Morphological and chemical characterizations of the interface of a hydroxyapatite-coated implant

被引:18
|
作者
Mimura, K
Watanabe, K
Okawa, S
Kobayashi, M
Miyakawa, O
机构
[1] Mimura Dent Clin, Shizuoka 4180114, Japan
[2] Niigata Univ, Grad Sch Med & Dent Sci, Div Dent Biomat Sci, Niigata 9518514, Japan
[3] Niigata Univ, Ctr Instrumental Anal, EMX Lab, Niigata 9518514, Japan
关键词
hydroxyapatite-coated implant; interface; bonding;
D O I
10.4012/dmj.23.353
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The present study aimed at morphological and chemical characterization of the coating-substrate interface of a commercially available dental implant coated with plasma-sprayed hydroxyapatite (HA). For this purpose, elements in the chemically and mechanically exposed substrate surfaces were analyzed by EPMA and XPS. A thin titanium oxide film containing Ca and P was found at the interface. When the implant was subjected to mechanical stress, a mixed mode of cohesive and interfacial fractures occurred. The cohesive fracture was due to separation of the oxide film from the substrate, while the interfacial fracture was due to exfoliation of the coating from the oxide film bonded to the substrate. Analysis showed diffusion of Ca into the metal substrate, hence indicating the presence of chemical bond at the interface. However, mechanical interlocking seemed to play the major role in the interfacial bond.
引用
收藏
页码:353 / 360
页数:8
相关论文
共 50 条
  • [21] Bone-hydroxyapatite interface in retrieved hydroxyapatite-coated titanium implants: A clinical and histologic report
    Piattelli, A
    Scarano, A
    Di Alberti, L
    Piattelli, M
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 1999, 14 (02) : 233 - 238
  • [22] Ultrastructural and biophysical studies of the bone tissue surrounding a hydroxyapatite-coated implant: Case report
    Tanaka, N
    Ichinose, S
    Suyama, T
    Kohama, G
    JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2000, 58 (06) : 685 - 690
  • [23] 36-month analysis of a hydroxyapatite-coated implant prospective clinical trial.
    McGlumphy, EA
    Larsen, PE
    Peterson, LJ
    Scheffer, DB
    JOURNAL OF DENTAL RESEARCH, 1997, 76 : 81 - 81
  • [24] Variability of hydroxyapatite-coated dental implants
    Gross, KA
    Berndt, CC
    Iacono, VJ
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 1998, 13 (05) : 601 - 610
  • [25] Hydroxyapatite-coated external fixation pins
    Moroni, A
    Pegreffi, F
    Cadossi, M
    Hoang-Kim, A
    Lio, V
    Giannini, S
    EXPERT REVIEW OF MEDICAL DEVICES, 2005, 2 (04) : 465 - 471
  • [26] HYDROXYAPATITE-COATED TOTAL HIP PROSTHESES
    WICKI, O
    MIKIC, S
    GERBER, P
    HELVETICA CHIRURGICA ACTA, 1990, 57 (01) : 107 - 115
  • [27] BIOCOMPATIBILITY OF HYDROXYAPATITE-COATED HIP PROSTHESES
    SANTAVIRTA, S
    NORDSTROM, D
    YLINEN, P
    KONTTINEN, YT
    SILVENNOINEN, T
    ROKKANEN, P
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1991, 110 (06) : 288 - 292
  • [28] Osseointegrated Hearing Implant Surgery Using a Novel Hydroxyapatite-Coated Concave Abutment Design
    Wilkie, Mark D.
    Chakravarthy, Kalyan M.
    Mamais, Constantinos
    Temple, Robert H.
    OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2014, 151 (06) : 1014 - 1019
  • [29] Periprosthetic Fractures Around a Cementless Hydroxyapatite-coated Implant: A New Fracture Pattern Is Described
    Capello, William N.
    D'Antonio, James A.
    Naughton, Marybeth
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2014, 472 (02) : 604 - 610
  • [30] The Soft Tissue Immunologic Response to Hydroxyapatite-Coated Transmucosal Implant Surfaces: A Study in Humans
    De Wilde, Elisabeth A. W. J.
    Jimbo, Ryo
    Wennerberg, Ann
    Naito, Yoshihito
    Coucke, Paul
    Bryington, Matthew S.
    Vandeweghe, Stefan
    De Bruyn, Hugo
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2015, 17 : E65 - E74