A novel calibration for L-shell x-ray fluorescence measurements of bone lead concentration using the strontium Kβ/Kα ratio

被引:2
作者
Gherase, Mihai R. [1 ]
Serna, Blaz [1 ]
Kroeker, Sarah [1 ]
机构
[1] Calif State Univ Fresno, Dept Phys, Fresno, CA 93740 USA
基金
美国国家卫生研究院;
关键词
bone; x-ray fluorescence; lead; strontium; calibration; UNITED-STATES; COHERENT NORMALIZATION; NATIONAL-HEALTH; CHILDREN; EXPOSURE; TOXICITY; PHANTOMS; TUBE; PB; AGE;
D O I
10.1088/1361-6579/abf886
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Objective. Lead (Pb) is a well-known toxic element. In vivo bone Pb concentration measurement is a long-term exposure metric complementary to blood Pb concentration measurement which is a metric of recent exposure. In vivo human tibia bone Pb measurements using Pb K-shell or L-shell x-ray fluorescence (KXRF or LXRF) emissions were developed in the 1980s. KXRF bone Pb measurements using Cd-109 gamma rays and coherent-to-fluorescence ratio to account for differences between phantom and in vivo measurements, was employed in human studies. Bone Pb LXRF method employed x-ray tubes. However, calibration procedures using ultrasound measurements of the soft tissue thickness (STT) proved inaccurate. Approach. In this study, bone and soft tissue (ST) phantoms simulated in vivo bone Pb measurements. Seven plaster-of-Paris cylindrical bone phantoms containing 1.01 mg g(-1) of strontium (Sr) were doped with Pb in 0, 8, 16, 29, 44, 59, and 74 mu g g(-1) concentrations. Polyoxymethylene (POM), resin, and wax were each used to fabricate four ST phantoms in the approximate 1-4 mm thickness range. Pb LXRF measurements were performed using a previously developed optimal grazing incidence position method. Main results. Linear attenuation coefficients measurements of ST materials indicated that POM and resin mimicked well attenuation of Pb x-rays in skin and adipose tissue, respectively. POM and resin data indicated a bone Pb detection limit of 20 mu g g(-1) for a 2 mm STT. Derived relationships between the Pb concentration, Pb LXRF and Sr K- beta /K- alpha ratio data did not require STT knowledge. Applied to POM and resin data, the new calibration method yielded unbiased results. Significance. In vivo bone Pb measurements in children were suggested following considerations of radiation dose, STT, detectability and distribution of Pb and Sr in bone. This research meets with the concerns regarding the negative effects of low levels of Pb exposure on neurodevelopment of children.
引用
收藏
页数:29
相关论文
共 50 条
  • [21] In vivo quantification of strontium in bone among adults using portable x-ray fluorescence
    Zhang, Xinxin
    Wells, Ellen M.
    Specht, Aaron J.
    Weisskopf, Marc G.
    Weuve, Jennifer
    Nie, Linda H.
    JOURNAL OF TRACE ELEMENTS IN MEDICINE AND BIOLOGY, 2022, 74
  • [22] Monitoring bone strontium intake in osteoporotic females self-supplementing with strontium citrate with a novel in-vivo X-ray fluorescence based diagnostic tool
    Moise, Helen
    Chettle, David R.
    Pejovic-Milic, Ana
    BONE, 2014, 61 : 48 - 54
  • [23] Calculation of depth-dependent elemental concentration with X-ray fluorescence using a layered calibration method
    Gherase, M. R.
    Fleming, D. E. B.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2011, 269 (10) : 1150 - 1156
  • [24] Evaluating calcium phosphate glass phantoms for the calibration of in vivo X-ray fluorescence spectrometry-based methods of bone strontium quantification
    Micsa, Matthew
    Da Silva, Eric
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2023, 38 (10) : 2019 - 2027
  • [25] X-ray fluorescence Kα/Kβ ratios for a layered specimen: Comparison of measurements and Monte Carlo calculations with the MCNPX code
    Trojek, T.
    Wegrzynek, D.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 619 (1-3) : 311 - 315
  • [26] Bias correction by use of errors-in-variables regression models in studies with K-X-ray fluorescence bone lead measurements
    Lamadrid-Figuero, Hector
    Tellez-Rojo, Martha M.
    Angeles, Gustavo
    Hernandez-Avila, Mauricio
    Hu, Howard
    ENVIRONMENTAL RESEARCH, 2011, 111 (01) : 17 - 20
  • [27] Detection of lead in bone phantoms and arsenic in soft tissue phantoms using synchrotron radiation and a portable x-ray fluorescence system
    Groskopf, Craig
    Bennett, Stephen R.
    Gherase, Mihai R.
    Fleming, David E. B.
    PHYSIOLOGICAL MEASUREMENT, 2017, 38 (02) : 374 - 386
  • [28] Effects of measurement distance and source activity on the precision of X-ray fluorescence measurements of lead in a bone phantom
    Fleming, DEB
    Chettle, DR
    McNeill, FE
    Weedon, AGP
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2004, 217 (03) : 471 - 478
  • [29] The feasibility of in vivo detection of lanthanum using a 241Am K x-ray fluorescence system
    Joanna Nguyen
    Keldani, Zaid
    Da Silva, Eric
    Pejovic-Milic, Ana
    Grafe, James L.
    PHYSIOLOGICAL MEASUREMENT, 2017, 38 (09) : 1766 - 1775
  • [30] Measurements of L-shell X-ray production cross-sections of Ag and Sb by low-energy electron impact
    Zhao, J. L.
    An, Z.
    Zhu, J. J.
    Tan, W. J.
    Liu, M. T.
    RADIATION PHYSICS AND CHEMISTRY, 2016, 122 : 66 - 72