Position tracking of moving liver lesion based on real-time registration between 2D ultrasound and 3D preoperative images

被引:14
作者
Weon, Chijun [1 ]
Nam, Woo Hyun [1 ]
Lee, Duhgoon [1 ]
Lee, Jae Young [2 ]
Ra, Jong Beom [1 ]
机构
[1] Korea Adv Inst Sci & Technol, Dept Elect Engn, Taejon 305701, South Korea
[2] Seoul Natl Univ Hosp, Dept Radiol, Seoul 110744, South Korea
关键词
ultrasound; 4D preoperative image; image-guided intervention; real-time registration; lesion tracking; NONRIGID REGISTRATION; CT IMAGES; INTENSITY; MOTION;
D O I
10.1118/1.4903945
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Registration between 2D ultrasound (US) and 3D preoperative magnetic resonance (MR) (or computed tomography, CT) images has been studied recently for US-guided intervention. However, the existing techniques have some limits, either in the registration speed or the performance. The purpose of this work is to develop a real-time and fully automatic registration system between two intermodal images of the liver, and subsequently an indirect lesion positioning/tracking algorithm based on the registration result, for image-guided interventions. Methods: The proposed position tracking system consists of three stages. In the preoperative stage, the authors acquire several 3D preoperative MR (or CT) images at different respiratory phases. Based on the transformations obtained from nonrigid registration of the acquired 3D images, they then generate a 4D preoperative image along the respiratory phase. In the intraoperative preparatory stage, they properly attach a 3D US transducer to the patient's body and fix its pose using a holding mechanism. They then acquire a couple of respiratory-controlled 3D US images. Via the rigid registration of these US images to the 3D preoperative images in the 4D image, the pose information of the fixed-pose 3D US transducer is determined with respect to the preoperative image coordinates. As feature(s) to use for the rigid registration, they may choose either internal liver vessels or the inferior vena cava. Since the latter is especially useful in patients with a diffuse liver disease, the authors newly propose using it. In the intraoperative real-time stage, they acquire 2D US images in real-time from the fixed-pose transducer. For each US image, they select candidates for its corresponding 2D preoperative slice from the 4D preoperative MR (or CT) image, based on the predetermined pose information of the transducer. The correct corresponding image is then found among those candidates via real-time 2D registration based on a gradient-based similarity measure. Finally, if needed, they obtain the position information of the liver lesion using the 3D preoperative image to which the registered 2D preoperative slice belongs. Results: The proposed method was applied to 23 clinical datasets and quantitative evaluations were conducted. With the exception of one clinical dataset that included US images of extremely low quality, 22 datasets of various liver status were successfully applied in the evaluation. Experimental results showed that the registration error between the anatomical features of US and preoperative MR images is less than 3 mm on average. The lesion tracking error was also found to be less than 5 mm at maximum. Conclusions: A new system has been proposed for real-time registration between 2D US and successive multiple 3D preoperative MR/CT images of the liver and was applied for indirect lesion tracking for image-guided intervention. The system is fully automatic and robust even with images that had low quality due to patient status. Through visual examinations and quantitative evaluations, it was verified that the proposed system can provide high lesion tracking accuracy as well as high registration accuracy, at performance levels which were acceptable for various clinical applications. (c) 2015 American Association of Physicists in Medicine.
引用
收藏
页码:335 / 347
页数:13
相关论文
共 50 条
  • [1] POSITION ESTIMATION OF MOVING LIVER LESION BASED ON REGISTRATION BETWEEN 2D ULTRASOUND AND 4D MR IMAGES
    Weon, Chi Jun
    Nam, Woo Hyun
    Lee, Duhgoon
    Hwang, Youngkyoo
    Kim, Jung-Bae
    Bang, Won-Chul
    Ra, Jong Beom
    2012 IEEE INTERNATIONAL CONFERENCE ON IMAGE PROCESSING (ICIP 2012), 2012, : 1677 - 1680
  • [2] SENSORLESS AND REAL-TIME REGISTRATION BETWEEN 2D ULTRASOUND AND PREOPERATIVE IMAGES OF THE LIVER
    Lee, Duhgoon
    Nam, Woo Hyun
    Hyun, Donggyu
    Lee, Jae Young
    Ra, Jong Beom
    2010 7TH IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING: FROM NANO TO MACRO, 2010, : 388 - 391
  • [3] 2D/3D Deep Registration for Real-Time Prostate Biopsy Navigation
    Dupuy, Tamara
    Beitone, Clement
    Troccaz, Jocelyne
    Voros, Sandrine
    MEDICAL IMAGING 2021: IMAGE-GUIDED PROCEDURES, ROBOTIC INTERVENTIONS, AND MODELING, 2021, 11598
  • [4] REAL-TIME 2D/3D REGISTRATION VIA CNN REGRESSION
    Miao, Shun
    Wang, Z. Jane
    Zheng, Yefeng
    Liao, Rui
    2016 IEEE 13TH INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING (ISBI), 2016, : 1430 - 1434
  • [5] A CNN Regression Approach for Real-Time 2D/3D Registration
    Miao, Shun
    Wang, Z. Jane
    Liao, Rui
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2016, 35 (05) : 1352 - 1363
  • [6] Real-time Moving Organ Tracking in Ultrasound Video Based on a 3D Organ Model
    Kim, Jung-Bae
    Hwang, Youngkyoo
    Bang, Won-Chul
    Kim, James D. K.
    Kim, ChangYeong
    2012 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2012, : 2659 - 2662
  • [7] Hybrid Feature-Based Diffeomorphic Registration for Tumor Tracking in 2-D Liver Ultrasound Images
    Cifor, Amalia
    Risser, Laurent
    Chung, Daniel
    Anderson, Ewan M.
    Schnabel, Julia A.
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2013, 32 (09) : 1647 - 1656
  • [8] Real-time motion tracking using 3D ultrasound
    Xu, Sheng
    Kruecker, Jochen
    Settlemier, Scott
    Wood, Bradford J.
    MEDICAL IMAGING 2007: VISUALIZATION AND IMAGE-GUIDED PROCEDURES, PTS 1 AND 2, 2007, 6509
  • [9] Real-time respiratory phase matching between 2D fluoroscopic images and 3D CT images for precise percutaneous lung biopsy
    Weon, Chijun
    Kim, Mina
    Park, Chang Min
    Ra, Jong Beom
    MEDICAL PHYSICS, 2017, 44 (11) : 5824 - 5834
  • [10] Rigid Registration of Untracked Freehand 2D Ultrasound Sweeps to 3D CT of Liver Tumours
    Cifor, Amalia
    Risser, Laurent
    Heinrich, Mattias P.
    Chung, Daniel
    Schnabel, Julia A.
    ABDOMINAL IMAGING: COMPUTATION AND CLINICAL APPLICATIONS, 2013, 8198 : 155 - 164