Anti-rabbit immunoglobulin G detection in complex medium by PM-RAIRS and QCM Influence of the antibody immobilisation method

被引:38
作者
Briand, Elisabeth
Salmain, Michele
Compere, Chantal
Pradier, Claire-Marie
机构
[1] Univ Paris 06, CNRS, UMR 7609, Lab React Surface, F-75252 Paris 05, France
[2] Ecole Natl Super Chim Paris, UMR 7576, Lab Chim & Biochim Complexes Mol, F-75231 Paris 05, France
[3] IFREMER, Ctr Brest, Dept Essais & Rech Technol Interfaces & Capteurs, F-29280 Plouzane, France
关键词
immunosensors; QCM; PM-RAIRS; rIgG immobilisation;
D O I
10.1016/j.bios.2006.12.009
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Two antibody immobilisation procedures were compared to set up an immunosensor for goat anti-rabbit immunoglobulin (anti-rIgG), i.e. rIgG covalently bound or immobilised via affinity to protein A (PrA). In both cases, the first layer of protein was covalently bound to a mixed self-assembled monolayer (SAM) of mercaptoundecanoic acid (MUA) and mercaptohexanol (C6OH) on a gold surface. The elaboration of the sensitive surfaces, as well as their selectivity and sensitivity were studied step by step by polarization modulation-reflection absorption infra-red spectroscopy (PM-RAIRS) and quartz crystal trticrobalance (QCM) with impedance measurement. QCM measurements showed that the viscoelastic properties of the antibody layer were markedly modified during the antigen recognition when the antibody was bound by affinity to PrA. The specific detection of antigen within a complex medium was assessed by PM-RAIRS thanks to the grafting of cobalt-carbonyl probes. Affinity constants between the immobilised rIgG and the anti-rIgG were determined from PM-RAIRS analysis. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:2884 / 2890
页数:7
相关论文
共 61 条
[31]   Comparison of different protein immobilization methods on quartz crystal microbalance surface in flow injection immunoassay [J].
Liu, YC ;
Wang, CM ;
Hsiung, KP .
ANALYTICAL BIOCHEMISTRY, 2001, 299 (02) :130-135
[32]   Lysozyme interaction with poly(HEMA)-based hydrogel [J].
Lord, MS ;
Stenzel, MH ;
Simmons, A ;
Milthorpe, BK .
BIOMATERIALS, 2006, 27 (08) :1341-1345
[33]   Oriented immobilization of antibodies and its applications in immunoassays and immunosensors [J].
Lu, B ;
Smyth, MR ;
OKennedy, R .
ANALYST, 1996, 121 (03) :R29-R32
[34]   A reusable and specific protein A-coated piezoelectric biosensor for flow injection immunoassay [J].
Lu, HC ;
Chen, HM ;
Lin, YS ;
Lin, JW .
BIOTECHNOLOGY PROGRESS, 2000, 16 (01) :116-124
[35]   The quartz crystal microbalance: mass sensitivity, viscoelasticity and acoustic amplification [J].
Lucklum, R ;
Hauptmann, P .
SENSORS AND ACTUATORS B-CHEMICAL, 2000, 70 (1-3) :30-36
[36]   The Δf-ΔR QCM technique:: an approach to an advanced sensor signal interpretation [J].
Lucklum, R ;
Hauptmann, P .
ELECTROCHIMICA ACTA, 2000, 45 (22-23) :3907-3916
[37]   Study of adsorption and viscoelastic properties of proteins with a quartz crystal microbalance by measuring the oscillation amplitude [J].
Marxer, CG ;
Coen, AC ;
Schlapbach, L .
JOURNAL OF COLLOID AND INTERFACE SCIENCE, 2003, 261 (02) :291-298
[38]   In situ analysis of a mercaptoundecanoic acid layer on gold in liquid phase, by PM-IRAS.: Evidence for chemical changes with the solvent [J].
Méthivier, C ;
Beccard, B ;
Pradier, CM .
LANGMUIR, 2003, 19 (21) :8807-8812
[39]   Miniaturized QCM-based flow system for immunosensor application in liquid [J].
Michalzik, M ;
Wilke, R ;
Büttgenbach, S .
SENSORS AND ACTUATORS B-CHEMICAL, 2005, 111 :410-415
[40]  
MURAMATSU H, 1987, ANAL CHEM, V59