A novel calibration for L-shell x-ray fluorescence measurements of bone lead concentration using the strontium Kβ/Kα ratio

被引:2
作者
Gherase, Mihai R. [1 ]
Serna, Blaz [1 ]
Kroeker, Sarah [1 ]
机构
[1] Calif State Univ Fresno, Dept Phys, Fresno, CA 93740 USA
基金
美国国家卫生研究院;
关键词
bone; x-ray fluorescence; lead; strontium; calibration; UNITED-STATES; COHERENT NORMALIZATION; NATIONAL-HEALTH; CHILDREN; EXPOSURE; TOXICITY; PHANTOMS; TUBE; PB; AGE;
D O I
10.1088/1361-6579/abf886
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Objective. Lead (Pb) is a well-known toxic element. In vivo bone Pb concentration measurement is a long-term exposure metric complementary to blood Pb concentration measurement which is a metric of recent exposure. In vivo human tibia bone Pb measurements using Pb K-shell or L-shell x-ray fluorescence (KXRF or LXRF) emissions were developed in the 1980s. KXRF bone Pb measurements using Cd-109 gamma rays and coherent-to-fluorescence ratio to account for differences between phantom and in vivo measurements, was employed in human studies. Bone Pb LXRF method employed x-ray tubes. However, calibration procedures using ultrasound measurements of the soft tissue thickness (STT) proved inaccurate. Approach. In this study, bone and soft tissue (ST) phantoms simulated in vivo bone Pb measurements. Seven plaster-of-Paris cylindrical bone phantoms containing 1.01 mg g(-1) of strontium (Sr) were doped with Pb in 0, 8, 16, 29, 44, 59, and 74 mu g g(-1) concentrations. Polyoxymethylene (POM), resin, and wax were each used to fabricate four ST phantoms in the approximate 1-4 mm thickness range. Pb LXRF measurements were performed using a previously developed optimal grazing incidence position method. Main results. Linear attenuation coefficients measurements of ST materials indicated that POM and resin mimicked well attenuation of Pb x-rays in skin and adipose tissue, respectively. POM and resin data indicated a bone Pb detection limit of 20 mu g g(-1) for a 2 mm STT. Derived relationships between the Pb concentration, Pb LXRF and Sr K- beta /K- alpha ratio data did not require STT knowledge. Applied to POM and resin data, the new calibration method yielded unbiased results. Significance. In vivo bone Pb measurements in children were suggested following considerations of radiation dose, STT, detectability and distribution of Pb and Sr in bone. This research meets with the concerns regarding the negative effects of low levels of Pb exposure on neurodevelopment of children.
引用
收藏
页数:29
相关论文
共 50 条
  • [1] A microbeam grazing-incidence approach to L-shell x-ray fluorescence measurements of lead concentration in bone and soft tissue phantoms
    Gherase, Mihai Raul
    Al-Hamdani, Summer
    PHYSIOLOGICAL MEASUREMENT, 2018, 39 (03)
  • [2] Measurements of Strontium Levels in Human Bone In Vivo Using Portable X-ray Fluorescence (XRF)
    Specht, Aaron J.
    Mostafaei, Farshad
    Lin, Yanfen
    Xu, Jian
    Nie, Linda H.
    APPLIED SPECTROSCOPY, 2017, 71 (08) : 1962 - 1968
  • [3] A Two-Dimensional K-Shell X-ray Fluorescence (2D-KXRF) Model for Soft Tissue Attenuation Corrections of Strontium Measurements in a Cortical Lamb Bone Sample
    Gherase, Mihai R.
    METROLOGY, 2023, 3 (04): : 325 - 346
  • [4] Experimental demonstration of direct L-shell x-ray fluorescence imaging of gold nanoparticles using a benchtop x-ray source
    Manohar, Nivedh
    Reynoso, Francisco J.
    Cho, Sang Hyun
    MEDICAL PHYSICS, 2013, 40 (08)
  • [5] THE INFLUENCE OF BONE-DENSITY ON IN VIVO K X-RAY FLUORESCENCE BONE-LEAD MEASUREMENTS
    Lodwick, Camille J.
    Lodwick, Jeffrey C.
    Spitz, Henry B.
    HEALTH PHYSICS, 2011, 100 (05): : 502 - 507
  • [6] Calibration of the 125I-induced x-ray fluorescence spectrometry-based system of in vivo bone strontium determinations using hydroxyapatite as a phantom material: a simulation study
    Da Silva, Eric
    Pejovic-Milic, Ana
    PHYSIOLOGICAL MEASUREMENT, 2017, 38 (06) : 1077 - 1093
  • [7] Development of a K-shell x-ray fluorescence measurement of cadmium in bone
    Carew, SE
    Gastaldo, J
    Roels, HA
    O'Meara, JM
    Chettle, DR
    X-RAY SPECTROMETRY, 2005, 34 (06) : 498 - 501
  • [8] Pure Hydroxyapatite Phantoms for the Calibration of in Vivo X-ray Fluorescence Systems of Bone Lead and Strontium Quantification
    Da Silva, Eric
    Kirkham, Brian
    Heyd, Darrick V.
    Pejovic-Milic, Ana
    ANALYTICAL CHEMISTRY, 2013, 85 (19) : 9189 - 9195
  • [9] Portable X-ray fluorescence for bone lead measurements of Australian eagles
    Hampton, Jordan O.
    Specht, Aaron J.
    Pay, James M.
    Pokras, Mark A.
    Bengsen, Andrew J.
    SCIENCE OF THE TOTAL ENVIRONMENT, 2021, 789
  • [10] Calibration and characterization of a digital X-ray fluorescence bone lead system
    Fleming, DEB
    Forbes, TA
    APPLIED RADIATION AND ISOTOPES, 2001, 55 (04) : 527 - 532