Assessment of QT-measurement accuracy using the 12-lead electrocardiogram derived from EASI leads

被引:15
作者
Martinez, Juan Pablo [1 ]
Laguna, Pablo
Olmos, Salvador
Pahlm, Olle
Pettersson, Jonas
Sornmo, Leif
机构
[1] Univ Zaragoza, Ctr Politecn Super, Dept Ingn Elect & Comunicac, Aragon Inst Engn Res, Aragon 50018, Spain
[2] Univ Lund Hosp, Dept Clin Physiol, S-22185 Lund, Sweden
[3] AstraZeneca R&D, Expt Med, Molndal, Sweden
[4] Lund Inst Technol, Dept Electrosci, Lund, Sweden
关键词
D O I
10.1016/j.jelectrocard.2006.08.089
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The purpose of the present study is to assess QT-interval measurements from the EASI 12-lead electrocardiogram (ECG) as compared with the standard 12-lead ECG. The QT interval was automatically determined in simultaneously recorded standard and EASI 12-lead ECGs, using a validated wavelet-based delineator. The agreement between the 2 sets of measurements was quantified both on a lead-by-lead basis and a multilead basis with global definitions of QRS onset and T-wave end. The results show that the agreement between QT-interval measurements from the 2 lead systems is acceptable, with negligible mean differences and with correlation coefficients ranging from 0.91 to 0.98 depending on the lead studied. Although the SD shows a clear dependence on the selected lead (ranging from 9.2 to 26.4 milliseconds), differences are within the accepted tolerances for automatic delineation. In a few patients, large differences were found, mainly because of changes in morphology present in both lead systems. QT intervals measured by the multilead approach were considerably more stable than single-lead measurements and resulted in a much better agreement between the 2 lead systems (correlation coefficient, 0.98; QT difference, 1.1 +/- 9.8 milliseconds). Thus, the EASI 12-lead ECG may be used for reliable QT monitoring when the multilead delineation approach is adopted. (c) 2007 Elsevier Inc. All rights reserved.
引用
收藏
页码:172 / 179
页数:8
相关论文
共 50 条
[21]   Detection of myocardial ischemia by vessel-specific leads derived from the 12-lead electrocardiogram and its subsets [J].
Horacek, B. Milan ;
Mirmoghisi, Maryam ;
Warren, James W. ;
Wagner, Galen S. ;
Wang, John J. .
JOURNAL OF ELECTROCARDIOLOGY, 2008, 41 (06) :508-517
[22]   The Earth is flat! The electrocardiogram has 12 leads! The electrocardiogram in the patient with ACS: looking beyond the 12-lead electrocardiogram [J].
Brady, William J. .
AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2007, 25 (09) :1073-1076
[23]   The spatial QRS-T angle in the Frank vectorcardiogram: accuracy of estimates derived from the 12-lead electrocardiogram [J].
Schreurs, Charlotte A. ;
Algra, Annemijn M. ;
Man, Sum-Che ;
Cannegieter, Suzanne C. ;
van der Wall, Ernst E. ;
Schalij, Martin J. ;
Kors, Jan A. ;
Swenne, Cees A. .
JOURNAL OF ELECTROCARDIOLOGY, 2010, 43 (04) :294-301
[24]   Feasibility of His bundle pacing facilitated by EASI derived 12-lead ECG [J].
Langanke, Alexander ;
Andreas, Klaus .
JOURNAL OF ELECTROCARDIOLOGY, 2023, 81 :272-276
[25]   Coefficients for the Derivation of Posterior and Right Sided Chest Leads From the 12-lead Electrocardiogram [J].
Jennings, Michael R. ;
Rababah, Ali S. ;
Biglarbeigi, Pardis ;
Brisk, Rob ;
Gueldenring, Daniel ;
Bond, Raymond ;
McLaughlin, James ;
Finlay, Dewar D. .
2020 COMPUTING IN CARDIOLOGY, 2020,
[26]   Eigen-Vector Based Leads for Reconstruction of the 12-Lead Electrocardiogram [J].
Finlay, D. D. ;
Nugent, C. D. ;
Donnelly, M. P. ;
Nelwan, S. P. .
COMPUTERS IN CARDIOLOGY 2008, VOLS 1 AND 2, 2008, :361-+
[27]   Derived 12-lead electrocardiogram in the assessment of ST-segment deviation and cardiac rhythm [J].
Chantad, D ;
Krittayaphong, R ;
Komoltri, C .
JOURNAL OF ELECTROCARDIOLOGY, 2006, 39 (01) :7-12
[28]   SENSITIVITY FOR TELEMED DIAGNOSIS OF MYOCARDIAL-INFARCTION BY USE OF 12-LEAD ELECTROCARDIOGRAM DERIVED FROM FRANK XYZ LEADS [J].
BRUCE, RA ;
BELANGER, L ;
BLACKMON, JR ;
TRIMBLE, S .
JOURNAL OF ELECTROCARDIOLOGY, 1982, 15 (02) :157-164
[29]   Measurement of Corrected QT Interval on Apple Watch Series 4 Electrocardiogram Has Moderate to Strong Agreement With 12-lead Electrocardiogram [J].
Aggarwal, Arjun ;
Saghir, Nabeel ;
Soneji, Nisha ;
Lopez, Suely Roman ;
Rodgers, George ;
Hong, Mauricio ;
Kurian, Thomas .
CIRCULATION, 2020, 142
[30]   Wearable 12-Lead ECG Acquisition Using a Novel Deep Learning Approach from Frank or EASI Leads with Clinical Validation [J].
Fu, Fan ;
Zhong, Dacheng ;
Liu, Jiamin ;
Xu, Tianxiang ;
Shen, Qin ;
Wang, Wei ;
Zhu, Songsheng ;
Li, Jianqing .
BIOENGINEERING-BASEL, 2024, 11 (03)