Buttress technique to insure accurate placement of AneuRx stent-grafts in patients with complex anatomy

被引:0
作者
Krajcer, Z [1 ]
Strickman, N [1 ]
Dougherty, K [1 ]
Howell, MH [1 ]
机构
[1] St Lukes Episcopal Hosp, Texas Heart Inst, Dept Cardiol, Houston, TX 77030 USA
关键词
abdominal aortic aneurysm; neck morphology; neck angulation; neck calcification; maldeployment;
D O I
10.1583/1545-1550(2002)009<0772:BTTIAP>2.0.CO;2
中图分类号
R61 [外科手术学];
学科分类号
摘要
Purpose: To describe a technique that prevents caudal displacement of the AneuRx stent-graft during removal of the nose cone and runners. Technique: The buttress technique uses a 16-F sheath positioned from the contralateral femoral approach at the inferior edge of the endograft's main body to provide safe retraction of the runners and the nose cone and accurate endograft deployment. Conclusions: If accurate deployment of the stent-graft according to the manufacturer's directions is not possible, this buttress technique can insure proper positioning. It adds no additional time or equipment to the procedure, and it appears to be a safe and easily applied maneuver.
引用
收藏
页码:772 / 776
页数:5
相关论文
共 8 条
[1]  
Ahn SS, 2001, ANN VASC SURG, V15, P123
[2]  
Broeders IAMJ, 2000, J ENDOVASC THER, V7, P389, DOI 10.1583/1545-1550(2000)007<0389:ASTTIT>2.0.CO
[3]  
2
[4]  
Howell M, 2001, J ENDOVASC THER, V8, P68, DOI 10.1583/1545-1550(2001)008<0068:PAACOF>2.0.CO
[5]  
2
[6]   Preliminary results of endovascular abdominal aortic aneurysm exclusion with the AneuRx stent-graft [J].
Howell, MH ;
Strickman, N ;
Mortazavi, A ;
Hallman, CH ;
Krajcer, Z .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2001, 38 (04) :1040-1046
[7]  
LAWRENCEBROWN MMD, 1999, DURABILITY VASCULAR, P375
[8]   Aneurysm rupture after endovascular repair using the AneuRx stent graft [J].
Zarins, CK ;
White, RA ;
Fogarty, TJ .
JOURNAL OF VASCULAR SURGERY, 2000, 31 (05) :960-970