Corneal endothelial cell damage after lens extraction using the fluid-based system compared to ultrasound phacoemulsification in human cadaver eyes
被引:8
|作者:
Sandoval, HP
[1
]
Al Sarraf, O
[1
]
Vroman, DT
[1
]
Solomon, KD
[1
]
机构:
[1] Med Univ S Carolina, Storm Eye Inst, Magee Res Ctr Vis Correct, Charleston, SC 29425 USA
来源:
关键词:
cataract extraction;
phacoemulsification;
fluid-based system;
endothelial cell loss;
AquaLase;
D O I:
10.1097/01.ico.0000130184.56514.51
中图分类号:
R77 [眼科学];
学科分类号:
100212 ;
摘要:
Purpose: To compare the endothelial cell damage after lens removal using a fluid-based system to conventional ultrasound. Methods: Twenty human cadaver eyes were randomized to undergo phacoemulsification using either a fluid-based system or conventional ultrasound. After surgery, each corneoscleral button was removed, stained and damaged endothelial cells per square millimeter were assessed using light microscopy. Results: Fluid-based system mean damaged endothelial cells/mm(2) was 60.2 +/- 24.1 compared with 60.4 +/- 42.6 when using conventional ultrasound (P = 0.248). Conclusion: The fluid-based system is as safe for the corneal endothelium as conventional ultrasound during cataract removal in human cadaver eyes.
引用
收藏
页码:720 / 722
页数:3