Low-power transceiver analog front-end circuits for bidirectional high data rate wireless telemetry in medical endoscopy applications

被引:52
|
作者
Chi, Baoyong [1 ]
Yao, Jinke
Han, Shuguang
Xie, Xiang
Li, Guolin
Wang, Zhihua
机构
[1] Tsinghua Univ, Inst Microelect, Beijing 100084, Peoples R China
[2] Tsinghua Univ, Dept Elect Engn, Beijing 100084, Peoples R China
[3] Tsinghua Univ, Inst Microelect, Beijing 100084, Peoples R China
基金
中国国家自然科学基金;
关键词
AGC; ASK modulator/demodulator; CMOS analog integrated circuit; endoscopy capsule system; low-power IC; RF integrated circuit; wireless telemetry;
D O I
10.1109/TBME.2006.889768
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
State-of-the-art endoscopy systems require electronics allowing for real-time, bidirectional data transfer. Proposed are 2.4-GHz low-power transceiver analog front-end circuits for bidirectional high data rate wireless telemetry in medical endoscopy applications. The prototype integrates a low-IF receiver analog front-end [low noise amplifier (LNA), double balanced down-converter, bandpass-filtered automatic gain controlled (AGC) loop and amplitude-shift keying (ASK) demodulator], and a direct up-conversion transmitter analog front-end [20-MHz IF phase-locked loop (PLL) with well-defined amplitude control circuit, ASK modulator, up-converter, and power amplifier] on a single chip together with an internal radio frequency oscillator and local oscillating (LO) buffers. Design tradeoffs have been made over the boundaries of the different building blocks to optimize the overall system performance. All building blocks feature circuit topologies that enable comfortable operation at low power consumption. The circuits have been implemented in a 0.25-mu m CMOS process. The measured sensitivity of the receiver analog front-end is -70 dBm with a data rate of 256 kbps, and the measured output power of the transmitter analog front-end could achieve -23 dBm with a data rate of 1 Mbps. The integrated circuit consumes a current of 6 mA in receiver mode and 5.6 mA in transmitter mode with a power supply of 2.5 V. This paper shows the feasibility of achieving the analog performance required by the wireless endoscopy capsule system in 0.25 mu m CMOS.
引用
收藏
页码:1291 / 1299
页数:9
相关论文
共 50 条
  • [1] A 2.4 GHz low power wireless transceiver analog front-end for endoscopy capsule system
    Baoyong Chi
    Jinke Yao
    Shuguang Han
    Xiang Xie
    Guolin Li
    Zhihua Wang
    Analog Integrated Circuits and Signal Processing, 2007, 51 : 59 - 71
  • [2] A 2.4 GHz low power wireless transceiver analog front-end for endoscopy capsule system
    Chi, Baoyong
    Yao, Jinke
    Han, Shuguang
    Xie, Xiang
    Li, Guolin
    Wang, Zhihua
    ANALOG INTEGRATED CIRCUITS AND SIGNAL PROCESSING, 2007, 51 (02) : 59 - 71
  • [3] A 2.4GHz low power wireless transceiver analog front-end for endoscopy capsule system
    Chi, Baoyong
    Yao, Jinke
    Han, Shuguang
    Xie, Xiang
    Li, Guolin
    Wang, Zhihua
    2006 IEEE INTERNATIONAL SYMPOSIUM ON CIRCUITS AND SYSTEMS, VOLS 1-11, PROCEEDINGS, 2006, : 4907 - +
  • [4] Low-power CMOS analog front-end for wireless communication
    Taha, AA
    Abo-El-Soud, MA
    AbedRassoul, RA
    Farrag, AK
    2002 IEEE PROCEEDINGS OF THE NINETEENTH NATIONAL RADIO SCIENCE CONFERENCE, VOLS 1 AND 2, 2002, : 467 - 474
  • [5] Low-Power Ultrawideband Wireless Telemetry Transceiver for Medical Sensor Applications
    Gao, Yuan
    Zheng, Yuanjin
    Diao, Shengxi
    Toh, Wei-Da
    Ang, Chyuen-Wei
    Je, Minkyu
    Heng, Chun-Huat
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2011, 58 (03) : 768 - 772
  • [6] Low power high data rate wireless endoscopy transceiver
    Chi, Baoyong
    Yao, Jinke
    Han, Shuguang
    Xie, Xiang
    Li, Guolin
    Wang, Zhihua
    MICROELECTRONICS JOURNAL, 2007, 38 (10-11) : 1070 - 1081
  • [7] A 2.4-GHz low-power superregenerative RF front-end for high data rate applications
    Moncunill-Geniz, F. Xavier
    Dehollain, Catherine
    Joehl, Norbert
    Declercq, Michel
    Pala-Schoenwaelder, Pere
    2006 EUROPEAN MICROWAVE CONFERENCE, VOLS 1-4, 2006, : 1047 - +
  • [8] A Low-Power High-Sensitivity Analog Front-End for PPG Sensor
    Lin, Binghui
    Atef, Mohamed
    Wang, Guoxing
    2017 39TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2017, : 861 - 864
  • [9] A Fully-Integrated Low-Power Analog Front-End for ZigBee Transmitter Applications
    Shi Zuochen
    Yang Yintang
    Li Di
    Liu Yang
    CHINESE JOURNAL OF ELECTRONICS, 2016, 25 (03) : 424 - 431
  • [10] A Fully-Integrated Low-Power Analog Front-End for ZigBee Transmitter Applications
    SHI Zuochen
    YANG Yintang
    LI Di
    LIU Yang
    Chinese Journal of Electronics, 2016, 25 (03) : 424 - 431