Evaluation of the eazyplex MRSA assay for the rapid detection of Staphylococcus aureus in pleural and synovial fluid

被引:10
|
作者
Henares, D. [1 ]
Brotons, P. [1 ,2 ]
Buyse, X. [1 ]
Latorre, I. [1 ,3 ,4 ]
de Paz, H. D. [1 ]
Munoz-Almagro, C. [1 ,2 ,5 ]
机构
[1] Hosp St Joan de Deu, Dept Microbiol Mol, Inst Recerca Pediat, Barcelona, Spain
[2] CIBER Epidemiol & Salud Publ CIBERESP, Barcelona, Spain
[3] Inst Invest Germans Trias & Pujol, Badalona, Spain
[4] CIBER Enfermedades Resp CIBERES, Barcelona, Spain
[5] Univ Int Catalunya, Sch Med, Barcelona, Spain
关键词
MSSA; MRSA; Pleural and synovial fluid; eazyplex MRSA; Performance evaluation; MEDIATED ISOTHERMAL AMPLIFICATION; METHICILLIN-RESISTANT; GENE;
D O I
10.1016/j.ijid.2017.04.013
中图分类号
R51 [传染病];
学科分类号
100401 ;
摘要
Objective: Loop-mediated isothermal amplification (LAMP) is a simple and sensitive technique for rapid microbiological diagnosis. The aim of this study was to evaluate the analytical and diagnostic performance of the LAMP eazyplex MRSA test for the direct detection and differentiation of methicillin-susceptible (MSSA) and methicillin-resistant Staphylococcus aureus (MRSA) in synovial/pleural fluid. Methods: Analytical validation included the determination of the limit of detection (LoD) and analytical specificity of the eazyplex MRSA test. A diagnostic evaluation of the eazyplex test against bacterial culture was performed on routine pleural/synovial samples collected prospectively from patients aged less than 18 years with complicated pneumonia with empyema or arthritis admitted to the Children's Hospital Sant Joan de Deu in Barcelona, Spain, between April 2015 and May 2016. Results: The new system appropriately detected a quality control panel of clinical samples with DNA of MSSA, MRSA, and other pathogens. The LoD was established at 6.4 x 10(3) CFU/ml for S. aureus and 1.0 x 10(4) CFU/ml for MRSA. Diagnostic validation of the eazyplex MRSA assay was performed on 51 prospective clinical invasive samples, resulting in sensitivity and specificity values of 83.3% and 97.8%, respectively, for S. aureus detection. The mean turnaround time was 70 min. Conclusions: The eazyplex MRSA assay was found to be a useful test for the rapid detection of S. aureus in invasive samples such as pleural/synovial fluid. (C) 2017 The Authors. Published by Elsevier Ltd on behalf of International Society for Infectious Diseases.
引用
收藏
页码:65 / 68
页数:4
相关论文
共 50 条
  • [31] Rapid Detection of Panton-Valentine Leukocidin in Staphylococcus aureus Cultures by Use of a Lateral Flow Assay Based on Monoclonal Antibodies
    Monecke, Stefan
    Mueller, Elke
    Buechler, Joseph
    Rejman, John
    Stieber, Bettina
    Akpaka, Patrick Eberechi
    Bandt, Dirk
    Burris, Rob
    Coombs, Geoffrey
    Hidalgo-Arroyo, G. Aida
    Hughes, Peter
    Kearns, Angela
    Molinos Abos, Sonia
    Pichon, Bruno
    Skakni, Leila
    Soderquist, Bo
    Ehricht, Ralf
    JOURNAL OF CLINICAL MICROBIOLOGY, 2013, 51 (02) : 487 - 495
  • [32] MOLECULAR EVALUATION OF METHICILLIN RESISTANT STAPHYLOCOCCUS AUREUS (MRSA) BY PCR TO BACTERIOLOGICAL METHODS
    Gholamhoseinpor, Farnosh
    Navaderi, Mohammad
    Mirzaei, Mehdi
    IRANIAN JOURNAL OF PUBLIC HEALTH, 2014, 43 : 237 - 237
  • [33] Evaluation of the BD GeneOhm StaphSR assay for detection of Staphylococcus aureus in patients in intensive care units
    Ho, Tai-Hua
    Huang, Yhu-Chering
    Lin, Tzou-Yien
    JOURNAL OF MICROBIOLOGY IMMUNOLOGY AND INFECTION, 2011, 44 (04) : 310 - 315
  • [34] Development of a rapid assay methodology for antimicrobial susceptibility testing of Staphylococcus aureus
    Koskinen, Janne O.
    Stenholm, Teppo
    Vaarno, Jonne
    Soukka, Jori
    Meltola, Niko J.
    Soini, Aleksi E.
    DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2008, 62 (03) : 306 - 316
  • [35] Evaluation of a Rapid Culture-Based Screening Test for Detection of Methicillin resistant Staphylococcus aureus
    Fwity, Boushra
    Lobmann, Ralf
    Ambrosch, Andreas
    POLISH JOURNAL OF MICROBIOLOGY, 2011, 60 (03) : 265 - 268
  • [36] Molecular Detection of Methicillin-resistant Staphylococcus aureus Isolated From Foods in Germany Using LAMP Assay
    Sheet, Omar Hashim
    Al-Mahmood, Omar Ahmed
    Taha, Ayman Hani
    Alsanjary, Raad Abdulghany
    Ploetz, Madeleine
    Abdulmawjood, Amir Abdulhak
    EGYPTIAN JOURNAL OF VETERINARY SCIENCE, 2025, 56 (06): : 1153 - 1160
  • [37] Rapid and simple detection of methicillin-resistance staphylococcus aureus by orfXloop-mediated isothermal amplification assay
    Jianyu Su
    Xiaochen Liu
    Hemiao Cui
    Yanyan Li
    Dingqiang Chen
    Yanmei Li
    Guangchao Yu
    BMC Biotechnology, 14
  • [38] Rapid and low-cost biosensor for the detection of Staphylococcus aureus
    Suaifan, Ghadeer A. R. Y.
    Alhogail, Sahar
    Zourob, Mohammed
    BIOSENSORS & BIOELECTRONICS, 2017, 90 : 230 - 237
  • [39] PCR based detection of nosocomial infection causing MRSA (Methicillin resistant Staphylococcus aureus)
    Iyer, Archana Panchapakesan
    Kumosani, Taha Abdulah
    BIOTECHNOLOGY AND FOOD SERVICE, 2011, 7 : 1 - 3
  • [40] Comparison of cefoxitin disc diffusion test, HiCrome rapid MRSA agar and PCR for mecA, PVL and femB genes for detection of methicillin-resistant Staphylococcus aureus (MRSA)
    Wijekoon, Sineth Jeewantha
    Abeygoonawardena, Harshi
    Karawgoda, Ridmy Amaya
    Chandrasiri, Nambage Shirani
    Senanayake, Nilanthi Priyadarshika
    CEYLON MEDICAL JOURNAL, 2023, 68 (02) : 81 - 84