Thermionic sodium ion source versus corona discharge in detection of alkaloids using ion mobility spectrometry

被引:4
作者
Asadi, Masood [1 ]
Valadbeigi, Younes [2 ]
Tabrizchi, Mahmoud [1 ]
机构
[1] Isfahan Univ Technol, Dept Chem, Esfahan 8415683111, Iran
[2] Imam Khomeini Int Univ, Fac Sci, Dept Chem, POB 288, Qazvin, Iran
关键词
Thermal ionization; Ion mobility spectrometry; Sodium ion; Alkaloids; Ionization source; IONIZATION SOURCE;
D O I
10.1007/s12127-019-00249-9
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
IMS spectra obtained by using a thermal ionization sodium ion (SI) source and a corona discharge (CD) source were compared in terms of resolution, selectivity and sensitivity. The reactant ion peak (RIP) for the SI source is a single narrow peak which appears in a shorter drift time compared to the main RIP of the CD ion source, (H2O)(n)H3O+. Comparison of the IMS spectra of selected alkaloids obtained with both the CD and the SI sources shows that the latter exhibits simpler pattern, including a narrow single peak for each compound, well separated from other samples peak. Furthermore, the product ion peaks (PIP) in the SI source appeared at higher drift times leading to higher peak-to-peak resolution. However, the sodium ion source is less sensitive to the selected alkaloids compared to the CD ion source.
引用
收藏
页码:51 / 58
页数:8
相关论文
共 41 条
[1]  
[Anonymous], 1992, P 1 WORKSH ION MOB S
[2]   Protonation of caffeine: A theoretical and experimental study [J].
Bahrami, Hamed ;
Tabrizchi, Mahmoud ;
Farrokhpour, Hossein .
CHEMICAL PHYSICS, 2013, 415 :222-227
[3]   ION MOBILITY DETECTOR FOR GAS-CHROMATOGRAPHY WITH A DIRECT PHOTO-IONIZATION SOURCE [J].
BAIM, MA ;
EATHERTON, RL ;
HILL, HH .
ANALYTICAL CHEMISTRY, 1983, 55 (11) :1761-1766
[4]   Detection of the gasoline components methyl tert-butyl ether, benzene, toluene, and m-xylene using ion mobility spectrometers with a radioactive and UV ionization source [J].
Baumbach, JI ;
Sielemann, S ;
Xie, ZY ;
Schmidt, H .
ANALYTICAL CHEMISTRY, 2003, 75 (06) :1483-1490
[5]   IONIZATION IN FLAMES .2. MASS-SPECTROMETRIC AND MOBILITY ANALYSES FOR FLAME IONIZATION DETECTOR [J].
BOLTON, HC ;
GRANT, J ;
MCWILLIAM, IG ;
NICHOLSON, AJC ;
SWINGLER, DL .
PROCEEDINGS OF THE ROYAL SOCIETY OF LONDON SERIES A-MATHEMATICAL PHYSICAL AND ENGINEERING SCIENCES, 1978, 360 (1701) :265-277
[6]  
Bramwell C.J., 2002, INT J ION MOBILITY S, V5, P87
[7]  
Clemmer DE, 1997, J MASS SPECTROM, V32, P577
[8]   Atmospheric. pressure air direct current glow discharge ionization source for ion mobility spectrometry [J].
Dong, Can ;
Wang, Weiguo ;
Li, Haiyang .
ANALYTICAL CHEMISTRY, 2008, 80 (10) :3925-3930
[9]  
Eiceman GA, 2014, ION MOBILITY SPECTROMETRY, 3RD EDITION, P1
[10]   QUANTITATIVE ASSESSMENT OF A CORONA DISCHARGE ION-SOURCE IN ATMOSPHERIC-PRESSURE IONIZATION-MASS SPECTROMETRY FOR AMBIENT AIR MONITORING [J].
EICEMAN, GA ;
KREMER, JH ;
SNYDER, AP ;
TOFFERI, JK .
INTERNATIONAL JOURNAL OF ENVIRONMENTAL ANALYTICAL CHEMISTRY, 1988, 33 (3-4) :161-183