDBA. Furthermore, two static bowtie filters and the influence of tube current modulation (TCM) are included in the comparison. Results: Our findings demonstrate by simulations that the presented trajectory optimization method determines reasonable DBA trajectories. The influence of TCM is strongly depending on the imaging task. The narrow bowtie filter allows for dose reductions of about 10% compared to the regular bowtie filter in the considered ROI imaging tasks. The DBAs are shown to realize substantially larger dose reductions. In our cardiac imaging scenario, the DBAs can reduce the effective dose by about 30% (z-sbDBA) or 60% (sbDBA). We can further verify that the noise characteristics are not adversely affected by the DBAs. Conclusion: Our research demonstrates that ROI imaging using the presented DBA concepts is a promising technique toward a more patient- and task-specific CT imaging requiring lower radiation dose. Both the sbDBA and the z-sbDBA are potential technical solutions for realizing ROI imaging in x-ray CT. (C) 2021 The Authors. Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine.
引用
收藏
页码:3479 / 3499
页数:21
相关论文
共 77 条
[1]  
Arenson JS., 2008, US Patent, Patent No. [7,330,535 B2, 7330535]
[2]   The noise power spectrum in CT with direct fan beam reconstruction [J].
Baek, Jongduk ;
Pelc, Norbert J. .
MEDICAL PHYSICS, 2010, 37 (05) :2074-2081
[3]  
Bartolac, 2011, MED PHYS, V38, P2
[4]   Compensator models for fluence field modulated computed tomography [J].
Bartolac, Steven ;
Jaffray, David .
MEDICAL PHYSICS, 2013, 40 (12)
[5]   CT scanning the early days [J].
Beckmann, EC .
BRITISH JOURNAL OF RADIOLOGY, 2006, 79 (937) :5-8
[6]   Current concepts - Computed tomography - An increasing source of radiation exposure [J].
Brenner, David J. ;
Hall, Eric J. .
NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (22) :2277-2284
[7]   Estimated risks of radiation-induced fatal cancer from pediatric CT [J].
Brenner, DJ ;
Elliston, CD ;
Hall, EJ ;
Berdon, WE .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 176 (02) :289-296
[8]  
BROOKS R A, 1976, Medical Physics (Woodbury), V3, P237, DOI 10.1118/1.594240
[9]   Association between power law coefficients of the anatomical noise power spectrum and lesion detectability in breast imaging modalities [J].
Chen, Lin ;
Abbey, Craig K. ;
Boone, John M. .
PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (06) :1663-1681
[10]   Dose Reduction in Helical CT: Dynamically Adjustable z-Axis X-Ray Beam Collimation [J].
Christner, Jodie A. ;
Zavaletta, Vanessa A. ;
Eusemann, Christian D. ;
Walz-Flannigan, Alisa I. ;
McCollough, Cynthia H. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 194 (01) :W49-W55