Uncertainty of measurement and validation of a ballistic method by RP-HPLC for determination of rhEPO in 2000, 4000 and 10000 IU pharmaceutical formulations

被引:0
作者
Alves Premazzi, Melissa Chamon [1 ]
Gomes de Castro, Eduardo da Silva [2 ]
机构
[1] Oswaldo Cruz Fdn FIOCRUZ, Biomanguinhos, Rio De Janeiro, RJ, Brazil
[2] IFRJ, Inst Fed Educ Ciencia & Tecnol Rio De Janeiro, Rio De Janeiro, RJ, Brazil
关键词
HPLC; Erythropoietin; Measurement uncertainty; Method validation; MASS-SPECTROMETRY; ERYTHROPOIETIN; CHROMATOGRAPHY; CARBOHYDRATE;
D O I
10.1016/j.microc.2022.107418
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A method by reverse phase High Performance Liquid Chromatography with spectrophotometric detection on UV range (HPLC-UV) simple and fast was developed and validated for determination of recombinant human Erythropoietin (rhEPO) in presence of human serum albumin (HSA) as stabilizer in three different presentations; 2000, 4000 and 10,000 international units (IU). A ballistic gradient was obtained for separation of both proteins of biopharmaceuticals at time lower than 5 min. The solvent system comprised: solvent A - deionized water (0.1% trifluoracetic acid v/v), solvent B - acetonitrile (0.05% trifluoracetic acid v/v). The stationary phase was an octyl silane, C8, analytical column from J. T. Baker (R). The flow rate was 1.5 mL min-1, while the column and autosampler temperatures was kept at 60 and 5 ?, respectively. The injection volume was 100 mu L and the wavelength monitored by detector was 214 nm. The method has been validated according to the following analytical figures of merit: selectivity, trueness, linearity, precision (repeatability and intermediate precision) and limits of detection and quantification. Moreover, additional studies as stability of sample solution, peak purity test and uncertainty of measurement were assessed. The present study used a reference standard of Erythropoietin (BRP) supplied by European Pharmacopoeia in addition to in-house reference material of Eryth-ropoietin produced in collaboration between Bio-Manguinhos (FIOCRUZ/BRASIL), The National Institute of Quality Control in Health, INCQS (FIOCRUZ/BRASIL) and The Center of Molecular Immunology, CIM (CUBA).
引用
收藏
页数:12
相关论文
共 27 条
[1]   Recombinant protein expression for therapeutic applications [J].
Andersen, DC ;
Krummen, L .
CURRENT OPINION IN BIOTECHNOLOGY, 2002, 13 (02) :117-123
[2]  
[Anonymous], ISO 5725 ACCURACY TR
[3]  
[Anonymous], 2012, JCGM
[4]  
[Anonymous], 1993, GUID EXPR UNC MEAS
[5]  
[Anonymous], 352017 ISO
[6]  
BIPM IEC IFCC ILAC ISO IUPAC IUPAP OIML, 2008, Guide to the Expression of Uncertainty in Measurement
[7]   AN ANALYSIS OF TRANSFORMATIONS [J].
BOX, GEP ;
COX, DR .
JOURNAL OF THE ROYAL STATISTICAL SOCIETY SERIES B-STATISTICAL METHODOLOGY, 1964, 26 (02) :211-252
[8]  
DAMM J. B. L., 1998, APPL GLYCOBIOLOGY GE, V2
[9]   Validation spreadsheet: A new tool for estimating the analytical figures of merit for the validation of univariate methods [J].
de Lima Ribeiro, Fabiana Alves ;
Castro Ferreira, Marcia Miguel ;
Morano, Sandra Campilongo ;
da Silva, Lucimara Rodrigues ;
Schneider, Rene Peter .
QUIMICA NOVA, 2008, 31 (01) :164-171
[10]   CRITICAL METROLOGICAL EVALUATION OF FUEL ANALYSES BY MEASUREMENT UNCERTAINTY [J].
de Oliveira, Elcio Cruz .