Audit feasibility for geometric distortion in magnetic resonance imaging for radiotherapy

被引:9
作者
Alzahrani, Meshal [1 ,2 ]
Broadbent, David A. [3 ]
Chuter, Robert [4 ,5 ]
Al-Qaisieh, Bashar [3 ]
Jackson, Steven [4 ]
Michael, Hutton [4 ]
Johnstone, Robert, I [6 ]
Shah, Simon [6 ]
Wetscherek, Andreas [7 ,8 ]
Chick, H. Joan [7 ,8 ]
Wyatt, Jonathan J. [9 ,10 ]
McCallum, Hazel Mhairi [9 ,10 ]
Speight, Richard [3 ]
机构
[1] King Abdulaziz Univ, Fac Appl Med Sci, Dept Diagnost Radiol, Jeddah, Saudi Arabia
[2] Univ Leeds, Leeds Inst Cardiovasc & Metab Med, Biomed Imaging Sci Dept, Leeds, W Yorkshire, England
[3] Leeds Teaching Hosp NHS Trust, Dept Med Phys & Engn, Leeds, W Yorkshire, England
[4] Christie NHS Fdn Trust, Christie Med Phys & Engn CMPE, Manchester, Lancs, England
[5] Univ Manchester, Fac Biol Med & Hlth, Div Canc Sci, Manchester, Lancs, England
[6] Guys & St Thomas NHS Fdn Trust, London, England
[7] Inst Canc Res, Joint Dept Phys, London, England
[8] Royal Marsden NHS Fdn Trust, London, England
[9] Newcastle Upon Tyne Hosp NHS Fdn Trust, Northern Ctr Canc Care, Newcastle Upon Tyne, Tyne & Wear, England
[10] Newcastle Univ, Ctr Canc, Newcastle Upon Tyne, Tyne & Wear, England
来源
PHYSICS & IMAGING IN RADIATION ONCOLOGY | 2020年 / 15卷
关键词
Magnetic resonance imaging; Geometric distortion; Radiotherapy; Quality assurance; Audit; RADIATION-THERAPY; MRI; PRECISION; ACCURACY; IMPACT;
D O I
10.1016/j.phro.2020.07.004
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: Magnetic Resonance Imaging (MRI) is increasingly being used in radiotherapy (RT). However, geometric distortions are a known challenge of using MRI in RT. The aim of this study was to demonstrate feasibility of a national audit of MRI geometric distortions. This was achieved by assessing large field of view (FOV) MRI distortions on a number of scanners used clinically for RT. Materials and methods: MRI scans of a large FOV MRI geometric distortion phantom were acquired on 11 MRI scanners that are used clinically for RT in the UK. The mean and maximum distortions and variance between scanners were reported at different distances from the isocentre. Results: For a small FOV representing a brain (100-150 mm from isocentre) all distortions were < 2 mm except for the maximum distortion of one scanner. For a large FOV representing a head and neck/pelvis (200-250 mm from isocentre) mean distortions were < 2 mm except for one scanner, maximum distortions were > 10 mm in some cases. The variance between scanners was low and was found to increase with distance from isocentre. Conclusions: This study demonstrated feasibility of the technique to be repeated in a country wide geometric distortion audit of all MRI scanners used clinically for RT. Recommendations were made for performing such an audit and how to derive acceptable limits of distortion in such an audit.
引用
收藏
页码:80 / 84
页数:5
相关论文
共 50 条
  • [31] Tumor-site specific geometric distortions in high field integrated magnetic resonance linear accelerator radiotherapy
    Hasler, Signe Winther
    Bernchou, Uffe
    Bertelsen, Anders
    van Veldhuizen, Elisabeth
    Schytte, Tine
    Hansen, Vibeke Nordmark
    Brink, Carsten
    Mahmood, Faisal
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2020, 15 : 100 - 104
  • [32] ON THE PROBLEM OF GEOMETRIC DISTORTION IN MAGNETIC-RESONANCE IMAGES FOR STEREOTAXIC NEUROSURGERY
    MICHIELS, J
    BOSMANS, H
    PELGRIMS, P
    VANDERMEULEN, D
    GYBELS, J
    MARCHAL, G
    SUETENS, P
    MAGNETIC RESONANCE IMAGING, 1994, 12 (05) : 749 - 765
  • [33] Feasibility of magnetic resonance guided radiotherapy for the treatment of bladder cancer
    Hunt, A.
    Hanson, I.
    Dunlop, A.
    Barnes, H.
    Bower, L.
    Chick, J.
    Cruickshank, C.
    Hall, E.
    Herbert, T.
    Lawes, R.
    McQuaid, D.
    McNair, H.
    Mitchell, A.
    Mohajer, J.
    Morgan, T.
    Oelfke, U.
    Smith, G.
    Nill, S.
    Huddart, R.
    Hafeez, S.
    CLINICAL AND TRANSLATIONAL RADIATION ONCOLOGY, 2020, 25 : 46 - 51
  • [34] Impact of Radiotherapy on Background Parenchymal Enhancement in Breast Magnetic Resonance Imaging
    Kim, Yun Ju
    Kim, Sung Hun
    Choi, Byung Gil
    Kang, Bong Joo
    Kim, Hyeon Sook
    Cha, Eun Suk
    Song, Byung Joo
    ASIAN PACIFIC JOURNAL OF CANCER PREVENTION, 2014, 15 (07) : 2939 - 2943
  • [35] Distortion correction of diffusion-weighted magnetic resonance imaging of the head and neck in radiotherapy position
    Winter, Rene M.
    Schmidt, Holger
    Leibfarth, Sara
    Zwirner, Kerstin
    Welz, Stefan
    Schwenzer, Nina F.
    la Fougere, Christian
    Nikolaou, Konstantin
    Gatidis, Sergios
    Zips, Daniel
    Thorwarth, Daniela
    ACTA ONCOLOGICA, 2017, 56 (11) : 1659 - 1663
  • [36] Rapid distortion correction enables accurate magnetic resonance imaging-guided real-time adaptive radiotherapy
    Liu, Paul Z. Y.
    Shan, Shanshan
    Waddington, David
    Whelan, Brendan
    Dong, Bin
    Liney, Gary
    Keall, Paul
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2023, 25
  • [37] A phantom based laser marking workflow to visually assess geometric image distortion in magnetic resonance guided radiotherapy
    Drobnitzky, Matthias
    vom Endt, Axel
    Dewdney, Andrew
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2021, 17 : 95 - 99
  • [38] Magnetic Resonance Imaging Distortion in Functional Neurosurgery
    Sankar, Tejas
    Lozano, Andres M.
    WORLD NEUROSURGERY, 2011, 75 (01) : 29 - 31
  • [39] Magnetic Resonance Guided Radiotherapy for Head and Neck Cancers
    Gharzai, Laila A.
    Rosen, Benjamin S.
    Mittal, Bharat
    Mierzwa, Michelle L.
    Yadav, Poonam
    JOURNAL OF CLINICAL MEDICINE, 2022, 11 (05)
  • [40] Development and validation of a novel large field of view phantom and a software module for the quality assurance of geometric distortion in magnetic resonance imaging
    Torfeh, Tarraf
    Hammoud, Rabih
    McGarry, Maeve
    Al-Hammadi, Noora
    Perkins, Gregory
    MAGNETIC RESONANCE IMAGING, 2015, 33 (07) : 939 - 949