Assessment of QT-measurement accuracy using the 12-lead electrocardiogram derived from EASI leads

被引:15
|
作者
Martinez, Juan Pablo [1 ]
Laguna, Pablo
Olmos, Salvador
Pahlm, Olle
Pettersson, Jonas
Sornmo, Leif
机构
[1] Univ Zaragoza, Ctr Politecn Super, Dept Ingn Elect & Comunicac, Aragon Inst Engn Res, Aragon 50018, Spain
[2] Univ Lund Hosp, Dept Clin Physiol, S-22185 Lund, Sweden
[3] AstraZeneca R&D, Expt Med, Molndal, Sweden
[4] Lund Inst Technol, Dept Electrosci, Lund, Sweden
关键词
D O I
10.1016/j.jelectrocard.2006.08.089
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The purpose of the present study is to assess QT-interval measurements from the EASI 12-lead electrocardiogram (ECG) as compared with the standard 12-lead ECG. The QT interval was automatically determined in simultaneously recorded standard and EASI 12-lead ECGs, using a validated wavelet-based delineator. The agreement between the 2 sets of measurements was quantified both on a lead-by-lead basis and a multilead basis with global definitions of QRS onset and T-wave end. The results show that the agreement between QT-interval measurements from the 2 lead systems is acceptable, with negligible mean differences and with correlation coefficients ranging from 0.91 to 0.98 depending on the lead studied. Although the SD shows a clear dependence on the selected lead (ranging from 9.2 to 26.4 milliseconds), differences are within the accepted tolerances for automatic delineation. In a few patients, large differences were found, mainly because of changes in morphology present in both lead systems. QT intervals measured by the multilead approach were considerably more stable than single-lead measurements and resulted in a much better agreement between the 2 lead systems (correlation coefficient, 0.98; QT difference, 1.1 +/- 9.8 milliseconds). Thus, the EASI 12-lead ECG may be used for reliable QT monitoring when the multilead delineation approach is adopted. (c) 2007 Elsevier Inc. All rights reserved.
引用
收藏
页码:172 / 179
页数:8
相关论文
共 50 条
  • [1] Accuracy of QT measurement in the EASI-derived 12-lead ECG
    Martinez, Juan Pablo
    Laguna, Pablo
    Olmos, Salvador
    Pahlm, Olle
    Petersson, Jonas
    Sornmo, Leif
    2006 28TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-15, 2006, : 4192 - +
  • [2] Diagnostic conclusions from the EASI-derived 12-lead electrocardiogram as compared with the standard 12-lead electrocardiogram in children
    Welinder, A
    Feild, DQ
    Liebman, J
    Maynard, C
    Wagner, GS
    Wettrell, G
    Pahlm, O
    AMERICAN HEART JOURNAL, 2006, 151 (05) : 1059 - 1064
  • [3] Accuracy of the EASI 12-lead electrocardiogram compared to the standard 12-lead electrocardiogram for diagnosing multiple cardiac abnormalities
    Drew, BJ
    Pelter, MM
    Wung, SF
    Adams, MG
    Taylor, C
    Evans, GT
    Foster, E
    JOURNAL OF ELECTROCARDIOLOGY, 1999, 32 : 38 - 47
  • [4] Detection of acute ischemia from the EASI-derived 12-lead electrocardiogram and from the 12-lead electrocardiogram acquired in clinical practice
    Sejersten, Maria
    Wagner, Galen S.
    Pahlm, Olle
    Warren, James W.
    Feldman, Charles L.
    Horacek, B. Milan
    JOURNAL OF ELECTROCARDIOLOGY, 2007, 40 (02) : 120 - 126
  • [5] DERIVING THE 12-LEAD ELECTROCARDIOGRAM FROM 4 (EASI) ELECTRODES
    DOWER, GE
    YAKUSH, A
    NAZZAL, SB
    JUTZY, RV
    RUIZ, CE
    JOURNAL OF ELECTROCARDIOLOGY, 1988, 21 : S182 - S187
  • [6] Effects of electrode placement errors in the EASI-derived 12-lead electrocardiogram
    Finlay, Dewar D.
    Nugent, Chris D.
    Nelwan, Stefan P.
    Bond, Raymond R.
    Donnelly, Mark P.
    Guldenring, Daniel
    JOURNAL OF ELECTROCARDIOLOGY, 2010, 43 (06) : 606 - 611
  • [7] Is the EASI 12-lead electrocardiogram comparable to the standard 12-lead electrocardiogram for diagnosing cardiac rhythm and myocardial ischemia?
    Drew, BJ
    Pelter, MM
    Adams, MG
    Wung, SF
    Taylor, CA
    Banks, AS
    CIRCULATION, 1999, 100 (18) : 736 - 736
  • [8] Comparison of waveforms in conventional 12-lead ECGs and those derived from EASI leads in children
    Pahlm, O
    Pettersson, J
    Thulin, A
    Feldman, CL
    Feild, DQ
    Wagner, GS
    JOURNAL OF ELECTROCARDIOLOGY, 2003, 36 (01) : 25 - 31
  • [9] Synthesis of the 12-Lead Electrocardiogram From Differential Leads
    Trobec, Roman
    Tomasic, Ivan
    IEEE TRANSACTIONS ON INFORMATION TECHNOLOGY IN BIOMEDICINE, 2011, 15 (04): : 615 - 621
  • [10] COMPARISON OF AUTOMATIC QT MEASUREMENT TECHNIQUES IN THE NORMAL 12-LEAD ELECTROCARDIOGRAM
    MCLAUGHLIN, NB
    CAMPBELL, RWF
    MURRAY, A
    BRITISH HEART JOURNAL, 1995, 74 (01): : 84 - 89