Synthesis and evaluation of thermoluminescence properties of ZrO2:Mg for radiotherapy dosimetry

被引:3
作者
Amarlou, Asieh [1 ]
Mohammadi, Kheirollah [2 ]
Banaee, Nooshin [3 ]
Nedaei, Hassan Ali [1 ,4 ]
机构
[1] Univ Tehran Med Sci, Dept Med Phys & Biomed Engn, Tehran, Iran
[2] Malek Ashtar Univ, Tehran, Iran
[3] Islamic Azad Univ, Cent Tehran Branch, Med Radiat Res Ctr, Tehran, Iran
[4] Univ Tehran Med Sci, Canc Inst, Radiat Oncol Res Ctr, Tehran, Iran
关键词
Thermoluminescence; Dosimetry; ZrO2; Mg; Radiotherapy; EXIT DOSE MEASUREMENTS; ZIRCONIUM-OXIDE; ELECTRON;
D O I
10.1007/s00411-021-00928-4
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
This study aimed to investigate the thermoluminescent properties of ZrO2:Mg irradiated with a 6 MV X-ray beam and its potential application in radiotherapy dosimetry. ZrO2 powder was synthesized using the sol-gel method and Mg was used as a dopant. Irradiations were performed with ZrO2:Mg chips located at the center of a 10 x 10 cm(2) radiation field at a source surface distance of 100 cm, below a stack of solid water slabs, at the depth of maximum absorbed dose. The investigated characteristics of the material included linearity with radiation dose, reproducibility, accuracy, sensitivity and fading. Regarding the intrinsic difference of the samples, the glow curves of the investigated ZrO2:Mg chips exposed to 1 Gy of 6 MV X-rays exhibited three or four peaks. The ZrO2:Mg samples showed a 47% fading at 24 h after irradiation, and the reproducibility of the thermoluminescence reading of ZrO2:Mg for equal irradiation conditions was +/- 21%. The thermoluminescence response of the investigated ZrO2:Mg samples to various absorbed doses from 0.5 to 2.5 Gy showed a gentle increase of the thermoluminescence intensity with increasing absorbed dose. The obtained results show that ZrO2:Mg is not an appropriate candidate for X-ray photons in radiotherapy, due to low thermoluminescence peak temperature, low reproducibility, low sensitivity to various absorbed doses and significant fading.
引用
收藏
页码:647 / 652
页数:6
相关论文
共 17 条
  • [1] Thermoluminescence of Eu-doped ZrO2 thin films exposed to ultraviolet and visible light
    Azorin, J
    Rivera, T
    Martinez, E
    Garcia, M
    [J]. RADIATION MEASUREMENTS, 1998, 29 (3-4) : 315 - 317
  • [2] ELECTRON AND PHOTON BEAMS FROM A 50 MEV RACETRACK MICROTRON
    BRAHME, A
    KRAEPELIEN, T
    SVENSSON, H
    [J]. ACTA RADIOLOGICA ONCOLOGY, 1980, 19 (04): : 305 - 319
  • [3] Edmund J.M., 2007, EFFECTS TEMPERATURE
  • [4] Automated TLD system for tumor dose estimation from exit dose measurements in external ream radiotherapy
    Ferguson, HM
    Lambert, GD
    Harrison, RM
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (04): : 899 - 905
  • [5] Modeling of the thermoluminescence mechanisms in ZrO2
    Kadari, Ahmed
    Kadri, Dahane
    [J]. APPLIED RADIATION AND ISOTOPES, 2013, 82 : 49 - 54
  • [6] Materials for thermoluminescent dosimetry: Current status and future trends
    Kortov, V.
    [J]. RADIATION MEASUREMENTS, 2007, 42 (4-5) : 576 - 581
  • [7] Entrance and exit dose measurements with semiconductors and thermoluminescent dosemeters: A comparison of methods and in vivo results
    Loncol, T
    Greffe, JL
    Vynckier, S
    Scalliet, P
    [J]. RADIOTHERAPY AND ONCOLOGY, 1996, 41 (02) : 179 - 187
  • [8] In vivo dosimetry in external beam radiotherapy
    Mijnheer, Ben
    Beddar, Sam
    Izewska, Joanna
    Reft, Chester
    [J]. MEDICAL PHYSICS, 2013, 40 (07)
  • [9] Montalvo TR, 2005, MATER SCI FORUM, V480, P373, DOI 10.4028/www.scientific.net/MSF.480-481.373
  • [10] Pagonis V., 2006, Numerical and Practical Exercises Thermoluminescence