Maximizing Oxygen Delivery During Mechanical Ventilation With a Portable Oxygen Concentrator

被引:17
作者
Rodriquez, Dario, Jr. [1 ]
Blakeman, Thomas C. [2 ]
Dorlac, Warren [1 ]
Johannigman, Jay A. [2 ]
Branson, Richard D. [2 ]
机构
[1] USAF, C STARS, Cincinnati, OH USA
[2] Univ Cincinnati, Cincinnati, OH USA
来源
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE | 2010年 / 69卷
关键词
Mechanical ventilation; Inspired oxygen; Oxygen delivery; Low flow oxygen; Oxygen concentrator; Tidal volume; Transport;
D O I
10.1097/TA.0b013e3181e44b27
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: Transportation of the critically ill or injured war fighter requires the coordinated care and judicious use of resources. Availability of oxygen (O-2) supplies for the mechanically ventilated patient is crucial. Size and weight of cylinders makes transport difficult and presents an increased risk of fire. A proposed solution is to use a portable oxygen concentrator (POC) for mechanical ventilation. We tested the SeQual Eclipse II POC paired with the Impact 754 and Pulmonetics LTV-1200 ventilators in the laboratory and evaluated the fraction of inspired oxygen (FIO2) across a range of minute volumes. Methods: Each ventilator was attached to a test lung and pressure, volume, flow, and inspired oxygen (FIO2) was measured by a gas or flow analyzer. Ventilators were tested at a tidal volume (V-T) of 500 mL; an inspiratory time of 1.0 second; respiratory rates of 10, 20, and 30 breaths per minute; and positive end-expiratory pressure of 0 and 10 cm H2O. The LTV 1200 was tested with and without the expiratory bias flow. The Eclipse II was modified to provide pulse dosing on inspiration at 3 volumes (64, 128, and 192 mL) and continuous flow at 1 L/min to 3 L/min. Six combinations of ventilator settings were used with each POC setting for evaluation. O-2 was injected at the ventilator gas outlet and patient y-piece for pulse dose and continuous flow. Additionally, continuous flow O-2 was injected into the oxygen inlet port of the LTV 1200, and a reservoir bag, on the inlet port of the Impact 754. All tests were done with both ventilators using continuous flow, wall source O-2 as a control. We also measured the FIO2 with the concentrator on the highest pulse dose setting while decreasing ventilator V-T to compensate for the added volume. Results: The delivered FIO2 was highest when oxygen was injected into the ventilator circuit at the patient y-piece using pulse dosing, with the VT corrected. The next highest FIO2 was with continuous flow at the inlet (LTV), and reservoir (Impact). Electrical power consumption was less during pulse dose operation. Summary: Oxygen is a finite resource, which is cumbersome to transport and may present a fire hazard. The relatively high FIO2 delivered by the POC makes this method of O-2 delivery a viable alternative to O-2 cylinders. However, patients requiring an FIO2 of 1.0 would require additional compressed oxygen. This system allows O-2 delivery up to 76% solely using electricity. An integrated ventilator or POC capable of automatically compensating V-T for POC output is desirable. Further patient testing needs to be done to validate these laboratory findings.
引用
收藏
页码:S87 / S93
页数:7
相关论文
共 9 条
[1]  
[Anonymous], 2000, N ENGL J MED, V342, P1301
[2]   Work of breathing characteristics of seven portable ventilators [J].
Austin, PN ;
Campbell, RS ;
Johannigman, JA ;
Branson, RD .
RESUSCITATION, 2001, 49 (02) :159-167
[3]  
AUSTIN PN, 2002, RESP CARE CLIN N AM, V8, P1
[4]   En-route care in the air: Snapshot of mechanical ventilation at 37,000 feet [J].
Barnes, Stephen L. ;
Branson, Richard ;
Gallo, Louis A. ;
Beck, George ;
Johannigman, Jay A. .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2008, 64 (02) :S129-S134
[5]  
BRANSON R, 1998, RESP CARE EQUIPMENT
[6]  
BRANSON RD, 2008, EVALUATION PORTABLE
[7]   Comparison of volume control and pressure control ventilation: Is flow waveform the differences [J].
Davis, K ;
Branson, RD ;
Campbell, RS ;
Porembka, DT .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 1996, 41 (05) :808-814
[8]   Autonomous control of oxygenation [J].
Johannigman, Jay A. ;
Muskat, Peter ;
Barnes, Stephen ;
Davis, Kenneth, Jr. ;
Beck, George ;
Branson, Richard D. .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2008, 64 (04) :S295-S301
[9]  
Kacmarek R M, 2000, Respir Care, V45, P84